Laboratory Instruments needed for مواصفات عطاء (9/2015)

Faculty of Medicine

No / Description / Qty / Country of origin
1 / Aesthesiometers
  • With a 12/100 mm nylon thread for measuring the tactile sensitivity of the Cornea and eyelid margin.
  • Can be used pre-operative or for post-operative recovery.
  • Provides direct reading on the instrument of the length of the very precisely calibrated filament.
  • The length of the thread like mono-filament, controlled by the forefinger, ranges from 60mm to 5mm.
  • Includes conversion table, which translates filament length in millimeters to a measure of mercury pressure (mm HG).
  • the tip of the monofilament is fully retracted into the instrument body
  • Includes a spare nylon filament and case
/ 2 / Japan,*EU, USA
2 / Anaerobic Jar
  • Anaerobic System with vacuum cum pressure gauge.
  • Supplied with transparent, unbreakable polycarbonate jar of 3.5-10 liter capacity.
  • Having sturdy aluminum lid, clamp and sealing ring with built in safety features.
  • Pressure valve with safety valve and two way pressure gauge.
  • Supplied with SS test tube carrier, SS petri-plate carrier, one box of 10 pcs of anaerobic indicator tablets.
  • Accessories – Autoclave portable to sterilize the media before using for culture and incubation in the anaerobic jar, Pair of Chuck and Clips, Dial Manometer Mounted on Chuck, Oxide Low Temperature Catalyst Sachet, 10 Plate Petri Dish Rack, , 48 Plate Petri Dish Rack, Anaerobia’s Indicator (test strip)
/ 2 / Japan, EU, USA
3 / Bench mark mini-centrifuge
  • RPM: 6000- 7000rpm
  • RCF 2,200 [×g]
  • Capacity : Micro tubeRotor 6 x 1.5/2.0ml
  • PCR Strip Rotor 2 x 8 strips or 16 x 0.2ml tubes
  • Operating temp. range: +4 C to 50 C
  • Power: 100V to 230V, 50/60 Hz
/ 3 / Japan, EU, USA
4 / Binocular Microscope
  • Body-Inter changeable, inclined Binocular body, 360º rotatable head , 30˚ incline, ( may have a camera).
  • Eyepieces-Highest quality 10 X wide angle anti fungus field eyepiece.
  • Objectives-Parfocal, anti-fungus coated 4x, 10x, 40x and 100x (oil immersion) with plan achromatic correction
  • Optical system-Infinity corrected
  • Stage –coated Horizontal mechanical stage preferably 100 x 140 mm with fine Vernier graduations designed with convenient coaxial adjustment lower position separate knobs for X and Y movement with rack & pinion ,for slide manipulation preferably through 30 x 70 mm
  • Sub stage-Abbe condenser focusable, continuously variable iris diaphragm
  • Illuminator-Built-in LED light source with white light.
  • Finish-A durable textured acid resistant finish.
  • Observation methods: Bright-field, Dark-field, Phase contrast with required accessories (Phase contrast attachment, Object marker, Dark-field ring unit, Mirror unit, Cord hanger).
  • Other Features: Should provide with storage box, dust cover, immersion oil.
  • Power: 200 to 240Vac 50 Hz supply.
/ 20 / Japan or EU
5 / Centrifuge (100 , 50 , 10 , 5 mL)
  • LED screen shown RPM/RCF (on steps 100 RPM/10xg) and time.
  • Intuitive handling through start, stops, short spin and play keys.
  • Rotor list on memory.
  • Timer count up/down, from 0 or at “set RPM/RCF” for test reproducibility.
  • Short spin key with speed adjustable.
  • Acceleration and braking selectable by operator
  • Automatic open lid.
  • Last values remain on memory.
  • Lid latch lock.
  • Microprocessor controlled.
  • Refrigerated with TemperatureRange (Metric) -9° to +40°C.
  • Capacity 4 x 100mL (400mL) at least
  • Induction motor maintenance free.
  • Quiet.
  • Option: free or locked adjustment of RPM/RCF along the run.
  • Accessories : ROTOR ( 12*15 , 18*15 , 24*15 , 6*50, 8*15 , 4*50 , 4*100 ) mL .
/ 2 / Japan, EU, USA
