Laboratory Capacity Building through the Use of Metrologically Traceable Reference Values in Proficiency Testing Programmes

Robert Kaarls, Lindsey Mackay, Angela Samuel, Della Wai-mei Sin, Chuen-shing Mok, Yee-lok Wong, Yiu-chung Yip

E-mail:

Table S1ANOVA table for between-bottle homogeneity study of Iron

Source of variation / SS / df / MS / F / p-value / Fcrit
Between Bottles / 171.7789 / 9 / 19.0865 / 1.2518 / 0.3637 / 3.0204
Within Bottles / 152.4726 / 10 / 15.2473
Total / 324.2515 / 19

Note: SS is the sum of squares; df is the degrees of freedom; MS is the mean square.

Table S2ANOVA table for between-bottle homogeneity study of Zinc

Source of variation / SS / df / MS / F / p-value / Fcrit
Between Bottles / 32.7904 / 9 / 3.6434 / 1.3984 / 0.3034 / 3.0204
Within Bottles / 26.0541 / 10 / 2.6054
Total / 58.8444 / 19

Note: SS is the sum of squares; df is the degrees of freedom; MS is the mean square.

Table S3ANOVA table for between-bottle homogeneity study of Arsenic (total)

Source of variation / SS / df / MS / F / p-value / Fcrit
Between Bottles / 8.9783 / 9 / 0.9976 / 1.2246 / 0.3762 / 3.0204
Within Bottles / 8.1466 / 10 / 0.8147
Total / 17.1249 / 19

Note: SS is the sum of squares; df is the degrees of freedom; MS is the mean square.

Table S4ANOVA table for between-bottle homogeneity study of Cadmium

Source of variation / SS / df / MS / F / p-value / Fcrit
Between Bottles / 0.00067 / 9 / 0.000075 / 2.0799 / 0.1347 / 3.0204
Within Bottles / 0.00036 / 10 / 0.000036
Total / 0.00103 / 19

Note: SS is the sum of squares; df is the degrees of freedom; MS is the mean square.

Table S5Summary of F values and relative standard uncertainty due to between-bottle in(homogeneity)

Analyte / ANOVA test / Relative standard uncertainty due to between-bottle in(homogeneity), ubb (%)
F / Fcrit
Fe / 1.2518 / 3.0204 / 1.05
Zn / 1.3984 / 3.0204 / 1.26
As / 1.2246 / 3.0204 / 0.95
Cd / 2.0799 / 3.0204 / 1.21

Table S6 Results of the stability tests at 20 ºC, 40 ºC and 50 ºC for Iron

Iron
Statistical parameter / 20 ºC / 40 ºC / 50 ºC
Slope of the regression line (b1) / -0.0107 / -0.3878 / -0.4327
Uncertainty associated with the slope [s(b1)] / 0.0024 / 0.3452 / 0.6426
Student’s t-factor
(t0.95, n-2) / 12.71 / 4.30 / 4.30
t0.95, n-2 × s(b1) / 0.0310 / 1.4844 / 2.7633
b1t0.95, n-2 × s(b1) / Yes / Yes / Yes
Statistical significance
(95% confidence interval) / No / No / No
Conclusion / The slope was insignificant.
No instability was observed. / The slope was insignificant.
No instability was observed. / The slope was insignificant.
No instability was observed.

Table S7 Results of the stability tests at 20 ºC, 40 ºC and 50 ºC for Zinc

Zinc
Statistical parameter / 20 ºC / 40 ºC / 50 ºC
Slope of the regression line (b1) / 0.0085 / 0.2035 / 0.1061
Uncertainty associated with the slope [s(b1)] / 0.0149 / 0.3014 / 0.0511
Student’s t-factor
(t0.95, n-2) / 12.71 / 4.30 / 4.30
t0.95, n-2 × s(b1) / 0.1889 / 1.2962 / 0.2196
b1t0.95, n-2 × s(b1) / Yes / Yes / Yes
Statistical significance
(95% confidence interval) / No / No / No
Conclusion / The slope was insignificant.
No instability was observed. / The slope was insignificant.
No instability was observed. / The slope was insignificant.
No instability was observed.

