LAMPIRAN A

SEBUTHARGA BAGI MEMBEKAL, MENGHANTAR, MENGUJI DAN MENTAULIAH PERALATAN PERUBATAN SATU (1) SETBRONCHOSCOPE UNTUK JABATAN PERUBATAN AM, HOSPITAL SIBU, SARAWAK BAGI TAHUN 2013

NO. SEBUTHARGA: HS/Q021/2013

BIL. / SPECIFICATION / QTY / COMPLIANCE TO REQUIREMENTS: YES / NO / COMMENTS / PROPOSED SPECIFICATION
1 / Video Bronchoscope / 1
1.1 / Features :
1.1.1.Detachable suction unit ensures very good hygiene
1.1.2.Ergonomic video connection with sealing cap ensures comfortable handling for the personnel
1.1.3.Optimally matched, non-rotational insertion sheath
1.1.4.Short insertion tip with smooth-running and non-twisting angle mechanism
1.1.5.Sensitive and stable control via ergonomic handle
1.2 / Specification :
1.2.1.Working channel / 2.0 - 2.5 mm
1.2.2.Direction of View / 0 degree
1.2.3.Angle of View / 80 - 120 degree
1.2.4.Deflection / up 90 - 180 degree , down 0- 100 degree
1.2.5.Depth of View / 3 - 50 mm
1.2.6.Working Length / 55 - 65 cm
1.2.7.Outer diameter / 5.0 - 6.5 mm
1.3 / Set Consisting of :
1.3.1.Video-Bronchoscope-Set, PAL, including a mounted suction device
1.3.2.Direction of view: 0° . Angle of view: <120°.
1.3.3.Deflection: up <180°, down <100°
1.3.4.Depth of view: 3-50 mm
1.3.5.Working length: 50-65 cm
1.3.6.Outer diameter: 5-6 mm
1.3.7.Instrument channel: no more than 2.2 mm consisting of:
1.3.7.1.Case / 1
1.3.7.2.Video Connecting Cable / 1
1.3.7.3.Mouth Piece / 1
1.3.7.4.Biopsy Forceps, spoon-shaped, round, flexible, double action jaws / 2
1.3.7.5.Leakage Tester, with bulb and manometer / 1
1.3.7.6.Pressure Compensation Cap, for ventilation during gas sterilization / 1
1.3.7.7.Cleaning Brush / 1
1.3.7.8.Irrigation Tube, for working channel / 1
1.3.7.9.Instrument Oil / 1
1.3.7.10.Protective Film / 1
2 / Adaptor, for Cold-Light- Fountains for use with Video-Endoscope with external Insufflation-Connection / 1
3 / Video Connecting Cable PAL , Videoscope/Image 1 hub, MID/High Res, approx 1/4", 1/6", 1/10", approx length 90-100 cm / 1
4 / Unit must be compatible with existing Karl Storz HDTV camera system in OT and ENT Clinic.
5 / WARRANTY
5.1One Year against Manufacturing Defect from Date Commissioning.
5.2Parts availability for at 10 years from date of testing and commissioning.
5.3Breakdown respond time should be within 48 hours within warranty period.
5.4Vendor to specified the warranty period for the parts and accessories.
6 / SERVICE
6.1Free Preventive Maintenance Service 1 X 6 Monthly During Warranty Period.
6.2Supplies have to provide at least two (2) copies of the Operation Manuals in both soft and hard copy during testing and commissioning.
6.3Supplies has to provide at least two (2) copies of Instructions for use Manuals with trouble Shoot Guide, Article Numbers Catalogue and ordering Information.
7 / TRAINING
Onsite Training By Application Specialist And Factory Trained Personnel On Operation Clinical Applications Running Self-Test And Basic Trouble Shooting.

Page 1 of 3