GRF-25-08 Rev. 16.1Page 1 of 51
Medical Devices – Substantial Change Review Form
Applicant Information and Instructions
Significant/ Substantial Change Notification Covering Product Family-
MDD 252. NSAI Use Only – Amendment Number-/
AIMD 253. NSAI Use Only – Amendment Number-/
TSE Human Blood Medicinal Substances
P/O number:
This is a:
Regular/ Standard Review Fast Track (expedited)
Modular submission (prior agreement only)
If OBLs apply to this product, please state the relevant product families below:
/; /; /
DECLARATION(s) BY APPLICANT
In signing this form, the manufacturer is verifying that the requirements of the Directive will be applied in full when the change has been implemented.
Signed on behalf of the Manufacturer: / Date:Name (please print):
Position / Title:
Contact person (if different to Manufacturer):
e-mail: / Phone:
INSTRUCTIONS
- Please complete all relevant sections of the form (excluding the NSAI Review sections).
- Please enter as much information onto the form as possible - avoid entering “see Technical File/Design Dossier”. If the data is in supporting documentation, please ensure that there is a clear reference to the exact location of this information.
- Please submit an unsigned version of this Application in Word as well as a signed copy - either scanned/secured (pdf) copy.
- All application forms and supporting data to be forwarded in soft copy via one of the following (Hard copies not required)
- NSAI upload facility : see
- CD or Memory stick to the appropriate address
Europe / N. America
NSAI
1 Swift Square,
Northwood,
Santry, Dublin 9
Ireland
Phone : (01) 807 3929
Fax : (01) 807 3996
/ NSAI Inc.
402 Amherst Street
Nashua
NH 03063
USA
Phone : (603) 882 4412
Fax : (603) 882 1985
- Supporting documents should be in SEARCHABLE format
- Applications and supporting documentation must be in English
- Please send a representative sample of the device(s). This is particularly important for new/novel devices. Any video of procedures/simulated use would also be helpful, if available.
APPLICANTS’ SUBMISSION CHECKLIST
Completed application form (Word format, .doc or .docx)
Application (min. Signed Declaration page(s)) scanned
Sample of device(s)
QMS certificates for any sites in Table 1 NOT registered with NSAI
Type Examination Certificate if required
Declaration of Conformity
Labelling & IFU
Essential Requirements Checklist
Performance/Complaint Analysis
Risk Management documentation
Sterilisation Validation(s) – if sterile/intended to be sterilised
Stability data – if necessary
Biocompatibility data – if necessary
Electrical Safety Testing data – if necessary
Software/firmware lifecycle documents – if necessary
Bench Testing data – if necessary
Clinical investigation(s) report(s) and supporting documents per MEDDEV 2.7.1.
NSAI Equivalence table RF-25-28 if using equivalent route
NSAI Post Market Surveillance (PMS) and Post Market Clinical Follow Up (PMCF) RF-25-27
Note – NSAI do not accept “For Information Only Data” (FIO); all attributes tested shall have clinically relevant specifications
For Transfers
Copy of existing Notified Body Certificate(s)
(If not already supplied)
Transition Plan
Contact details for existing Notified Body
(NSAI will not contact the existing Notified Body prior to agreement with the Manufacturer)
For Own Brand Labeller (OBL)
Copy of the OEM (maker’s) CE Certificate and last assessment reports from their Notified Body
Copy of the OEM’s Declaration of Conformity
Traceability between OEM cert & devices covered in this application
Copy of the contract between OBL & OEM outlining responsibilities
Index for Technical documentation/Design dossier (version/date), listing supporting documents
Copy of the OEM’s Instructions for Use/Labelling
For Tissue of Animal Origin falling under Commission Regulation 722/2012
Please complete Section 16.1
For Human Blood Derivatives
Please complete Section 16.2
For Medicinal Substances.
Please complete Section 16.3
Table of Contents:
Applicant Information and Instructions
DECLARATION(s) BY APPLICANT
INSTRUCTIONS
APPLICANTS’ SUBMISSION CHECKLIST
Section 1 – Manufacturer and Product Details
Section 2 – Nature of Change
Section 3 – Intended Use of the Device
Section 4 – Correct Classification and Appropriate Annex
Section 5 – Existing Legislation
Section 6 – Labelling and IFU
Section 7 – Solutions to Essential Requirements and Harmonised Standards
Section 8 – Performance/ Complaint Analysis
Section 9 – Risk Management
Section 10 – Sterilisation
10.1 – Sterilisation Validation
10.2 – Maintenance of Sterility over shelf life
Section 11 –Biocompatibility
Section 12 –Medical Electrical Equipment and Systems
Section 13 – Device Testing
13.1 – Device Design Testing
13.2 - Device Stability
Section 14 - Clinical Testing (Animal Model)
14.1 In Vivo Clinical Performance
14.2 In Vivo Device Performance
Section 15 – Clinical Performance (Human)
15.1 Clinical Evaluation
Section 16.0-Appendices
16.1 – Regulation 722/2012 : TSE
16.2 – Human Blood Derivative(s)