6 / Colony counter
  • Pressure sensitive counting
  • Digital display and audible tone confirms each count made.
  • Magnify petri plate images
  • Easy counting process on touchpad screen.
  • Hold function to stop counting at any stage
  • Alarm for counting deviations and out of range concentration.
  • Streamlined design enables fast and accurate count & reduces operator fatigue.
  • Bright white energy saving LED lighting with Bio-Cote antimicrobial protection.
  • Choice of light or dark background
  • Connectivity to printer or computer
  • Marks counted colonies on screen
  • Accessories : Marking Pen , Magnifier Lens
/ 1 / EU, USA
7 / Digital magnetic stirrer with heating
  • LEDDigitalMulti-point Magnetic Hotplate Stirrer
  • Display target and actual temperatures
  • External temperature sensor , display and control actual medium temperature, control accuracy of ±1°C.
  • Separated safety circuits, fixed safety temperature of 580°C. Automatically stops heating once exceed secure temperature
  • HOT warning indicates residual hotplate temperature.
  • LCD display would show Hot when hotplate temperature over 50°C warning point, even if switched off.
  • Up t0 .5 L stirring capacity per point
  • Digital timer function .
/ 2 / Japan, EU, USA
8 / Electrolyte Instrument
  • Electrolyte results can be obtained from just 95 μl of whole blood, serum, plasma, acetate, bicarbonate dialysate or bicarbonate dialysate, or pre-diluted urine using a simple Yes/No dialogue.
  • Several kinds of sample container – collection tube, syringe, capillary or sample cup.
  • All reagents, as well as waste, are sealed in the convenient container.
  • The Electrolyte Analyzer monitors the amount of remaining reagent and prompts the user to insert a fresh container when required.
  • Calibration: A two-point calibration is performed every four hours and a one-point calibration with each test.
  • Analyzer can automatically enter stand-by mode during idle periods.
  • choice of different, interchangeable electrolyte configurations to fit your immediate and future testing needs, changing the configuration by substitute the electrode and recalibrate with no need to change reagents or replace any tubing.
  • Sample Type: whole blood, serum, plasma, urine dialysate, aqueous stds, QC.
  • Sample Application: syringe, sample cup, collection tube, and capillary tube.
  • Analysis Time : 50 seconds
  • Data Management : quality control storage: 3 levels, 34 days; calculation of mean, SD and CV.
  • Normal Values: flagging of abnormal results; user programmable ranges.
  • Display : dot matrix, 2 lines, 16 characters wide
  • Printout: built-in, thermal roll printer; 16 characters wide.
  • Computer Interface : RS-232C (standard serial port)
  • Yes / No easy operation.
/ 1 / Japan, EU, USA
9 / Electrophoresis set (Horizontal)
  • Arrow on the side of the base indicates the direction of the run
  • Color-coded, labeled electrodes and labeled base.
  • Clear plastic
  • UV-transparent gel trays with fluorescent ruler
  • Gel-casting gates
  • Combs (multichannel pipet–compatible combs, fixed-height drop-in combs, adjustable-height combs, and preparative combs)
  • Cell size W x L x H approx. 18 x 25 x 7
  • Gel tray size W x L approx. 15 x 7 & 15 x 10 cm
  • Sample throughput 10-60
  • Base buffer volume ~650 ml
/ 1 / Japan, EU, USA
10 / Electrophoresis set (vertical)
  • Numbers of gels: 1-4
  • Gel size (WxL): approx. 8.6 x 6.8
  • Total buffer volume: from 700 to 1000 ml
  • Typical running time for SDS-PAGE: 35-45 (at 200 V constant)
  • Dimensions (W x L x H): approx. 12 x 16 x 18
/ 1 / Japan, EU, USA
11 / Power supply
  • Output specification: 500 V, 205 A, 500 W