Table S8 Results of the stability tests at 20 ºC, 40 ºC and 50 ºC for Arsenic (total)

Arsenic (total)
Statistical parameter / 20 ºC / 40 ºC / 50 ºC
Slope of the regression line (b1) / -0.0163 / 0.1281 / -0.0111
Uncertainty associated with the slope [s(b1)] / 0.0073 / 0.1973 / 0.1134
Student’s t-factor
(t0.95, n-2) / 12.71 / 4.30 / 4.30
t0.95, n-2 × s(b1) / 0.0926 / 0.8482 / 0.4874
b1t0.95, n-2 × s(b1) / Yes / Yes / Yes
Statistical significance
(95% confidence interval) / No / No / No
Conclusion / The slope was insignificant.
No instability was observed. / The slope was insignificant.
No instability was observed. / The slope was insignificant.
No instability was observed.

Table S9 Results of the stability tests at 20 ºC, 40 ºC and 50 ºC for Cadmium

Cadmium
Statistical parameter / 20 ºC / 40 ºC / 50 ºC
Slope of the regression line (b1) / 0.00007 / 0.0008 / -0.0006
Uncertainty associated with the slope [s(b1)] / 0.00002 / 0.0015 / 0.0011
Student’s t-factor
(t0.95, n-2) / 12.71 / 4.30 / 4.30
t0.95, n-2 × s(b1) / 0.00026 / 0.0064 / 0.0047
b1t0.95, n-2 × s(b1) / Yes / Yes / Yes
Statistical significance
(95% confidence interval) / No / No / No
Conclusion / The slope was insignificant.
No instability was observed. / The slope was insignificant.
No instability was observed. / The slope was insignificant.
No instability was observed.

Table S10Summary of p-values for Iron, Zinc, Arsenic (total) and Cadmium

Analyte / p-value for significance test for 1
20 ºC / 40 ºC / 50 ºC
Fe / 0.1422 / 0.3781 / 0.5701
Zn / 0.6685 / 0.5691 / 0.1733
As / 0.2674 / 0.5826 / 0.9310
Cd / 0.1729 / 0.6490 / 0.6399

Table S11Technical Information - Methods of Analysis of Iron for APMP PT 11-01

Lab Code / Digestion technique / Digestion medium / Matrix separation / Quantification / Source(s) of Calibration standard(s)
1 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Merck CertiPUR Reference Material
2 / *** / *** / *** / *** / ***
3 / *** / *** / *** / *** / ***
4 / Microwave-assisted digestion / HNO3 / Yes / External calibration / ISTD & ESTD
5 / Dry ashing / HNO3/ HCl / No / External calibration / Merck
6 / Microwave-assisted digestion / HNO3 / No / External calibration / Fe in Fe(NO3)2 0.5M(Merck)
7 / Microwave-assisted digestion / HNO3 / No / Internal calibration / Wako
8 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Standard solution Merck® 1000 mg/L
9 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / NIST - SRM3126a
10 / Wet digestion / HNO3/ HClO4 / No / External calibration / Traceable to NIST Standard
11 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Fluka & Spectrasol
12 / Dry ashing / HNO3 / --- / External calibration / Baker Instra-Analyzed reagent 1000 mg/L
13 / Dry ashing / --- / No / External calibration / Fluka,
1000ppm stock solution, Fe in 2% nitric acid
14 / Dry ashing / HCl / No / External calibration / Fluka Iron standard for AAS (1000 mg/L)

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S11Technical Information - Methods of Analysis of Iron for APMP PT 11-01 (Cont’d)

Lab Code / Use of internal standard(s) / Analytical instrument(s) / Correction for recovery / Method accreditation / Additional information
1 / No / Flame AAS / Yes: 89.8% / No / ---
2 / *** / *** / *** / *** / ***
3 / *** / *** / *** / *** / ***
4 / Yes: Sc / ICP-MS / Yes: 99.27% / Yes / ---
5 / No / Flame AAS / No / No / ---
6 / No / Flame AAS / No / No / ---
7 / Yes: DORM-3 / Flame AAS / No / No / Analytical Method: AOAC Official Method 999.10:2002
8 / No / ICP-AES / Yes: 95% / No / ---
9 / No / Flame AAS / Yes: 103.4% / No / Is attached
10 / No / Flame AAS / No / No / ---
11 / No / Flame AAS / No / No / ---
12 / No / Flame AAS / No / Yes / ---
13 / No / Flame AAS / No / Yes / ---
14 / No / Flame AAS / No / No / ---