16.3 – Devices incorporating Medicinal Substances
Section 17 – Design Dossier
Section 18 – Critical Process Changes
Section 19 - NSAI Queries
19.1 Queries, Responses and Dispositions.
19.2 Query Status Table 14:
Section 20 - Conclusion of Assessment
Approval
Section 1 – Manufacturer and Product Details
Table 1 – Manufacturers Information & Summary Product data
(Legal) Manufacturer’s Name(Legal) Manufacturer’s Address
Design Site(s):
Manufacturing Site(s):
(i.e. sites of actual manufacture)
Assembly Site(s) if applic.:
Sterilisation Site(s) if applic.:
Scope of Site(s):
(i.e. as shown on the QMS cert)
Name and address of EU Authorised Representative
(if applicable)
Product/Product Family Name:
(In compliance with NB/MED/2.5.1/REC4 & NBOG’S Best Practice Guide 2006-2)
GMDN Reference Number: / See
Declaration of Conformity included - Location within submission :
MDD ONLY :
Class / III IIb IIa Is Im / Rule(s)
Rationale
Conformity Assessment / Annex II / V (+VII) / V + III / VI
Full QA / Prodn QA / + Type Testing / Product QA
AIMD ONLY :
Conformity Assessment / Annex 2 / Annex 3 + 5
Full QA / Production QA + Type Testing
ALL DEVICES :
Clinical Strategy-
- Clinical data from: / Clinical Investigation
Literature (Equivalence)
Combination
Date of this application (i.e. date of Declaration of Applicant)
NSAI REVIEW
NSAI registration(s) if new site(s) -
If Annex II is “Design” included in scope Yes No
Scope of registration(s) adequate to cover the product family
Client QMS Certificates are valid and scope of registration(s) remains adequate for product family under review
Product family name in line with GMDN & remains valid
DoC reviewed - lists model numbers
Type Testing Certificate required – No Yes
if YES – Supplied, Valid
NSAI REVIEW
Technical Reviewer Date:
COMMENTS:
Clinical Reviewer (if applicable) Date:
COMMENTS:
Additional Reviewer (if applicable) Date:
COMMENTS:
QUERIES No YesNumber(s)
Section 2 – Nature of Change
- Device Description:
- Please provide a clear, detailed description of the change(s):
- Did the change(s) arise from a vigilance or performance issue
Yes No
If “Yes” – please advise -
If”No” what was the rationale/motivation for each change(s) made:
- Does the change involve an addition to the members of the product family Yes No
If “Yes” – please complete Table 2 below:
Table 2 – New Model/Catalogue Numbers
Sub-Family / NEW Model/Catalogue Number(s) / Description / Class- Please advise what models are affected by the change
Table 3 – Affected Model/Catalogue Number
Sub-Family / AFFECTED Model/Catalogue Number(s) / Description / Class- Impact of change
Table 4-Impact of Change
Impact / Yes / No- Is there a change to the solutions (documented evidence ) to the ERs?
- Is there a change to the Harmonised Standards?
- Does the change introduce new hazards which have not been previously addressed?
- Does the change adversely affect the risk profile associated with the existing hazards?
- Does the change alter the details on intended use and /or compliance with the essential requirements given in the design/type approval dossier?
- Does the change mean that the device will have different end users or be used in a different manner?
- Does the change mean that the clinical data /performance evaluation data for the original device is not sufficient to confirm conformity of the changed device with the required characteristics and performance?
NSAI REVIEW
Change adequately described
Modifications/changes considered vs. approved product family
Validity of reason(s) for change
NSAI REVIEW
Technical Reviewer Date:
COMMENTS:
Clinical Reviewer (if applicable) Date:
COMMENTS:
Additional Reviewer (if applicable) Date:
COMMENTS:
QUERIES No YesNumber(s)
Section 3 – Intended Use of the Device
- Please state the current Intended Use –
- For class III and Class IIb implantables ONLY,
Please outline the “Indications for use” (i.e. specific patient population intended for this device/family).
