NN / name / quantity1 / Centrifuge
BENCHTOP CENTRIFUGE -1
Performance
• max.RCF:4,226
• max.capacity:4x100ml
. max.32x15ml
. max.8 slide carriers for cyto preparations
Design
• metal housing and lid
• stainless steel centrifuging chamber
• practical lid locking device
• viewing port in the lid
Safety
• lid locking and holding
• imbalance switch-off
• automatic rotor recognition
• emergency lid lock release
Display and controls
• foil keypad
. fast,precise and easy entry of the
parameters via selector keys
• speed RPM
.entry in increments of 100
• time t/min
.entry in minutes,max.99 min.
• impulse key
.for short centrifuging
•RCF key
.Indication of the RCF during centrifugation,as long
as the key is pressed.This requires the previous entry of the rotor radius. / 1
2 / Noiseless compressor for chromatograph activation
ZERO AIR GAS GENERATOR
Model Name ZA Range
Gases Produced:Zero Air
Applications:GC,GC-ECD,GC-FID,GC-FPD,THA
The ZA Range of Zero Air Generators is designed specifically for use as
Flame gas for GC Applications,which require zero air for operation.Units
Are available with or without internal air compressors and some models are
further available as 19 Rack Versions.
Technical Specification
Max Flow Rate (Outlet) for specified CH4 Concentration
1,500cc/min-30,000cc/minATP
Outlet CH4 Concentration (as Methane):
<0.1ppm
MaxInlet CH4 Concentration (as Methane):
100ppm
Particies >0.01micron:
None
Electrical Requirements:
ZA015-ZA180-110v 60Hz 4.0A/220v 50Hz 2.0A
ZA300-110v 60Hz 4.5A/220v 50Hz 8.9A
Electrical Requirements with internal air compressor:
ZA035A-ZA180A-110v 60Hz 7.0A/220v 50Hz 305A
Dimensions:
ZA015-ZA180-25x16x51cm/17x18x35 ins
ZA300-43x41x88cm/17x16x35 ins
Dimensions with internal air compressor
ZA035A-ZA180A-43x41x88cm/17x18x35 ins
Dimensions 19 Rack:
ZA015-ZA180-18x48x43cm/7x19x17 ins
Service Plan Notes
Complete and Standard Service Plans available for all models / 1
3 / Distiller (distillation apparatus) with a receiver
AUTOMATIC WATER STILL
Capacity 8.0 L/Hour
Dimension(WxDxHmm) 680x360x650
Heater 6.0KW
Cooling Water Approx.60-80l
Material Inner Stainless Steel Boiler
Stainless Steel Storage Tank
Outer Powder Coating Steel Case
Reservoir 16-20l SUS Srorage Tank
Safely Device Over Temp Protector
Water Float Level Sensor
Water Supply Cut Off
Operation Fully Automatic Operation
Signal lamp Distill,Full,Water
Electric supply 220V,50Hz / 1
4. / Redistiller
AUTOMATIC DOUBLE WATER STILL
Capacity 4.0L/Hour
Dimension (WxDxHmm) 570x600x600
Heater 9.0KW
Cooling Water Approx.35-60l
Material Inner Stainless Steel Boiler
Stainless Steel Storage Tank
Outer Powder Coating Steel Case
Reservoir 11L SUS Storage Tank 16-20l SUS Storage Tank
Safely Device Over Temp Protector
Water Float Level Sensor
Water Supply Cut Off
Operation Fully Automatic Operation
Signal lamp Distill,Full,Water
Electric supply 380V,50/60Hz,3Phase / 1
5 / Drying cupboard for chemical ware
UNIVERSAL DRYING OVEN
-Rapid Heat up 25 min up to 250ie
-Digital Timer built –in Controller provide a efficient and convenient use and operation
Dimension Inner 400x350x400
(WxDxHmm) Outer 545x500x770
Capacity 56 l
Heater 1.5Kw
TemperatureRange Ambient +5ºC to 250°C
Accuracy ±1.0°C
Uniformity ±3.0°C at 120°C
Convection Natural Convection
Controller PID Multi –Funcion ,Alarm,Auto tunning
