Worksheet Page: ______

RD#:______

Date:______LAB#:______

Analyst:______Batch # ______

QC Lot#:______

Profiler Plus and/or COfiler Sample/CE Set-up

Standards and Controls:

1) Reagent Blank: as specified in the extraction procedure used

2) Substrate Control: if used in the extraction process

3) QC sample: one required per case

4) Positive Amplification Control: extracted DNA of specific, known type supplied with the

Profiler Plus or COfiler Typing kit plus amplification reagents;

5) Negative Amplification Control: Profiler Plus or COfiler amplification reagents, but no added DNA

__1. Remove samples from the refrigerator/freezer and allow thawing if necessary. Remove the Blue Ice trays from the freezer. Turn on thermal cycler and run File No. 13. (File 13 is programmed at 95oC for 999 minutes.)

__2. Prepare a GS-500 Internal Size Standard Master Mix solution as follows:

(#Samples_____ +2 = _____) X 26 mL de-ionized formamide = ______mL

(#Samples_____ +2 = _____) X 0.6mL GS-500 = ______mL

Gently mix and centrifuge briefly.

__3. Aliquot 26 mL of each of GS-500 Master Mix into each of the required 0.5 mL injection sample tubes. The tubes are labeled with the tray position and sample description.

__4. Transfer 0 to 3.0 mL of each of the samples or 1.5 mL Profiler Plus or COfiler Allelic Ladder to the appropriately labeled tube containing the GS-500 Master Mix. Gently mix each sample while adding to the GS-500 Master Mix using the pipette tip. One tube does not contain amplified product (GS500+formamide only). Cap each injection sample tube with a septum.

__5. Place all injection sample tubes in a thermal cycler and denature for 3 minutes at 95oC

(File No. 13 is programmed to heat at 95oC for 999 minutes).

__6. Snap-chill the injection samples in the Blue Ice block for at least 3 minutes.

__7. Place each injection sample tube in a 48 well tray according to the ABI Prism 310 CE Set-up sheet and the tray designation on the tube.

__8. Set-up the ABI Prism 310 CE as per SOP

Number of injections on capillary:______New capillary needed?: Yes___ No___.