Element Installation Report

Date: / Report Date
Customer: / Customer
EDAX Order Number S#: / S#
Analyzer Serial Number: / Analyzer s/n
Ardexus Product Number: / Ardexus #

Rev. 1.3

May 31, 2016

Element Installation Report / Page 3 of 4
Customer Information
Customer: / Customer
User: / User Name
Address: / Address
Address: / Address 2
Address: / Address 3
City: / City
State: / State
Zip Code: / Zip code
Country: / Country
Phone: / Phone
System Information
Analyzer Model: / S/N: / Analyzer s/n
Microscope Vendor:
Microscope Model: / S/N:
EDAX Computer Model: / S/N:
Operating System: / Svc Pk:
Printer: / S/N:
Monitor: / S/N:
Order Number (S#): / S#
User Interface / Electrical
Check Montor function: / OKN/AYesNo / Check Mains Supply: / OKN/AYesNo
Check Keyboard function: / OKN/AYesNo / Measured Input Voltage (in VAC):
Check Mouse function: / OKN/AYesNo / Line Frequency: / 50 Hz60 Hz
Check Touchscreen function: / OKN/AYesNo / Check Ground / Earthing: / OKN/AYesNo
Analyzer / Column and/or Stage control? / N/AColumn CtrlStage CtrlColumn + Stage
Check cooling fans in Analyzer: / OKN/AYesNo / Check HD write-cache setting: / OKN/AYesNo
Check Hyperthreading is disabled: / OKN/AYesNo / Check condition of all cables connections: / OKN/AYesNo
Check PCBs inside EDAX Analyzer (seated properly): / OKN/AYesNo
Imaging / Stage and Column Control
Check software & hardware / OKN/AYesNo / Stage Control: / OKN/AYesNo
Check and adjust imaging / OKN/AYesNo / Column Control / OKN/AYesNo
Check mapping / OKN/AYesNo / DCOM Control (Yes/No): / YesNo
Line Measurements, Scan Extent: / Microscope Software Version:
X = / Port Data: / IP Address
Y =
Use SG2 Shell program, print results / (PDF + Printer): / OKN/AYesNo

Login Setup

Check Administrator and User logins with appropriate privileges: / YesNoOKN/A
Check Windows Login: / YesNoOKN/A
Check TEAM Element Software Login: / YesNoOKN/A
Show the user where they can add their company’s logo in TEAM Element: / YesNoOKN/A

EDS Information

Detector Model: / S/N:
Detector Part Number:
EDAX Software (select): / TEAM ElementTEAM Element + Genesis
EDAX Sales Order Number (S#): / S# / Ardexus product Number: Ardexus #

EDS Detector

Adjust mechanical setup of the detector: / YesNoOKN/A
Adjust the hard stops of the detector if applicable: / YesNoOKN/A
Verify the vacuum integrity of the microscope: / YesNoOKN/A
Use eDPP Shell program, calibrate all Amp Times using Al and Cu: / YesNoOKN/A
Use eDPP Shell program, print results (PDF + Printer): / YesNoOKN/A
Record MnKα resolution (eV):
Verify detector cool down: / YesNoOKN/A

Customer Instruction for EDS

Check
EDS Detector handling including startup, window precautions: / YesNoOKN/A
System calibration, geometry parameters: / YesNoOKN/A
TEAM Element overview of Spectrum, Point Analysis, Mapping and Report functions: / YesNoOKN/A
Recording and storing spectra, identifying elements, HPD function, background operations: / YesNoOKN/A
Mapping: Image acquisition, spectral mapping, phase mapping, setting up mapping: / YesNoOKN/A
Qualitative and quantitative Line Scan: / YesNoOKN/A
Processing of stored data sets (Quantitative mapping, RGB): / YesNoOKN/A
Evaluation / printout of spectra, storing of results for analysis report, data storage, report generator: / YesNoOKN/A
Show location and access to software movies: / YesNoOKN/A
Advise customer on installation of remote software: / YesNoOKN/A
Number of hours spent on user training:
Basic Operation Training by Service complete (Yes/No)? / YesNo
Advise customer to store system restoration software: / YesNo
Advise customer to store critical hardware: Tools, boxes, blanking port, etc.: / YesNo

Refer to EDAX web site for additional training: http://www.edax.com/Support/training/index.aspx

Comments / Notes:


Installation Documentation Complete

Installation problems reported to TFS? Data entered in Ardexus?