Timer Automatic timer 0-99 min,99hr,continuous
Material Inner Stainless Steel(SUS 304)
Outer Steel with epoxy powder coating
Shelves Adjustable 2 or 3 shelves
Insulation Glass Wool
Electric supply 110V,60Hz or 220V,50/60Hz with solid schuko plug / 1
6 / Rotary evaporator for samples concentration
Rotary Evaporator
Condenser Vertical
Capacity 4.3L
Flasks Range: 50ml to 3000ml
Speed control 20~280RPM / 1
7 / Automated extractors (from solid samples)
Extraction systems with linear sample configuration
Extraction unit with 4 in-line sample positions,250ml
Dimensions(WxHxD) approx.53x74x32cm
(unit w/o glassware)
Weight 15,1 kg
(unit w/o glassware)
Rated voltage 230VAC,50/60Hz
Power 1440W / 4
8 / Mufflefurnace (minimum 600° C; volume: 5-7 L)
Electric Muffle Furnace
Controller Microprocessor
Dimensions Inner 150x300x100
(W xD xHmm) Dimensions Outer 420x530x510
Chamber Capacity 4.5L
Watt 3000w
Temperature Up to 1200C° Max
Output Regulator SSR unit control
Thermocouple K type CA Sensor
Heating Element Kanthal
Insulation Vacuum Molded Ceramic Fiber
Material Stainless steel
Electric Supply 220V 1Phase or 3 Phase 50/60Hz / 1
9 / Shaker for flasks (volume: up to 0.5 L)
Orbitalshakers
Shaking motion circular
Orbital diameter 10mm
Perm.shaking weight with attachment 7,5kg
Motor rating input/output 45/10W
Permissible ON-time 100%
Speed range 20-500 rpm
Speed display LED-Line digital
Timer /5-50 min /0-9h 59 min
General information
Dimensions WxDxH 360x420x98mm
Weight 8,1kg 8,8kg
Permissible ambient temperature 5-50°C
Protection class according to DIN EN 60529 IP21
Interface RS232/analog / 1
10 / Laboratory grinder for solid samples
Mill
Grinding procedures Batch
Grinding head
Motor rating input/output 300/160 W
Speed range 28.000 rpm
Useful volume 80ml
Duty cycle on/off
Circumferential speed 53 m/sec
Granularity of the task 10mm / 1
11 / Sieving stack to sieve the samples ( 0.05-1.5 mm)
Analytical Sieve Shakers
Analytical sieve shakers are used in research and development,qualitycontrol of raw materials,interim and finished products as well as in production monitoring.
-Sieving with 3D effect.
-For wet and dry sieving.
-Short sieving times.
-Low-noise,maintenance-free. / 1
12 / Analytical balance with a set of weights
Analytical Balance GLP /GMP/ISO 9000
Metal housing
Max.weighing range-220 g
Min weighing range-0,1mg
Read out-0.1mg
RS-232Сinterface / 1
13 / Laboratory electronic balance with a set of weights
Лабораторные электронные весы.
GLP/GMP/ISO9000
Max.weighing range-820 g
Min.weighing range10,0 mg
Read out -10,0 mg
Internal adjusting CAL
RS-232C interface / 1
14 / Refractometer
DIGITAL HAND-HELD“POCKET”REFRACTOMETERS
Feature
• Calibration with water only
• Extremely water resistant (IP65)
• Light&Compact,100g
• Revolutionary new feature ELI*
• Automatic Temperature Compensation(A.T.C.)
• Measurement in 3 seconds
• Ergonomically designed for one-handed operation
• Equipped with a strap holder option
• Convenient storagecase
• Brix(0.0~53.0%) / 1
15 / Hydrogen generator for chromatograph activation
Hydrogen Generator
Model Name:PH Range
Gases Produced:Hydrogen
Applications:GC,GC-FID,GC-FPD,GC-NPD,GC-TCD
The PH Hydrogen Generators are designed to offer a pure
Source of hydrogen for your GC
Application,whether you use hydrogen for FID or Carrier Gas.
