Chapter 4 – Calibration
Why calibrate?
-required by ______and ______
-ensures that results are as accurate as possible (eliminate equipment variations)
-necessary to ensure the audiometer produces a pure tone at the correct frequency and specified level via the selected transducer
-ensure that the signal is not distorted and is free of unwanted noise interference
Parameters of Calibration
-follow standards and procedures outlines in the equipment’s manual
-do regularly scheduled tests
-do listening tests
Basic Equipment
-basic calibration equipment for checking output levels:
- voltmeter or multimeter
- condenser ______
- 6-cm3 coupler
- 2-cm3 coupler
- ____ gram weight
- mechanical coupler for bone vibrator measurement
- ______meter or spectrum analyzer
- also possibly:
- oscilloscope
- graphic level recorder
- electric counter
- frequency analyzer
- distortion meter
Checking the Calibration of Puretone Audiometers
-basic signal
- read the instructions
-biologic test
- listen to the signal at different settings and through each tranducer
-frequency check
- use an electric counter to tell the exact ______of the output
-linearity check
- Change ______from highest dB level and check each step down individually
-distortion check
- distortion appears as a lack of linear attenuation
- presence of energy at other frequencies
- more common at high output levels
-rise-fall time
- put output directly into a storage ______
- ______time is the time between pressing the button and the signal being at its intended strength
- ______time is the time it takes from the stopping of the signal presentation for the level to return to its steady value
Ear Phone Calibration
-generally, the ear phones are tested on a group of people using real-ear methods
-can also use a coupler method which acts as an “artificial ear”
Bone Vibrator Calibration
-more difficult than ear phone because a ______cannot be used as the pick up device
-preferred method the “artificial ______procedure” which uses a coupler
Speech Audiometers
- very similar to pure tone checks
-check linearity and level
-match output ______of the oscillator to input ______of the audiometer
Monitoring (VU) Meters
-monitoring meters indicate ______and are found on the face of most audiometers
-calibrated relative to the input it is monitoring (not an ______value)
-on a speech audiometer it is used to monitor the speech signal and lets audiologist adjust the input ______tone for recorded speech materials
Sound Field Testing
-check any ______“hum”
-check internal ______of loudspeaker
Calibration of Ancillary Equipment
-masking generator
- there are standard definitions for white ______, weighted random noise for masking of speech, and ______noise for masking of pure tones.
- it is important to first do an electronic check of the ______since they vary greatly. This verifies that any variation in the bandwidth is due to the transducer rather than the ______itself.
-Compact disc and tape players
- Follow instruction manual that comes with the specific piece of equipment for general maintenance
- Important to calibrate these as recorded speech testing is an integral part of audiometric testing
- Other equipment that requires calibration:
- Auditory ______Response Units (ABR)
- ______Emission Devices (OAEs)
-Acoustic ______Devices
- Measure frequency activator
- Be sure air tube is connected properly
- Check the reflex activating system
Test Room Standards
-Ambient______must be below a standard level ( referred to as MPANLs or Maximum Permissible Ambient Noise Levels)
-Different for _____ conduction testing than bone conduction or sound field testing
-check the room using a ______meter to verify that this standard is met
-important to check testing booths as well since fans, transducer hum, etc can affect testing