Warranty Start Date:
Customer: / Customer
User: / User Name
Signature:
Date:
EDAX Representative:
Date:

Contact information

Sales: / Applications:
Phone: / Phone:
E-Mail: / E-mail:
Service:
Phone:
E-Mail:

Print Pages 1 to 4

EDAX Inc. / Ametek B.V. / Ametek Nordic AB
Ametek, Materials Analysis Division / Ringbaan Nord 103 / EDAX Business Unit
91 McKee Drive / P.O. Box 4144 / Krossgatan 36
Mahwah, NJ 07430 / 5004JC Tilburg / SE-162 50 Vällingby
USA / The Netherlands / Sweden
Tel: 201 529 4880 / Tel: +31 13 536000 / Tel: +46(0)8 519 060 31
Fax: 201 529 4760 / Fax: +31 13 535 6279 / Fax: +46(0)8 519 060 34
Mob: +46(0)73 367 0785
Ametek GMBH / EDAX Business Unit / Ametek Inc. Beijing
EDAX Business Unit / Ametek Co. Ltd / Representative Office
Kreuzberger Ring 6 / 3F, Shiba NBF Tower / Room 2202, 22/F Citic Building
D65205 Wiesbaden / 1-30 ShibaDaimon 1-chrome Minato-ku / 19 Jianguimenwai Dajie
Germany / Tokyo 105-0012 Japan / Beijing, 100004, P.R. China
Tel: +49(0)611 97304 50 / Tel: +81 3 6809 2404 / Tel: +86 10 8526 2111-22
Fax: +49(0)611 97304 55 / Fax: +81 3 6809 2410 / Fax: +86 10 8526 21414
Other Region
EDAX Business Unit / EDAX Business Unit / Company name
Ametek Instruments India Pvt., Ltd / Ametek Singapore Pte Ltd. / Company address
601, Raaj Chambers / No. 43 Changi South Avenue 2 / Company Address 2
Off Old Nagardas Road, Andheri (E) / #04-01 Singapore 486164 / Company addrfess 3
Mumbai – 4100 069 / Tel: +65-6484-2388 / Phone number
Tel: +91-22-6196-8200 / Fax: +65-6481-6588 / Fax number
Fax: +91-22-2836-3613
Element Installation Report / Page 3 of 4

Skip this page. This page does not apply!

Detector Performance

Resolution / Check
MnKα resolution / eV / YesNoOKN/A
BKα resolution / eV / YesNoOKN/A
CKα resolution / eV / YesNoOKN/A
NKα resolution / eV / YesNoOKN/A
OKα resolution / eV / YesNoOKN/A
FKα resolution / eV / YesNoOKN/A
Check Peak shift between 20% and 40% Dead Time / eV / YesNoOKN/A
Check low energy efficiency at 20kV using Cu (Net Int. CuLα/CuKα) / YesNoOKN/A

SEC Factor Correction

Element / Sample / HV / Wt% / Wt% / Peak pos. / Exp. SEC / Meas.SEC / Check
BKα / CaB6 / 10kV / B=61.81 / Ca=38.19 / 0.18-4 / YesNoOKN/A
CKα / SiC / 5kV / C=29.95 / Si=70.05 / 0.4-2 / YesNoOKN/A
NKα / CrN / 15kV / N=21.22 / Cr=78.78 / 0.1-1 / YesNoOKN/A
OKα / Fe203 / 15kV / O=30.06 / Fe=69.94 / 0.2-1 / YesNoOKN/A
FKα / CaF2 / 10kV / F=48.66 / Ca=51.33 / 0.2-1 / YesNoOKN/A

Help

Entering Data

The Acceptance report is made with fields that should be filled out electronically. Because of different regional requirements and configurations, some fields will change based on region and configurations. Always start at the top and work down.

If the form does not open and allow editing, select View and then Edit Document first.

Filling out the form automatically populates some fields later in the form, such as Customer and Sales numbers.

Fill in the shaded fields with information. You can use the mouse to move from field-to-field. Do NOT use [Enter].

In many cases, a pull down selection can be made to make data entry easy.

The [DEL] key in a field clears the entry.

Security

This document uses macros to work properly. Be sure to Enable Macros in Trust Center settings.

Select View, then Edit in Word 2013 to allow the document to be edited.

Using pull-down choices

Select OK when a check of a section is completed successfully to acknowledge it is done.

Select Yes to confirm a configuration.

Select No if the check is not done, or not successful.

Select N/A if the check does not apply to the configuration, or can not be done for some reason.