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S12Technical Information - Methods of Analysis of Zincfor APMP PT 11-01

Lab Code / Digestion technique / Digestion medium / Matrix separation / Quantification / Source(s) of Calibration standard(s)
1 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Merck CertiPUR Reference Material
2 / *** / *** / *** / *** / ***
3 / *** / *** / *** / *** / ***
4 / Microwave-assisted digestion / HNO3 / Yes / External calibration / ISTD & ESTD
5 / Dry ashing / HNO3/ HCl / No / External calibration / Merck
6 / Microwave-assisted digestion / HNO3 / No / External calibration / Zn in Zn(NO3)2 0.5M(Merck)
7 / Microwave-assisted digestion / HNO3 / No / Internal calibration / Wako
8 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Standard solution Merck® 1000 mg/L
9 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / NIST - SRM3168a
10 / Wet digestion / HNO3/ HClO4 / No / External calibration / Traceable to NIST Standard
11 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Fluka & Spectrasol
12 / Dry ashing / HNO3 / --- / External calibration / Fluka TraceCERT
1000 mg/L
13 / Dry ashing / --- / No / External calibration / Fluka,
1000ppm stock solution, Zn in 2% nitric acid
14 / Dry ashing / HCl / No / External calibration / Fluka Zinc standard for AAS (1000 mg/L)

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S12Technical Information - Methods of Analysis of Zincfor APMP PT 11-01 (Cont’d)

Lab Code / Use of internal standard(s) / Analytical instrument(s) / Correction for recovery / Method accreditation / Additional information
1 / No / Flame AAS / Yes: 92.2% / No / ---
2 / *** / *** / *** / *** / ***
3 / *** / *** / *** / *** / ***
4 / Yes: Sc / ICP-MS / Yes: 97.3% / --- / ---
5 / No / Flame AAS / No / No / ---
6 / No / Flame AAS / No / Yes / ---
7 / Yes: DORM-3 / Graphite AAS / No / No / Analytical Method: AOAC Official Method 999.10:2002
8 / No / ICP-AES / Yes: 95% / No / ---
9 / No / Flame AAS / Yes: 100.7% / No / Is attached
10 / No / Flame AAS / No / No / ---
11 / No / Flame AAS / No / No / ---
12 / No / Flame AAS / No / Yes / ---
13 / No / Flame AAS / No / No / ---
14 / No / Flame AAS / No / No / ---

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S13Technical Information - Methods of Analysis of Arsenic (total)for APMP PT 11-01

Lab Code / Digestion technique / Digestion medium / Matrix separation / Quantification / Source(s) of Calibration standard(s)
1 / --- / --- / --- / --- / ---
2 / *** / *** / *** / *** / ***
3 / *** / *** / *** / *** / ***
4 / Microwave-assisted digestion / HNO3 / Yes / External calibration / ISTD & ESTD
5 / Dry ashing / HNO3/ HCl / No / External calibration / Fluka
6 / Microwave-assisted digestion / HNO3 / No / External calibration / As in H3AsO4 0.5M(Merck)
7 / --- / --- / --- / --- / ---
8 / --- / --- / --- / --- / ---
9 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / NIST - SRM83d
10 / Wet digestion / HNO3/ HClO4 / No / External calibration / Traceable to NIST Standard
11 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Fluka & Spectrasol
12 / --- / --- / --- / --- / ---
13 / Microwave-assisted digestion / HNO3/ H2O2 / --- / External calibration / ---
14 / --- / --- / --- / --- / ---

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S13Technical Information - Methods of Analysis of Arsenic (total) for APMP PT 11-01 (Cont’d)

Lab Code / Use of internal standard(s) / Analytical instrument(s) / Correction for recovery / Method accreditation / Additional information
1 / --- / --- / --- / --- / ---
2 / *** / *** / *** / *** / ***
3 / *** / *** / *** / *** / ***
4 / Yes: Rh / ICP-MS / Yes: 97.8% / Yes / ---
5 / No / Graphite AAS / No / No / ---
6 / No / ICP-MS / No / Yes / ---
7 / --- / --- / --- / --- / ---
8 / --- / --- / --- / --- / ---
9 / No / Hydride generation AAS / Yes: 94.3% / No / Is attached
10 / No / Hydride generation AAS / No / No / ---
11 / No / Hydride generation AAS / No / Yes / ---
12 / --- / --- / --- / --- / ---
13 / No / Graphite AAS / No / No / ---
14 / --- / --- / --- / --- / ---