  • Output range (programmable): 10–500 V, fully adjustable in 1 V steps
    0.01–2.5 A, fully adjustable in 0.001 A steps
    1–500 W, fully adjustable in 1 W steps
  • Type of output: Constant voltage, constant current, or constant power with automatic crossover
  • Output terminals: 4 pair recessed banana jacks floating in parallel
  • Timer control: 1 min–99 hr 59 min, fully adjustable
  • Volt-hour control: 99,000 V-hr
  • Real-time clock
  • Automatic recovery after power failure
  • Safety features: No-load detection, sudden load change detection, ground leak detection, overload/short circuit protection, overvoltage detection, input line protection, auto power-up after power failure
  • Input power (nominal): 100–120/220–240 VAC, 50/60 Hz, auto-switching
  • Suitable for horizontal and vertical electrophoresis
  • Operating conditions: 0–40°C, 0–90% humidity in the absence of condensation
/ 2 / Japan, EU, USA
12 / Embedding station
  • Single workstation that enables both specimen orientation and paraffin embedding.
  • finger touch plate and optional foot pedal for precise control of paraffin flow
  • dispensing tower knob to adjust the flow rate to specific tasks
  • automatic ON/OFF switch that allows manual override
  • temperature control over key areas for uniform regulation
  • Individual temperature control for left and right Thermal Storage Compartments
  • Panel display of temperatures in all heated and cooled areas.
  • Temperature range of paraffin: ambient to 99°C
  • Temperature range of embedding platform: ambient to 99°C
  • Temperature range of forceps warmer: ambient to 99°C
  • Safety cut-out: 100°C
  • Temperature range of cooling table: ambient to -30°C.
  • Designed to orient even small and big samples (Mega Cassettes).
  • easy-to-reach forceps wells
  • Paraffin Chamber Capacity : 4 Liters
  • Left and Right Thermal Chamber Capacity: 1.5 liters each
  • The tissue embedding cassettes and base mold trays are removable and can be set at different temperatures.
  • Easy-to-clean surfaces with no hidden corners to catch paraffin or debris.
/ 1 / Japan, EU, USA
13 / Fluorescence microscope
  • Viewing head: Trinocular, 30° inclined, Interpupilary distance 48-75mm
  • Camera: CMOS 3.0 megapixels, 1600 x 1280
  • Objectives: 4X, 10X, 40X, 100X
  • Nosepiece: Quadruple
  • Total Magnification: 40X - 1000 X
  • Stage: Double layer mechanical 185 x 142 / 75 x 55mm
  • Condenser: Swing Condenser NA 0.9/0.25
  • Focus: Coaxial coarse plus fine. Fine division = .001mm
  • Filter: Blue Excitation: BP 460-490, DM500, BA520 Green Excitation: BP 510-550, DM570, BA590
  • Illumination: Kohler External Aspherical Collector, halogen lamp 6V/20W
  • Software: Image Capture in BMP and JPG, Live Video on PC Monitor, Functions Include: Previewing, Saving, Measuring, Comparing, Counting and Deleting
/ 1 / Japan, EU, USA
14 / Gel documentation system
  • System optimization at setup
  • Comprehensive, automated, quantitative analysis of protein and DNA samples in seconds
  • Focused images at all times, regardless of zoom level or sample position
  • Appropriate, automatic, and consistent flat fielding correction of image data for every application
  • Automatic correction of imaging artifacts
  • Applications: Chemiluminescence / Fluorescence / Colorimetry/ Densitometry / Gel Documentation
  • Lighting: Trans-UV Light / Epi White Light
  • Image Lab Analysis software
  • Supercooled CCD camera
/ 1 / Japan, EU, USA
15 / Glassware washer
  • Digital Microprocessor Control with digital display.
  • Have top and bottom racks that accommodate accessory inserts for a wide variety of glassware shapes and sizes, primarily beakers and other wide-mouth glassware.