Technical Specification
Hydrogen: 200-600cc/min
Purity: 99.9995%
Dewpoint: -40°C/-40°F
Delivery Pressure: 0-6.9 bar /0-100ps
Gas Flow Indicator : Yes
Dual Pressure Safety: Yes
System Check: Yes
Leak Detection: Yes
Water Purity Required
<1.0µS-cm Conductivity/>1.0M Ohm-cm Resistivity
Outlet Port
1/8 OD tube
Electrical Requirements
230v 50Hz Va300/
Dimensions(cm/ins)WxDxH
31x49x42/12.2x19.3x16.5
Service Plan Notes
Complete and Standard Plans are available for all models / 1
16 / Nitrogen generator for chromatograph activation
UHP Nitrogen Generators
Model Name: NG Range
Gases Produced:Nitrogen
Applications:DMA,DSC,GC,ICP,TGA,TMA,TOD
The NG Range of Nitrogen Generators is designed specifically for use as make up and carrier gas for GC
applications, which require ultra high purity nitrogen for operation.
The NG Range is also suitable for ICP, DSC,TGA,TOD,TMA and DMA.
Technical Specification
Nitrogen: 250cc/min-4,000cc/minATP
Purity Oz: 99.9995%*
Max Outlet Pressure: 80psi/5.5bar
Pressure Dewpoint: -70°C/-94°F
Gas Outlet: 1x1/4"BSP Female
Electrical Requirements compressor –less systems: 230v 50Hz 0.5A
Electrical Requirements with internal air compressor: 230v 50Hz 2.2A
Dimensions:
-43x41x126cm/
Service Plan Notes
Complete and Standard Service Plans are available / 1
Laboratory Glassware
Measuring cylinder tall form with hexagonal base 1000ml - 21 pcs
Measuring cylinder tall form with hexagonal base 100 ml- 20pcs
Measuring cylinder tall form with hexagonal base 50ml - 16 pcs
Beakers, borosilicate glass 3.3, graduated, low form 250ml- 25 pcs
Beakers, borosilicate glass 3.3, graduated, low form 100 ml- 20pcs
Beakers, borosilicate glass 3.3, graduated, low form 600ml- 3 pcs
Glenammer
100m Diax180 mic Test sieve
Separatory funnels acc to Squibb,borosilicate glass with interchargeable
N5 plag of DTEE 2000 ml
18 / UV-VIS spectrophotometer
UV-3600
UV-VIS Spectrophotometers
The UV-3600 UV-VIS-NIR Spectrophotometer is a new additional model in our product lineup of high-end UV-VIS-NIR spectrophotometers.
The UV-3600, a successor model to the UV-3150/3101, is a unique UV-VIS-NIR instrument that provides for optimum sensitivity from 185-3,300nm.
Features
High Sensitivity - The Only 3 Detector System
Equipped with 3 detectors: A PMT detector (photomultiplier tube) for the ultraviolet and visible regions, and InGaAs and PbS detectors for near infrared region. The InGaAs detector bridges the gap between the PMT - PbS switching wavelength where sensitivity is typically low to ensure high sensitivity over the entire measurement wavelength range. With <0.00003 Abs noise at 1500 nm, the world´s lowest noise level is achieved.
The following figures represents relationships between detectors and measurable ranges
High Resolution - WideMeasurementRange & Ultra-low Stray Light
With its high performance double monochromator, ultra-low stray light (0.00005% or less at 340 nm) is achieved at high resolution (highest resolution 0.1 nm). With a measurement wavelength range of 185 - 3300 nm, measurement is possible over a wide range including the ultraviolet, visible and near infrared regions. This allows spectroscopic analysis in a wide variety of fields.
Software
All-in-one Software
UVProbe software for the UV-3600 includes functions for spectrum measurement, quantitation, kinetics and a report generator. Multiple security and audit-trail functions ensure the reliability of data.
Specifications
Wavelength range / 185nm - 3300nm
Spectral Bandwidth / 8 steps in ultraviolet/visible region 0.1, 0.2, 0.5, 1, 2, 3, 5, 8 nm
10 steps in near-infrared region 0.2, 0.5, 1, 2, 3, 5, 8, 12, 20, 32 nm
Resolution / 0.1 nm
Sampling Pitch / 0.01 - 5 nm
Wavelength accuracy / UV/VIS region: ±0.2 nm NIR region: ±0. 8 nm
Wavelength repeatability / UV/VIS region: less than ±0.08 nm NIR region: less than ±0.32 nm
Stray light / < 0.00008% (220 nm, NaI)
< 0.00005% (340 nm, NaNO2)
< 0.0005% (1420 nm, H2O)
< 0.005% (2365 nm, CHCl3)
Photometric range / -6 to 6 Abs
Noise / 0.00005 Abs or less (500 nm), 0.00008 Abs or less (900 nm), 0.00003 Abs or less (1500 nm)
Slit width 2nm, RMS value at 1 sec. response
Photometric System / Double beam
Dimensions / 1020(W) x 660(D) x 275(H) mm
Accessories
Abundant Accessories for a Wide Variety of Applications
The multipurpose large-sample compartment and integrating sphere attachment enable measurement of solid samples, and with the guaranteed precision of the ASR Series (Absolute Specular Reflectance) attachments, high accuracy, absolute reflectance measurements are possible. Various cell holders like the constant-temperature cell holder and ultra-micro cell holder can also be mounted to enable a wide range of measurement possibilities.