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S14Technical Information - Methods of Analysis of Cadmiumfor APMP PT 11-01

Lab Code / Digestion technique / Digestion medium / Matrix separation / Quantification / Source(s) of Calibration standard(s)
1 / --- / --- / --- / --- / ---
2 / *** / *** / *** / *** / ***
3 / *** / *** / *** / *** / ***
4 / Microwave-assisted digestion / HNO3 / Yes / External calibration / ISTD & ESTD
5 / Dry ashing / HNO3/ HCl / No / External calibration / Merck
6 / Microwave-assisted digestion / HNO3 / No / External calibration / Cd in Cd(NO3)2 0.5M(Merck)
7 / Microwave-assisted digestion / HNO3 / No / Internal calibration / Wako
8 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Standard solution Merck® 1000 mg/L
9 / Microwave-assisted digestion / HNO3/ H2O2 / No / Standard additions / NIST - SRM3108
10 / Wet digestion / HNO3/ HClO4 / --- / External calibration / Traceable to NIST Standard
11 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Fluka & Spectrasol
12 / --- / --- / --- / --- / ---
13 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Fluka,
1000ppm stock solution, Cd in 2% nitric acid
14 / Dry ashing / HCl / No / External calibration / Fluka
Cadmium standard for AAS (1000 mg/L)

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S14Technical Information - Methods of Analysis of Cadmium for APMP PT 11-01 (Cont’d)

Lab Code / Use of internal standard(s) / Analytical instrument(s) / Correction for recovery / Method accreditation / Additional information
1 / --- / --- / --- / --- / ---
2 / *** / *** / *** / *** / ***
3 / *** / *** / *** / *** / ***
4 / Yes: Rh / ICP-MS / Yes: 101% / Yes / ---
5 / No / Flame AAS / No / No / ---
6 / No / ICP-MS / No / Yes / ---
7 / Yes: DORM-3 / Graphite AAS / No / No / Analytical Method: AOAC Official Method 999.10:2002
8 / No / ICP-AES / Yes: 95% / No / ---
9 / No / Graphite AAS / Yes: 99.9% / No / Is attached
10 / No / Flame AAS / No / No / ---
11 / No / Graphite AAS / No / Yes / ---
12 / --- / --- / --- / --- / ---
13 / No / Graphite AAS / No / No / ---
14 / No / Flame AAS / No / No / ---

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S15Technical Information - Methods of Analysis of Ironfor APLAC T082

Lab Code / Digestion technique / Digestion medium / Matrix separation / Quantification / Source(s) of Calibration standard(s)
1 / Microwave-assisted digestion / HNO3/ H2O2/ Aqua regia / No / External calibration / Analytica Praha,
CzechRepublic
2 / Microwave-assisted digestion / HNO3/ H2O2/ Aqua regia / No / External calibration / Merck
3 / --- / --- / --- / --- / ---
4 / Wet digestion / HNO3/ H2O2 / No / External calibration / Std Fe: Merck 1000±2mg/L
5 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / NIM
6 / Microwave-assisted digestion / HNO3/ H2O2 / --- / External calibration / Agilent Technologies
7 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / National Institute of Metrology P.R. China
8 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / ---
9 / Wet digestion / HNO3 / No / External calibration / High Purity Standards
10 / Wet digestion / Aqua regia / No / External calibration / ACR standards
11 / Microwave-assisted digestion / HNO3/ HCl/ H2O2 / No / External calibration / CPA Chem
12 / Wet digestion / HNO3/ HCl / No / External calibration / Choice Analytical
13 / Wet digestion / HNO3/ HCl/ H2O2 / No / External calibration / NIST SRM 3126a
14 / Wet digestion
(Method 3050B) / HNO3/ HCl/ H2O2 / No (All samples were filtered) / Internal calibration / Inorganic Ventures
15 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / SCP Science
16 / Wet digestion / Aqua regia / --- / External calibration / Delta Scientific (Supplier)
17 / Microwave-assisted digestion / HNO3 / No / Internal calibration / Spex Certiprep
18 / Acid digestion / HNO3/ HCl/ H2O2 / No / External calibration / Inorganic Ventures
19 / Acid digestion / HNO3/ HCl/ H2O2 / No / External calibration / Inorganic Ventures
20 / Wet digestion / HNO3 / No / Internal calibration / Inorganic Ventures