  • Specialized inserts for BOD bottles, petri dishes, culture tubes and small utensils.
  • Upper and lower racks, of stainless steel.
  • brushed stainless steel door, tank, upper and lower wash arms
  • LCD information center for customizing cycles and monitoring operation, including internal water temperature
  • Water recirculation
  • Dual pumps, one for washing and one for draining, to reduce potential for cross contamination.
  • Built-in purified water pump.
  • Built-in steam generator.
  • Manual-fill detergent dispenser for powder or liquid detergent
  • Manual-fill rinse aid solution dispenser.
  • Built-in forced air drying.
  • Factory-set cycle programs.
  • Two user-set cycle programs
  • Programs may be locked to prevent unauthorized changes.
  • Alarms with display.
  • Protection from debris provided by filter traps.
  • Conservative water usage, with Configure to your own requirements with choice of de-ionized water or tap water rinse
/ 1 / Japan, EU, USA
16 / Ice maker
  • Fully automatic controller
  • Constructed from sturdy, corrosion-resistant ,stainless steel exterior
  • Available with an air-or water-or remote-cooled condenser.
  • Simple and reliable electromechanic controls.
  • Easily accessible front panel mount ON-OFF Switch
  • Routine Maintenance visible alarm light on front panel
  • Ergonomically designed ice storage access, with disappearing door.
  • Front panel in & out airflow for built-in installation
  • Front access condenser air filter, removable and washable.
  • Water system protected by patented anti-scale system
  • PWD System for remote water drain (easy-fit installation)
  • AgION antimicrobial agent embedded in moulded plastics parts.
/ 2 / Japan, EU, USA
17 / Infrared microsterilizer (instead of Bunsen burner)
  • Safe, convenient and compact sterilizer for inoculating loops, needles and glass tubes, pipette mouths, metal and borosilicate glass instruments
  • Asbestos-free core element producing 825℃
  • Rapid and complete infrared sterilization in 5-7 seconds
  • Prevent infectious spatter and cross contamination
  • Adjustable angle heating chamber
  • Can be used in anaerobic and aerobic chamber
  • Safeguard users without open flame and hazardous gas
  • Ceramic funnel tube enclosed in stainless-steel perforated guard and stainless-steel support stand
/ 10 / Japan, EU, USA
18 / Mini Dry bath
  • LCD-display.
  • Temperature range from - 10°C to +100°C.
  • The lid of the dry block heat bath may be closed.
  • Block with different capacity for tubes.
  • Temperature stability.
  • The temperature settings are made by touch pad
/ 2 / Japan, EU, USA
19 / Moisture oven
  • Temperature range: 40 to 250ºC
  • Fluctuation +/- 0.75ºC.
  • Fan circulation.
  • Easy clean powder coated body.
  • Aluminum coated mild steel chamber.
  • Microprocessor digital controller with dual display of set point and actual temperature.
  • Safety overheat thermostat.
  • Exterior is constructed from sheet steel finished in an easy clean powder coated paint.
/ 2 / Japan, EU, USA
20 / Ophthalmoscopes
  • Halogen HPX® lamp illumination for true tissue color
  • 18 aperture/filter combinations for greater versatility
  • Polarizing filter virtually eliminates corneal reflection
  • Detect corneal abrasions with the cobalt blue filter
  • Compatible with all 3.5 V power sources
/ 2 / Japan, EU, USA
21 / Oven
•Stainless Steel w x h x d: 585 x 704 x 434 mm
•fully insulated stainless steel door with 2-point locking
•Interior - Heating Concept
•w x h x d: 400 x 320 x 250 mm
•easy-to-clean interior, made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides
•nearly gas-tight interior
•Max. loading of chamber: 60 kg