Page 1

Selecting a Region

Be sure to start the report with the correct region where the system is installed. If the region you need is not one of the available selections, you can add a region on page 4 of the form. Then select “Other” region pull down. The Region information will update after a region is selected and the next field is selected (or click on the [Tab] key).

Date of Report

The date field on page 1 should be the date the report is filled out.

Customer, EDAX Order #, Analyzer serial number

These are needed to identify the system.

The customer name should be the name of the Company where the system is installed.

The EDAX Order # (Sales Number or S#), can be found on the Shipping paper work, or in the “About” window from the TEAM or Genesis software.

***The Ardexus Product number can be obtained from the local Sales/Service office.

The Analyzer serial number can be found on an EDAX label on the computer workstation.

Page 2

Customer Information – Fill out as required. If all of the address fields are not needed, enter a [-] in the empty field. The Customer field will be automatically carried over from Page 1.

System information – Enter the information as available.

The Analyzer Model in many cases will be the EDAX Computer, such as an “X1” or “HPZ230”. The Analyzer serial number will be carried over from the first page.

Enter the rest of the information as configured. If a printer is not supplied, enter “Not supplied” or N/A.

The EDAX Order Number will be carried over from Page 1.

Hyperthreading

To check if Hyperthreading is enabled

Type wmic in Windows run box

Type cpu Get NumberofCores,NumberofLogicalProcessors /Format:List n the window

If the number of Logical Processors > number of Cores, then Hyperthreading is enabled

Type exit to quit

Reboot and change Hyperthreading settings in the BIOS

Hard Disk write-cache buffer flushing: it is recommended to turn off the Windows write-cache buffer flushing option. Leaving it on can slow down the system response times. This setting is in the Windows Device Manager – Disk Drives – Policies tab

Electrical

Check the Mains Supply. Measure the actual input voltage where the EDAX system will plug into.

Record the Line Frequency in Hz where the EDAX system will plug into.

Check the Ground/Earthing where the EDAX system will plug into. Request information from the customer or building operations as needed. Select the appropriate response from the pull down selections.

Imaging – Set up the imaging for the microscope model per the specific microscope interface documentation. Use the service Shell to configure the imaging interface parameters.

Print the parameters from the Shell to a PDF and printer, if available. Select the response from the pull down selections.

Stage and Column Control – Select whether the system is configured with stage and/or column control software per the order. Add any specific notes as desired, such as IP address, or Com port used to the microscope control notes field.

Login Setup – This section confirms that the User has been shown how to log in to Windows and the EDAX software.

Page 3

EDS Information – List the actual information from the label on the detector.

Select the software installed on the system and version information.

EDAX Sales and Ardexus numbers are automatically filled in from previous entries.

EDS Detector – The detector should be installed and calibrated according to installation procedures in the service manual (on disk in c:\EDAX Service\Service Manuals) and the calibration Help files in the Shell program. Check each of the requested functions during installation and setup.

Note the data requested in each of the fields.

Mechanical stops (for detectors with a slide) should be set so the analyze position matches geometry on included data sheets, without touching any part in microscope.

The vacuum level of the microscope should be checked against the vacuum level before the detector was installed, per the installation procedure in the service manual.

The detector should be calibrated in the Shell program per the calibration procedure appropriate for the detector type.

The best resolution should be recorded. Note that the resolution is typically tested at a count rate of 10 kcps, but may vary per the individual detector. See the factory data sheets shipped with the detector for the actual test conditions used by the factory. If unable to achieve the resolution specifications, test using the same count rate used in the factory tests. The best resolution value may be at any Amp Time, any eV/Ch.

Customer Instructions for EDS

The customer should be provided with an overview during the installation so they can operate the system safety and collect basic analyses.

Comments / Notes – A free form area to add notes about the system that should be recorded.

Page 4

Installation documents – If any quality problems occurred during the installation, such as missing or damaged parts, they should be reported via the TFS system. See the local service manager for access.

Enter the Warranty start date.

The Customer and User names will be carried over from previous entries.

There is a line for a customer’s physical signature, if the form is printed.

EDAX Representative is the installing service engineer’s name and the date.

Contact information

Enter the information for the customer with the regional EDAX contacts for service, sales and applications.

Printing – The report is designed to be filled out electronically.

Under the Print menu option, set the pages to print, select the Page per sheet option, and select the Paper Size appropriate for your region. Select Scale to Paper Size for correct formatting.