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S15Technical Information - Methods of Analysis of Ironfor APLAC T082 (Cont’d)

Lab Code / Digestion technique / Digestion medium / Matrix separation / Quantification / Source(s) of Calibration standard(s)
21 / Wet digestion / HNO3/ HCl/ H2O2 / Yes / Internal calibration / Certified mixes from
Inorganic Ventures
22 / Microwave-assisted digestion / HNO3/ H2O2 / No / Internal calibration / SCP Science
23 / Dry ashing / HCl / No / Standard additions / 2ppm; 5ppm; 10ppm;
15ppm, 20ppm
24 / Microwave-assisted digestion / HNO3/ H2O2 / No / Standard additions / Merck
25 / Microwave-assisted digestion / HNO3/ HCl/ H2O2 / No / External calibration / Merck
26 / Dry ashing / HNO3/ HCl / No / External calibration / Merck
27 / Microwave-assisted digestion / HNO3/ HCl / No / Internal calibration / MRC's
28 / Wet digestion / HNO3/ HCl / No / External calibration / Merck; Batch: HC818177;
Exp: 08/31/2011; Concentration: 999±2 mg/L
29 / Wet digestion / HNO3/ HClO4 / --- / --- / ---
30 / Dry ashing / Aqua regia / No / External calibration / Merck standard solution
31 / Wet digestion / HNO3 / Yes / Internal calibration / Four standard solution 1000ppm
32 / Wet digestion / HNO3 / Yes / Standard additions / Inorganic Ventures
CCS 4,5,6
33 / Wet digestion / HNO3/ H2SO4/HClO4 / No / Internal calibration / Iron standard solution (JCSS)
34 / Microwave-assisted digestion / HNO3/ H2O2 / Yes / External calibration / Merck
35 / Microwave-assisted digestion / HNO3 / No / External calibration / Merck ICP Multi Element Standard Solution
36 / Microwave-assisted digestion / HNO3 / --- / External calibration / High Purity
37 / *** / *** / *** / *** / ***
38 / Microwave-assisted digestion in MARS / HNO3/ H2O2 (5ml and 2ml accordingly) / No / External calibration / Calibration solutions were prepared by dilution of VNIIM Standard Material - aqueous solution of Fe ions: SM 07.01.012

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S15Technical Information - Methods of Analysis of Ironfor APLAC T082 (Cont’d)

Lab Code / Digestion technique / Digestion medium / Matrix separation / Quantification / Source(s) of Calibration standard(s)
39 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Merck certified
40 / Dry ashing / HNO3 / No / External calibration / Merck ICP-MS
Mixed metal standard
41 / Wet digestion / HNO3/ HCl / Yes (filtration post digestion) / External calibration / Merck Iron ICP Standard 10000ppm
42 / Wet digestion / HNO3/ HCl / No / Internal calibration / InorganicVentures
43 / Wet digestion / HNO3, Trace HF / No / External calibration / High Purity Standards
44 / --- / --- / --- / --- / ---
45 / Dry ashing / HCl / Yes / Internal calibration / ---
46 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / Fluka & Spectrasol
47 / Microwave-assisted digestion / HNO3/ H2O2 / --- / External calibration / High Purity Standard
48 / Wet digestion / HNO3/ HCl / No / Internal calibration / GB Standard
49 / --- / --- / --- / --- / ---
50 / --- / --- / --- / --- / ---
51 / Microwave-assisted digestion / HNO3/ H2O2 / No / Internal calibration / GBW 08616
52 / --- / --- / --- / --- / ---
53 / Microwave-assisted digestion / HNO3 / No / External calibration / GB Grade
54 / Wet digestion / HNO3 / No / Internal calibration / SRM
55 / *** / *** / *** / *** / ***
56 / Microwave-assisted digestion / HNO3/ HCl/ H2O2 / No / External calibration / Mineral source
57 / *** / *** / *** / *** / ***
58 / Dry ashing / HCl / No / External calibration / ---

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S15Technical Information - Methods of Analysis of Ironfor APLAC T082 (Cont’d)