•TemperatureRange min. 5°C above ambient up to +80°C / 2 / Japan, EU, USA
22 / PH meter
  • Range (PH): -2.00 to 16.00
  • Range (mV): ± 2000
  • Range (Temperature °C): -10 to 110
  • Resolution (pH): 0.1/0.01
  • Resolution (Temperature): 0.1
  • Resolution (mV): 0.1/1
  • Accuracy (pH): ±0.01 pH
  • Accuracy (Temperature): ±0.5°C
  • Automatic temperature compensation (ATC) maintains reading accuracy.
  • Five-point pH calibration, with automatic buffer recognition for both USA (pH 1.68, 4.01, 7.00, 10.01, 12.45) and NIST (pH 1.68, 4.01, 6.86, 9.18, 12.45) pH buffers.
  • Universal 110/240 VAC power supply and adapters.
/ 4 / Japan, EU, USA
23 / Post mortem hammers
9.5" (24.1cm), with hook handle / 3 / Japan, EU, USA
24 / Refrigerated centrifuge
•Speed range: 250 to 15,000 rpm
•Timer: 1 to 60 min + short spin function
•Power: 220 V/ 50 Hz
•Digital display
•Safety lock system
•Temp range: -10 to 40 °C
•Rotor for: 1.5 ml and 2 ml
0.5 ml and 0.6 ml
0.25 ml and 0.4 ml
0.2 ml PCR tubes / 1 / Japan, EU, USA
25 / Refrigerator
  • 32-3/4’ wide
  • Upfront temperature controls
  • Adjustable spill-resistant shelves
  • Adjustable-humidity drawers
  • Temperature Management Features: AirTower in Freezer
  • Defrost Type: Frost Free
  • Door locker
/ 7 / Japan, USA,
**Far east
26 / Rotator Instrument
•Speed range: 1 to 40 rpm
•Capacity: 12 x 50ml, 20 x 15 ml or 48 x 2.0ml
•Timer: 1 Sec. to 9 hour or continues
•Dimensions (wxdxh): 21 x 27 x 32 cm
•Operating Temp. Range: +4°C to +60°C
•Power: 100 to 240V, 50-60Hz, 25W / 1 / Japan, EU, USA
27 / Shaker Incubator
  • Maximum load: 8 kg
  • Platform Dimensions: approx. 44 x 45 x 2cm
  • Temp Uniformity: +/- 0.5 C at 37 C
  • Temp range: -8 to 60 C
  • Exterior Dimensions w x d x h 55.9 x 64.8 x 71.2 cm
  • Interior Dimensions w x d x h 48.2 x 45.7 x 41.9 cm
  • Capacity: approx. 90L
  • Shaking Speed: 30 to 300 rpm
  • Power: 220V, 50-60Hz
  • Load Flexibility
  • Shaking Platform Included
  • Independent Controls for Temperature, Timer, and Shaker
  • Accommodates flasks up to 3 L as well as test tube racks, Microplate holders, Petri-dishes, plates, trays, etc.
  • Over-temperature Alarm
/ 2 / Japan, EU, USA
28 / Slide warmer
  • Thermal heater ensures heat transfer over the entire surface
  • Anodized black surface provides contrast with samples
  • Thermoplast setting range from room temperature to 70°C ±2°
  • Power 220V, 60Hz, 300 Watts
  • 25" x 8" (64 cm x 20 cm)
  • Surface holds about 66 slides
/ 3 / Japan, EU, USA
29 / Spirometer
Measurements: VC, FEV0.75, FEV1, FEV3, FEV6, FVC, PEF FEV0.75/VC, FEV0.75/FVC, FEV1/VC, FEV1/FVC, FEV3/VC, FEV3/FVC, FEV0.75/FEV6, FEV1/FEV6, MEF75, MEF50, MEF25, MEF25-75, FEF50/VC, FEF50FVC, MVVind, FET, MET, IC, FIV1, FIVC, PIF, FIV1,/FIVC, FIF25, FIF50, FIF75, FIF50/MEF50,VT, ERV, IRV [all expiratory measurements with baseline, post bronchodilator 1 and 2, % predicted, % change and normal range]
Transducer: Gold Standard Bi-Directional Digital Volume
Resolution: 10ml volume 0.03l/s flow
Accuracy: ± 3% (to ATS recommendations)
Storage: 1000 patients' tests including Flow/Volume loops and Volume/Time curves
Power: 100 - 240V 0.3A 47 - 63Hz. DC output 9V 0.8A
Dimensions: 150h x 251w x 40d mm. Transducer 50 x 60 x 90mm
Temperature: 0 to +40°C
Operating humidity: 30% to 90% RH
Storage temperature: 0 to +70°C
Storage humidity: 10% to 90% RH / 2 / Japan, EU, USA
30 / Stereo Microscope
  • Zoom drive: horizontal knob system
  • Knob rotation tension adjustment
  • Click stop for each zoom magnification: ON-OFF switching possible
  • Zoom ration values: 7 : 1 (0.8X to 5.6)
  • Zoom magnification indication: 0.8. 1, 1.25, 1.6, 2, 2.5, 3.2, 4, 5, 5.6
  • Binocular tube view inclination angle 45