- Is there a change in Intended Use Yes No
- Is there a change in Indications for Use Yes No
NSAI REVIEW
Agreed that this section is not required Yes No
Reviewer Date:
Comments:
- Please enter a full description of the revised intended use and/ or indications for use of the device-
- Does this change impact the classification/rule Yes No
If “No” please justify -
NSAI REVIEW
Revised Intended Use clearly stated
Revised Intended Use in line with labelling
Impact of Revised Intended Use assessed – including classification/rule
NSAI REVIEW
Technical Reviewer Date:
COMMENTS:
Clinical Reviewer (if applicable) Date:
COMMENTS:
Additional Reviewer (if applicable) Date:
COMMENTS:
QUERIES No Yes Number(s)
Section 4 – Correct Classification and Appropriate Annex
- Please state the current MDD Classification- Rule:
- Is there a change in Classification Yes No
NSAI REVIEW
Agreed that this section is not required Yes No
Reviewer: Date:
Comments:
- MDD Only – Please enter the revised Classification of the device based on the change, along with the Rule & Rationale for the choice
Class
Rule
Rationale
NSAI REVIEW
All Rules reviewed
Intended Use described supports Classification
Agree with Classification - Yes No
NSAI REVIEW
Technical Reviewer Date:
COMMENTS:
Additional Reviewer (if applicable) Date:
COMMENTS:
QUERIES No YesNumber(s)
Section 5 – Existing Legislation
1. Is the product OBL (Own Brand Labelled) Yes No
*You are not the actual Maker of the device – and the identical product carries CE marking by the Maker.
NSAI REVIEW
Agreed that this section is not required Yes No
Reviewer: Date:
Comments:
Please complete all questions in this section:
- Has there been a change to the Maker’s CE cert Yes No
If “Yes” – please provide the Notified Body’s review of the change
Report/ Document Ref.
- If the Maker uses NSAI as their Notified Body, please provide the file reference for the review of the change(s) –
MDD : 252. /
AIMD: 253. /
NSAI REVIEW
Revised OEM cert
Report provide
Review adequate
NSAI REVIEW
Technical Reviewer:Date:
COMMENTS:
Additional Reviewer (if applicable): Date:
COMMENTS:
QUERIES No YesNumber(s)
Section 6 – Labelling and IFU
- Is there a change to the Labelling/IFU Yes No
If “No” please justify -
NSAI REVIEW
Agreed that this section is not required Yes No
Reviewer: Date:
Comments:
Please supply a sample of the revised draft labelling & IFU in English.
- Location of the sample Label(s) & IFU in the supporting documentation
- Please clarify the exact nature of change(s) to the labelling/IFU based on the proposed change(s) under review –
- Are the requirements of EN 15223-1 & EN 1041 being met
Yes No
Version of Standard –
If “No” Please justify -
- Is the device Medical Electrical Equipment Yes No
If “Yes”
a) Are the requirements of EN 60601-1 being met:
Yes No N/A
Version of Standard:
If “No” please justify -
b) Are the requirements of EN 60601-1-6 being met:
Yes No N/A
Version of Standard:
If “No” please justify -
6. Do any vertical labelling standards apply Yes No
(i.e. device-specific standards that require labelling)
If “Yes”, please list (including current version) -
If compliance with these vertical labelling standards is not claimed, please justify -
NSAI REVIEW
Labels reviewed – details
All relevant parts of ER 13, particularly 13.3 (MDD) or ER 14 (AIMD) met
IFU reviewed – details
All relevant parts of ER 13, particularly 13.6 (MDD) or ER 15 (AIMD) met
Ensure traceability between the Risk Management File and the CER and labelling
Symbols & Markings on ME Equipment/System are per Table D.1, D.2, D.3 of EN 60601-1
Markings & IFU in line with EN 60601-1-2 (EMC)
For OBL – OEM IFU claims in line with OBL IFU
Vertical labelling standards -
NOTE – If an issue is identified with the sample labelling, ensure that the corrective action(s) are applied to all labelling operations.
NSAI REVIEW
Technical Reviewer Date:
COMMENTS:
Clinical Reviewer (if applicable) Date:
COMMENTS:
Additional Reviewer (if applicable) Date:
COMMENTS:
QUERIES No YesNumber(s)
Section 7 – Solutions to Essential Requirements and Harmonised Standards
Please indicate how relevant Essential Requirements (Annex I) of the Directive are met for the proposed changes.
- Location of the revised solutions to Essential Requirements in the supporting documentation
FOR CLASS I STERILE devices: In particular ER 8
FOR CLASS I MEASURING devices: In particular ER 10
- For devices other than Class I Sterile &/or Measuring please ensure there is specific reference to the design elements associated with this application and demonstrate how the design outputs meet the design inputs through design verification and validation
- Please list the relevant revised/updated Harmonised Standards in Table 5 below
Table 5 – Applicable Harmonised Standards List
Standard / Year / Full ComplianceYes/No
FOR MDD CLASS I STERILE devices: In particular Harmonised Standards relating to sterility
FOR MDD CLASS I MEASURING device: In particular Harmonised Standards relating to the measuring function.
