Optoelectronics 1: Devices for Optical Communications

VCSEL Laser characteristics

Objective:

To measure the light current characteristic of a Vertical Cavity Surface Emitting Laser Diode (VCSEL) and

hence to determine the lasing threshold and the differential quantum efficiency, above and below threshold.

WARNING: Class IIIB laser hazard

WARNING: Laser light is always dangerous. The light from this laser is

invisible to human eyes and thus is even more dangerous. DO NOT stare at

any point where the beam may emerge from, such as the fibre output or the

connectorised laser diode package. The laser can be powered up using the key

on the front panel of the Thorlabs laser diode driver. Only power up the laser

while there is a fibre patchcord securely attached to the laser, with the other

end of the patchcord firmly connected to an optical power meter. At all other

times, YOU MUST TURN OFF THE LASER DIODE CURRENT. Do not

allow any person(s) not involved with this specific exercise to interfere with the

setup.

Please reread the above instructions and be certain that you understand the

warning before proceeding.

Equipment:

Honeywell HFE4080-32X VCSEL Laser mounted in an ST package on a laser heatsink

ILX 3412 precision Laser diode Driver and associated mains supply and laser interconnect lead

Interconnecting ST connectorised 62.5/125 µm fibre patchcord

Fiber Optic power level meter, Megger OTP 620.

WARNING: The maximum laser diode current must never exceed 12 mA otherwise

permanent damage to the laser will result.

Optical Communications Systems Laboratory, Dr. Gerald Farrell 2002

When the exercise is complete please return all test equipment and leads used to storage and ensure that all protective

dust covers are replaced on fibres, optical sources etc..

Method:

Connect the laser to the ILX laser driver (the front cover is shown below) and the laser optical output to

the optical power meter. The laser driver output can be controlled to within 0.1 mA. Note that to protect

the laser the current output is initially off when the mains power switch is turned on.

Turn on the laser driver using the power button. Turn the laser driver output current control to zero (fully

anti-clockwise) prior to turning on the laser current. To turn on the laser current press the small output

button once (below the current set knob). Rotating the laser output knob clockwise will increase the

current, the value of which is shown on the display. If the display does not show an increase in current

recheck your connections.

To switch off the laser current depress output button once more. To protect the laser from transient

damage do not disconnect the laser from the laser driver at any time when the laser current is enabled. The

correct sequence to disassemble the experiment is to turn off the laser output, then disconnect the laser

from the ILX laser driver, then turn off the ILX driver using the power button.

ILX 3412 Laser driver

Measure the laser diode light-current characteristic, by varying the laser diode current from about 1 mA up

to a maximum of 12 mA, in small increments (typically 0.2 mA, but close to threshold smaller increments of

0.1 mA will be needed for accuracy. Monitor the output of the laser via the supplied ST connectorised

62.5/125 µm optical fibre patchcord, connected to the optical power meter. Set the optical power meter

to measure µW at 850 nm. WARNING: The maximum laser diode current must never exceed 12 mA

otherwise permanent damage to the laser will result.

Precautions: Only power up the laser while there is a fibre patchcord securely

attached to the laser, with the other end of the patchcord firmly connected to an

Optical Communications Systems Laboratory, Dr. Gerald Farrell 2002

When the exercise is complete please return all test equipment and leads used to storage and ensure that all protective

dust covers are replaced on fibres, optical sources etc..

optical power meter. At all other times and when you are not taking measurements,

YOU MUST TURN OFF THE LASER DIODE CURRENT.

Optical Communications Systems Laboratory, Dr. Gerald Farrell 2002

When the exercise is complete please return all test equipment and leads used to storage and ensure that all protective

dust covers are replaced on fibres, optical sources etc..

Results:

Plot the laser diode light-current curve as shown in the sample below and compare with that supplied in the

manufacturers data sheet (supplied). Hence determine the laser threshold and the differential quantum

efficiency, above and below threshold.

The laser threshold is the point where the laser changes its operating mode from a spontaneous emission

(like a light emitting diode (LED)) to stimulated emission (Lasing). By convention the threshold can be

found by drawing a line parallel to the characteristic above threshold. The point where the line intercepts

the X or Current axis is the threshold.

The differential slope efficiency (dL/dI) is the slope of the characteristic at a particular current. The SE has

units of µW per mA (or mW per mA depending on the laser optical power). The SE is low below

threshold and high above. In your results estimate the SE below threshold and at least at three

points above threshold.

In principle the SE should be constant above and below threshold, but in practice the SE above threshold

varies as shown in the diagram below. Serious variations in the SE above threshold along with so-called

"kinks" in the laser characteristic may point to internal defects in the laser. It is for this reason that in laser

production each light-current characteristic is analysed and serious slope changes and kinks are identified in

order in order to eliminate faulty lasers prior to packaging. Check your laser characteristic for kinks. Are

there any present in your opinion and at what drive current?

dL/dI

slope

efficiency

variation

emission

Stimulated

Kink

Spontaneous

emission

Drive Current

Laser threshold

current

Laser diode characteristics

Optical Communications Systems Laboratory, Dr. Gerald Farrell 2002

When the exercise is complete please return all test equipment and leads used to storage and ensure that all protective

dust covers are replaced on fibres, optical sources etc..