  • Mounting diameter 76mm
  • Focusing stroke 120mm
/ 10 / Japan or EU
31 / Stethoscopes / 5 / Japan, EU, USA
32 / Stimulator
  • Wave Form: Square
  • Stimulus Amplitude: 0-100 volts
  • Stimulus Duration: 0.04ms to 50ms
  • Frequency: 1-1000 pulses per second
  • Sync-Output: 10v square wave
  • Stimulator Protection: Isolated transformer
  • Fuse: 1 amp, 3 AG
  • Signal Magnet Output: Activated magnet (2 ohms or more) once for each stimulus at frequencies of up to 20 stimuli per second.
  • Operational Modes: Continuous automatic, continuous manual, and manual single pulse control.
  • Power: 220V, 35 W max, 50-60 Hz
/ 1 / Japan, EU, USA
33 / Tissue processor (Bench top)
  • Basket Capacity: 100 cassettes
  • Reagent Capacity: 1.8L
  • Height: 50 to 70cm.
  • Voltage: 100-240V Hertz 50/60Hz
  • Ten reagent vessels and two paraffin baths for single basket operation
  • Includes ventilation system, Two paraffin baths, One processing basket
  • Diameter approx. 85cm
  • Power: 100-240V 50/60Hz
/ 1 / Japan, EU, USA
34 / Top loading Balance
  • Capacity (g) 1020
  • Readability (g): 0.001
  • Repeatability (g): 0.001
  • Linearity +/- (g): 0.002
/ 3 / Japan, EU, USA
35 / Tuning forks set / 1 / Japan, EU, USA
36 / UV spectrophotometer
  • Wavelength range : 190 to 1000 nm
  • Wavelength accuracy : ± 1 nm
  • Wavelength resolution : 0.1 nm
  • Photometric ranges:
Transmittance: 0 to 200% T
Absorbance: - 0.7 to 2.5 A
Concentration: 0 to 9999 C
  • Source lamp: long life halogen lamb
  • Power : 220 V/ 50 Hz
  • Input/output RS-232 and parallel port for printer.
/ 2 / Japan, EU, USA
37 / Vortex mixer
  • Speed range: 100 to 3000 rpm
  • Power: 220 V/ 50 Hz
  • Touch (continuous) and hand-free operation
  • Suction cups or rubber pad
/ 5 / Japan, EU, USA
38 / Water bath with Lid
  • Capacity: from 6 to 15L
  • Temperature range Ambient: +5°C to 99.9°C
  • Temperature stability: +/- 0.5C
  • Internal dimensions (w x d x h),mm: 300 x150 x 200 or 300 x 325 x 200
  • Overall dimensions (w x d x h), mm: 335 x 230 x 280 or 335 x 408 x 280
  • Heater Power: From 500W to 1000W
  • Electrical supply 230V, 50Hz
/ 3 / Japan, EU, USA
39 / Reverse osmosis system
  • Water flow rates from 6 to 24 liters per hour
  • CSA and CE listed
  • Can be wall or bench mounted
  • supplied with everything needed for operation—arrives complete with prefilter, MPS (Membrane Protection System) cartridge, extruded carbon filter and membrane(s) s
  • Membrane Protection System cartridge
  • Automatic flush cleanses the RO membranes
  • A series of internal pressure gauges to monitor membrane operating pressure and pressure drops across the pretreatment filters
  • Fully automatic operation
  • Monitor system operation with indicator lights on the control panel alerting of vital information including membrane performance and reservoir level
  • Control panel displays the status of the system including "Stand By" indicating that the reservoir is full
  • includes a special Membrane Protection System (MPS)
  • System alerts if the carbon needs changing or the membrane is not functioning
  • Suitable for laboratory uses.
/ 1 / Japan, EU, USA
40 / CO2 incubator
  • TemperatureRange 4 °C above ambient to 50 °C
  • Control ± 0.1 °C
  • Stability ± 0.1 °C at 37 °C
  • Uniformity (on platform) ± 0.25 °C at 37 °C
  • (ambient temperature between 18 and 25 °C)
  • CO2 0.2 to 20 %
  • PlatformDimensions 612 x 356 mm (24 x 14 in)
  • OverallDimensions (W x D x H) 834 x 730 x 848 mm (32.8 x 28.7 x 33.4 in)
/ 1 / Japan, EU, USA
41 / Orbital shaker
  • Speed mixing: from 4-100 rpm's.
  • Orbit diameter: 1.5 inches (38.1 mm).
  • Platform size: 12 in. x 16 in. (304.8 x 406.4 mm).
  • Non-slip rubber surface