NSAI REVIEW
Class I Sterile
ER 8 adequately addressed -
Harmonised Standards for Sterilisation -
Class I Measuring
ER 10 adequately addressed -
Harmonised Standard(s) for Measuring -
Other classes of MDD devices & AIMDs
ERC Document Rev.
ERs identified as N/A
Agree with non-applicability
Has each subpart in the ERs been adequately addressed, if applicable
Solutions to ERs complete
Ref to the supporting documentation adequate
Harmonised standards used
Adequately demonstrates that design outputs meet design inputs through design verification & validation activities
NSAI REVIEW
Technical Reviewer Date:
COMMENTS:
Clinical Reviewer (if applicable) Date:
COMMENTS:
Additional Reviewer (if applicable) Date:
COMMENTS:
QUERIES No YesNumber(s)
Section 8 – Performance/ Complaint Analysis
- Is the change due to Vigilance or Performance Issue Yes No
NSAI REVIEW
Agreed that this section is not required Yes No
Reviewer Date:
Comments:
- If “Yes” please give details -
- HasNSAI received the Vigilance Report(s) Yes No
If “Yes” please provide the relevant Unique Identifier number(s) –
`If “No” please:
(a)Justify:
(b)If applicable, please submit a copy of the Competent Authority report(s) along with the completed NSAI Vigilance Form located at [ -“Vigilance Reporting”] to
- Has this product been the subject of product recalls or Incident Reports in other Regulatory geographies outside EU? If yes, please summarize and provide details with supporting documentation.
- For those failure modes associated with the identified Root Causes, please clarify if the Occurrence Rates outlined in the Risk Management File required an update based on the observed real world rates.
NSAI REVIEW
Vigilance reports reviewed
Trending data supplied reviewed
Negative review/refusal assessed
NSAI REVIEW
Technical Reviewer Date:
COMMENTS:
Clinical Reviewer: Date:
COMMENTS:
Additional Reviewer Date:
COMMENTS:
QUERIES No YesNumber(s)
Section 9 – Risk Management
- Did the proposed change affect or change any existing risks
Yes No
If “No” please justify -
- Did the proposed change introduce any new risks Yes No
If “No” please justify -
- Was the Risk review documented Yes No
(e.g. during change control process, update to FMEA, Memo to file etc.)
If “No” please justify -
- If no update to Risk Management File, please provide rationale:
NSAI REVIEW
Agreed that this section is not required Yes No
Reviewer Date:
Comments:
MANUFACTURERS – Where reviewed and updated, please provide
Risk Mgmt File - Risk Analysis, Risk Management Report, including a signed and dated conclusion regarding residual/remaining risks, which clearly identifies or highlights risk(s) in relation to the proposed change(s).
NOTE: Additional information may be required for devices containing tissue of animal origin, human blood derivatives or medicinal substances: see APPENDICES.
FOR CLASS I STERILE devices: In particular risks related to sterility
FOR CLASS I MEASURING devices: In particular risks related to measuring function.
- Risk management per EN ISO 14971 latest version Yes No
Version of Standard:
If “No” please justify -
- Which of the multifunctional team provided the clinical input – i.e. risks associated with the clinical use of the device
Please justify the appropriateness of their expertise -
- Please provide a traceability matrix linking the contraindications, warnings and precautions from Risk Management File to the Instructions For Use and Clinical Evaluation Report.
- If applicable, have relevant risk documentation for Medical Electrical Equipment, systems &/or Software been included Yes No
If “No” please justify -
NSAI REVIEW
Signed & dated conclusion regarding residual/remaining risks
Residual/remaining risks acceptable, based on intended application of the device
Compliance with ISO 14971 – version:
Suitable cross-functional team
Contra-indications/Warning/Precautions in IFU originate in Risk file
Risk analysis adequate
Solutions adopted by Manufacturer conform to safety principles, taking account of the generally accepted state of the art
Adequately covers the proposed changes
NSAI REVIEW
Technical Reviewer Date:
COMMENTS:
Clinical Reviewer (if applicable) Date:
Clinical risks (actual risks involved in the use, application of the treatment or deployment of the device) identified by suitably qualified person
COMMENTS:
Additional Reviewer (if applicable) Date:
COMMENTS:
QUERIES No YesNumber(s)
Section 10 – Sterilisation
10.1 – Sterilisation Validation
- Is the product provided sterile or intended to be sterilised prior to use Yes No
- Does the proposed change affect sterilisation Yes No
If “No” please justify -
NSAI REVIEW
Agreed that this section is not required Yes No
Reviewer: Date:
Comments:
MANUFACTURERS - Please provide the necessary sterilisation validation protocol(s) & report(s) and populate Table 6 below for the proposed changes