UV-2450/2550
UV-VIS Spectrophotometers
Features
This single monochromator system provides the analyst with the reliability and performance our customers have come to expect in Shimadzu instruments. The low stray light, wide dynamic range, small beam size combination makes this instrument well suited to handle a variety of practical situations.
Software
From small sample volumes to highly absorbing samples, the UV Probe software gives you the power and versatility of an extremely sophisticated system in a package that's easy to learn and use. Additional software enhancements like Kubelka-Munk conversion, area calculations and peak picks make this an instrument for the real world. What's more, it can resolve the triplet peaks in the vapor spectrum of benzene near 250nm with ease.
The UV-2550
An Unparalleled Instrument For Your Demanding Applications
When you can't dilute the sample or reduce the path length, think of the UV-2550. The double blazed grating, double monochromator (DDM) in this instrument is just waiting for those samples. Shimadzu has achieved the world's lowest level of stray light and highest energy throughput with this unique design. By creating two blaze wavelengths (200nm and 600nm), Shimadzu has overcome the energy loss and sensitivity restrictions usually associated with double monochromator instruments. The face of the grating is actually divided into two unique areas for short and long wavelength measurements. The sodium iodide spectra for the double monochromator UV-2550 and a single monochromator instrument described below put this information into practical terms.
Accessories
Capillary Cuvettes
Micro Sample Holder
Specifications
Wavelength Range / 190 to 900nm (performance guaranteed range). Extendable to 1,100nm through the use of the optional photomultiplier. (The measurable range maybe restricted in the shorter wavelength side depending on the type of photomultipler used.)
Monochromator System / UV-2450: Single monochromator with a high-performance blazed holographic grating in the aberration corrected Czerny-Turner mounting.
UV-2550: Double monochromator with a high-performance double-blazed holographic grating in the aberration corrected Czerny-Turner mounting.
Resolution / 0.1nm
Spectral Bandwidth / 0.1, 0.2, 0.5, 1, 2 and 5nm
Wavelength Accuracy / ±0.3nm
Wavelength Repeatability / ±0.1nm
Wavelength Scanning Speed / FAST, MEDIUM, SLOW, and SUPER SLOW
Light Source / 50W halogen lamp (2,000 hours of life) and D2lamp (500 hours of life)
Light Source lamp switching / Selectable between 282nm and 393nm
Stray Light / UV-2450: Less than 0.015% at 220nm and 340nm
UV-2550: Less than 0.0003% at 220nm and 340nm
Detector / Photomultiplier R-928
Photometric System / Double beam, direct ratio system with dynode feedback
Photometric Mode / Absorbance (Abs.), transmittance (%), reflectance (%) and energy (E).
Photometric Range / Abosrbance: -4~5 Abs. (0.001 Abs. increments)
Transmittance: 0~999.9% (0.01 increments)
Reflectance: 0~999.9% (0.01% increments)
Photometric Accuracy / ±0.002Abs(0~0.5Abs), ±0.004Abs(0.5~1Abs),±0.3T (0~100%T)
(all determined with NIST 930D standard filter)
Photometric Repeatability / 0.001Abs (0~0.5Abs), ±0.1%T
Baseline Correction / Selectable with storage in firmware
Baseline Flatness / Within ±0.001Abs.
(excluding noise, 2nm slit width and SLOW wavelength scanning speed)
Drift / Less than 0.0004 Abs. per hour (after 2 hours warm-up)
Dimensions / 570 (W) x 660 (D) x 275 (H) (mm)
Weight / 36kg
Power Supply / AC 100V/120V/220V/240V, 50/60Hz 250VA (swithc-selectable)