  • Operates in ambient temperatures from -10C. to 65 C. and 80% relative humidity in a non-condensing environment.
  • Cold-room safe
/ 2 / Japan, EU, USA
42 / 3 Bodies Mortuary Freezer
  • High accuracy temperature control system to keep the temperature insidethe refrigerator accurate and stable at -5 to 5 C
  • The set point can be adjusted and corrected with a correctness range of 0.5 C
  • Safe control system
  • Safe door lock mechanism
  • High efficiency compressor
  • Temperature can be set manually according to demand
  • Tray size: 190X60X40 cm
  • AC: single phase 220V 50HZ
/ 2 / Japan, EU, USA or Local
43 / Bio-Monitor
  • Used to detect and display bioelectrical events associated with cardiac and striated muscle activity.
  • Provides both visual (flashing light) and audible (beep tone) responses.
  • Comes with all the accessories needed to perform experiment, including:
-Determining Mean Heartbeat
-Heart Rate Differences by Sex and Height
-Effects of Rapid Breathing and Exercising
/ 1 / Japan, EU, USA
44 / Balanced Biological Bio-Amplifier
  • A multi-purpose, medium sensitivity amplifier for use in conducting cardiac muscle potentials, nerve and striated muscle potentials, and cortical and electroencephalographic signals.
/ 1 / Japan, EU, USA
45 / Blood Pressure Experiments Kit
  • With a blood pressure cuff and stethoscope.
/ 4 / Japan, EU, USA
46 / Ergograph
  • For studying the physiology of muscle fatigue.
  • Used with a kymograph
  • Provides fatigue curve tracings from the student's index finger working against an adjustable tension coil spring.
/ 1 / Japan, EU, USA
47 / Isolated Organ / Tissue Bath
  • Study the effects of drugs or electrical stimuli on isolated organs and smooth muscle segments of the intestine, uterus, ureter, trachea, vessel segments, etc.
  • With Muscle chamber accessories.
  • Comes complete with heater/agitator and muscle chambers.
/ 1 / Japan, EU, USA
48 / Morgue Trolley
  • Trolley made of stainless steel
  • Trolley wheels have arrest system
  • Dimensions (L x W x H): 180 x 70 x 90 cm
/ 2 / EU or Local
49 / Human Skeleton + separated bones / 10 / EU, USA, Far East
50 / HD LED TV screen 42”
  • LED backlighting technology
  • Thinner bezel improves display's aesthetics
  • RS-232C allows users to lock out external buttons, synchronize multiple displays and remotely power on/off displays
  • USB media player allows users to view .jpg images
  • Full 1080p high-definition resolution provides stunning detail
  • Built-in NTSC/ATSC analog/digital tuner allows for high-definition broadcast capabilities
  • Built-in, low-profile 10W speakers
/ 4 / Far East
51 / Data show projector / 2 / Far East, EU
52 / Desktop Computer / 2 / Far East
53 /

Educational Microscope

High-intensity LED Eco-illumination

Camera compatibility: A digital camera can be attached when the trinocular eyepiece tube is used.

  • Series objective lenses for flat images 4x, 10x, 20x, 40x, 60x, 100x (Oil)
  • Trinocular eyepiece tube allows image capturing
  • Observation methods: Bright-field, Dark-field, Phase contrast with required accessories (Phase contrast attachment, Object marker, Dark-field ring unit, Mirror unit, Cord hanger).
  • Objectives and sliders dedicated for phase contrast and darkfield observation, even an unstained specimen that is colourless and transparent can be observed with clear contrast.
  • The condenser comes with an aperture diaphragm with position guide markings for each objective magnification to make operations quick and easy.

Double-plate mechanical stage

Easy halogen lamp replacement

Anti-mold design

/ 1 / Japan, EU, USA
54 /

Cabinets for keeping dry samples

/ 4 / Local, EU, Far East

* EU: European countries