Lab Code / Digestion technique / Digestion medium / Matrix separation / Quantification / Source(s) of Calibration standard(s)
59 / Dry ashing / HNO3/ HCl / No / External calibration / Standard iron 1000 ppm (g/ml) - Merck
60 / Wet digestion / HNO3/ HF / No / External calibration / NIST
61 / Wet digestion / HNO3/ H2O2 / No / External calibration / Merck
62 / Microwave-assisted digestion / HNO3 / No / External calibration / Perkin Elmer
63 / Dry ashing / HNO3 / --- / Internal calibration / Merck
64 / Dry ashing / HNO3 / No / External calibration / ---
65 / Dry ashing / HNO3 / No / External calibration / Dilute from Merck Standard Solution 1000 µg/L
66 / *** / *** / *** / *** / ***
67 / *** / *** / *** / *** / ***
68 / Wet digestion / HNO3/ H2O2 / No / Internal calibration / Traceable to Standard Reference Materials from NIST
69 / Dry ashing / HNO3 / No / External calibration / AccuStandard (Traceable to National Institute of Standards and Technology Reference Materials)
70 / Dry ashing / HNO3/ HCl / No / External calibration / NIST
71 / Microwave-assisted digestion / HNO3/ H2O2 / No / Standard additions / UWR

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S15Technical Information - Methods of Analysis of Ironfor APLAC T082 (Cont’d)

Lab Code / Use of internal standard(s) / Analytical instrument(s) / Correction for recovery / Method accreditation / Additional information
1 / No / Flame AAS / No / Yes / ---
2 / No / Flame AAS / No / Yes / ---
3 / --- / --- / --- / --- / ---
4 / No / Flame AAS / No / Yes / ---
5 / Yes: Rh / ICP-MS / Yes: 100.5% / Yes / ---
6 / Agilent Technologies / ICP-MS / No / Yes / ---
7 / No / ICP-AES / No / No / ---
8 / Yes: Sc / ICP-MS / No / Yes / ---
9 / Yes / ICP-AES / No / Yes / Results on sample as received, not adjusted for moisture.
10 / Te / ICP-AES / No / Yes / ---
11 / Yes: Lu / ICP-AES / No / Yes / ---
12 / Yes: Lutetium / ICP-AES / No / Yes / ---
13 / Yes: Yttrium / ICP-AES / Yes: 86.94% / Yes: AIHA IHPAT / ---
14 / Yes: Indium, Yttrium / ICP-AES / No / Yes / ---
15 / Yes: Indium, Scandium, Yttrium / ICP-MS / No / Yes / ---
16 / No / ICP-AES / No / Yes / ---
17 / Yes: Sc / ICP-MS / No / Yes / ---
18 / No / ICP-MS / Yes: 99.6% / Yes / ---
19 / Yes: Sc, In, Lu / High Res ICPMS / Yes / Yes / ---
20 / Yes: Inorganic Ventures / ICP-MS / No / Yes / ---

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S15Technical Information - Methods of Analysis of Ironfor APLAC T082 (Cont’d)

Lab Code / Use of internal standard(s) / Analytical instrument(s) / Correction for recovery / Method accreditation / Additional information
21 / Yes: Yttrium / ICP-OES / No / Yes / ---
22 / Yes: In, Lu, Re, Rh, Tb, Te, Y, Ge / ICP-MS / No / Yes / ---
23 / No / Flame AAS / No / Yes / ---
24 / No / ICP-OES / Yes: 102% / Yes / ---
25 / No / ICP-AES / No / No / ---
26 / No / ICP-AES / No / Yes / ---
27 / Yes: MRC / Flame AAS / No / No / ---
28 / No / Flame AAS / Yes: 83.0% / No / ---
29 / --- / Flame AAS / Yes: 99.9% / Yes, AOAC 968.08 (2002) / ---
30 / No / Flame AAS / No / Yes, AOAC 999.11 / ---
31 / No / Flame AAS / No / Yes / ---
32 / No / ICP-AES / No / Yes / ---
33 / Yes: Y (Yttrium) / ICP-AES / No / No / ---
34 / No / ICP-OES / No / Yes / ---
35 / No / ICP-AES / No / No / ---
36 / No / ICP-AES / No / Yes / ---
37 / *** / *** / *** / *** / ***
38 / Yes: solution of In115 / ICP-MS / Yes: Moisture content is 13.61% / --- / Bottle#210, sample mass for analysis about 0.5g
39 / No / ICP-AES / No / No / ---
40 / No / Flame AAS / No / Yes / ---

