Advanced Radiotherapy Consulting Mammosite

-  Mammosite is a catheter balloon used as a HDR brachy breast applicator

-  Rx Dose = 10 fx at 340 cGy/fx for a total dose of 34 Gy delivered BID

-  Rx Point is 1 cm from balloon

-  Used for early stage breast cancer (preferably node negative)

-  Protocol allows for 3 (+) nodes

-  Physics Notes

o  Total Dose = 34 Gy

o  Symmetry of Balloon 2mm asymmetric

o  Min distance to skin is (should be 7mm; shall be 5mm)

o  PTV_Eval 95% Volume covering 90% Dose

o  Max skin dose: 145%

o  Tissue Volume V150 50cc

o  Tissue Volume V200 10cc

o  Uninvolved Normal Breast: 60% receives 50% Rx Dose

-  Contouring Notes

o  Subtract balloon volume from breast volume

o  PTV_eval = PTV – 5mm from skin – pectoral muscle

o  Breast Volume based on external tangential fields

o  Evaluate Air Pockets vs Volume

Advanced Radiotherapy Consulting 3D PBI

-  3D Partial Breast Irradiation used in leau of Mammosite

-  Rx Dose = 10 fx at 385 cGy/fx for a total dose of 38.5 Gy delivered BID

-  Rx Point Volume is PTV_eval

-  Used for early stage breast cancer (preferably node negative)

-  Protocol allows for 3 (+) nodes

-  Physics Notes

o  Total Dose = 38.5 Gy

o  PTV_Eval 90% Volume covering 90% Dose

o  Max Dose in Plan: 120%

o  Heart Dose: 5%V 5% Dose (right sided lesions)

o  Heart Dose: 40%V 5% Dose (left sided lesions)

o  Thyroid Max: 3%

o  Ipsilateral Lung Dose: 15%V 30% Rx Dose

o  Contralateral Lung Dose: 15%V 5% Rx Dose

o  Whole Breast Reference Volume

§  60%V at 50% Rx Dose

§  35%V at Rx Dose

-  Contouring Notes

o  Subtract balloon volume from breast volume

o  PTV_eval = PTV – 5mm from skin – pectoral muscle

o  Breast Volume based on external tangential fields