SAMS Questions
Session: Advances in Breast imaging: Digital Breast Tomosynthesis
QC Requirements for the Hologic Dimensions 3D and GE SenoClaire Digital Breast Tomosynthesis Units -Michael S. Glaser, M.S. DABR, Alliance Medical Physics, Alpharetta, Georgia
- The HologicSelenia Dimensions 3D Physicist QC protocol includes which of the following test requirements for artifact evaluation:
- Test in 2D mode + 3D Tomo mode + Dicom printer
- Test in 2D mode only
- Test in 3D Tomo mode only
- Test in 2D mode using Large focal spot only
- Test in 2D mode using Small focal spot only
The correct answer is
A. Test in 2D mode + 3D Tomomode + Dicom printer
Source: HologicSelenia Dimensions QC Manual:MAN-01965, Rev. 008 (July/2014)
- The HologicSelenia Dimensions 3D Technologist QC protocol includes which of the following set of comprehensive test requirements:
- Weekly artifact evaluation + phantom image
- Weekly artifact evaluation + phantom image + semiannual geometry calibration
- Weekly SNR + CNR in the 3D Tomo mode
- Annual Breast entrance exposure in the 3D Tomo mode
- Annual Collimation assessment in the 3D Tomo mode
The correct answer is
B. Weekly artifact evaluation + phantom image + semiannual geometry calibration
Source: HologicSelenia Dimensions QC Manual:MAN-01965, Rev. 008 (July/2014)
- When scoring the ACR mammo phantom on both the HologicSelenia Dimensions 3D Tomo and GE SenoClaire3D Tomo, the minimum passing score for the 3DTomo mode is:
- 5 fibers, 4 speck groups, 4 masses
- 4 fibers, 3 speck groups, 3 masses
- 6 fibers, 5 speck groups, 5 masses
- 3 fibers, 3 speck groups, 3 masses
- 1 fibers, 1 speck groups, 1 masses
The correct answer is
B. 4 fibers, 3 speck groups, 3 masses
Source: HologicSelenia Dimensions QC Manual:MAN-01965, Rev. 008 (July/2014) GE SenoClaire QC Manual: 5415892-4-1-EN, Rev 1, 2013-14
- The Average Glandular Dose ( AGD)per view for both the Hologic Dimensions 3D Tomo unit and the GE SenoClaire 3D Tomo unit must meet the following performance criteria:
- AGD < 1.5mGy for 2D mode mode
- AGD < 1.5mGy for 3D mode mode
- AGD < 3.0 mGy for 2D mode + 3D Combo mode
- AGD < 6.0 mGy for 2D mode + 3D Combo mode
- AGD < 6.0 mGy for 3D mode mode
The correct answer is
C. AGD < 3.0 mGy for 2D mode + 3D mode
Source: HologicSelenia Dimensions QC Manual:MAN-01965, Rev. 008 (July/2014)
GE SenoClaire QC Manual: 5415892-4-1-EN, Rev 1, 2013-14
- The GE SenoClaire 3D TomoPhysicist QC protocol includes which of the following annual test requirements:
- AOP 2D + 3D SNR test only
- Phantom IQ 3D test only
- Flatfield 3D test only
- AOP 2D + 3D SNR + Phantom IQ 3D
- AOP 2D + 3D SNR + Phantom IQ 3D + Flatfield 3D
The correct answer is
E: AOP 2D + 3D SNR + Phantom IQ 3D + Flatfield 3D
Source: GE SenoClaire QC Manual: 5415892-4-1-EN, Rev 1, 2013-14