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S15Technical Information - Methods of Analysis of Ironfor APLAC T082 (Cont’d)

Lab Code / Use of internal standard(s) / Analytical instrument(s) / Correction for recovery / Method accreditation / Additional information
41 / No / ICP-AES / No / Yes / Our staff froze the sample as per routine SOP and assumed that this will not affect the sample results.
42 / Yes: Scandium / ICP-MS / No / Yes / ---
43 / Yes / ICP-AES / No / Yes / ---
44 / --- / --- / --- / --- / ---
45 / Yes / Flame AAS / No / No / ---
46 / No / Flame AAS / No / No / ---
47 / Yes / Flame AAS / --- / No / ---
48 / Yes: Indium / ICP-AES / No / No / ---
49 / --- / --- / --- / --- / ---
50 / --- / --- / --- / --- / ---
51 / Yes: Yttrium / ICP-AES / No / No / ---
52 / --- / --- / --- / --- / ---
53 / No / ICP-AES / No / No / ---
54 / Yes: Sc / ICP-AES / No / No / ---
55 / *** / *** / *** / *** / ***
56 / No / Flame AAS / No / No / ---
57 / *** / *** / *** / *** / ***
58 / No / Flame AAS / Yes: 91% / No / ---
59 / No / Flame AAS /
Graphite AAS / No / No / AOAC 999.11:2005
60 / No / ICP-AES / No / Yes / ---

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S15Technical Information - Methods of Analysis of Ironfor APLAC T082 (Cont’d)

Lab Code / Use of internal standard(s) / Analytical instrument(s) / Correction for recovery / Method accreditation / Additional information
61 / No / ICP-AES / No / No / ---
62 / No / ICP-OES / No / No / ---
63 / No / Flame AAS / No / Yes / ---
64 / No / ICP-MS / No / No / ---
65 / No / Flame AAS / Yes: 95.24% / Yes / ---
66 / *** / *** / *** / *** / ***
67 / *** / *** / *** / *** / ***
68 / No / Flame AAS / No / No / ---
69 / No / Flame AAS / No / No / ---
70 / Yes: Cr / ICP-AES / No / Yes / ---
71 / No / Flame AAS / No / Yes / ---

“***” The registered participant did not submit the result. “---” Data or information was not provided.

Table S16Technical Information - Methods of Analysis of Zincfor APLAC T082

Lab Code / Digestion technique / Digestion medium / Matrix separation / Quantification / Source(s) of Calibration standard(s)
1 / Microwave-assisted digestion / HNO3/ H2O2/ Aqua regia / No / External calibration / Analytica Praha,
CzechRepublic
2 / Microwave-assisted digestion / HNO3/ H2O2/ Aqua regia / No / External calibration / Merck
3 / Dry Ashing / Aqua regia / No / Internal calibration / 5
4 / --- / --- / --- / --- / ---
5 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / NIM
6 / Microwave-assisted digestion / HNO3/ H2O2 / --- / External calibration / Agilent Technologies
7 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / National Institute of Metrology P.R. China
8 / --- / --- / --- / --- / ---
9 / Wet digestion / HNO3 / No / External calibration / High Purity Standards
10 / Wet digestion / HNO3 / No / External calibration / ACR standards
11 / Microwave-assisted digestion / HNO3/ HCl/ H2O2 / No / External calibration / CPA Chem
12 / Wet digestion / HNO3/ HCl / No / External calibration / Choice Analytical
13 / Wet digestion / HNO3/ HCl/ H2O2 / No / External calibration / NIST SRM 3168a
14 / Wet digestion
(Method 3050B) / HNO3/ HCl/ H2O2 / All samples were filtered / Internal calibration / Inorganic Ventures
15 / Microwave-assisted digestion / HNO3/ H2O2 / No / External calibration / SCP Science
16 / Wet digestion / Aqua regia / --- / External calibration / Delta Scientfic
17 / Microwave-assisted digestion / HNO3 / No / Internal calibration / Spex Certiprep
18 / Acid digestion / HNO3/ HCl/ H2O2 / No / External calibration / Inorganic Ventures
19 / Acid digestion / HNO3/ HCl/ H2O2 / No / External calibration / Inorganic Ventures
20 / Wet digestion / HNO3 / No / Internal calibration / Inorganic Ventures

“***” The registered participant did not submit the result. “---” Data or information was not provided.