WisconsinCenter for Applied Microelectronics
1550 Engineering DrivePhone: 608/262-6877
Madison, WI 53706Fax: 608/265-2614
Deflecking Aluminum Etch Bench Operationand Quick Dump Rinser
Rev. 8/7/12
Material Restrictions:
All materials allowed for use in this equipment are listed in PDF form on the WCAM My Web Space. To view the listing following these steps:
- Open the UW website for My WebSpace.
- Log on using your UW NetID and password.
- Click on the star in the upper left corner.
- Under Bookmarks click on Group Directories.
- Scroll down the list of organizations to WCAM.
- Click on WCAM.
- Double click on the first file folder to open “Approved Materials.”
- Double click on the WebsiteRpt to view the approved materials for equipment.
- Within the PDF you can perform a search.
Personal Protective Equipment (PPE)
While working at etch bath you must wear:
- Face shield and safety glasses
- Chemical apron
- Chemical gloves over the nitrile gloves
Materials
- Silicon, silicon on insulator, quartz
- NO METALS, metal tweezers, glass, Ga/As wafers
Etch Bath
Prepare
- Log onto CRESS.
- If system is off, press PWR on the ETCH BATH controller.
- The system will start in a holding status, press HOLD to release.
- Allow the bath temperature to stabilize at 40C. Approximately 15-20 minutes.
- Insert wafers in the either 3-inch or 4-inch wafer holder. Pieces should be locked in the basket and then the basket placed inside the 3-inch wafer holder.
- Don all PPE – apron, face shield and chemical gloves – before using the etch bath.
- Lift off the white tank lid and set aside.
Wet Etching
- With PPE on, slowly lower holder/basket into acid tank.
- Replace the white tank lid.
- Rinse and remove chemical gloves to operate etch bath controller.
- With chemical gloves off, press START on the controller to begin the timer. The default setting is 5 minutes. Or, manually time etch period.
- At the end of 5 minutes, press STOP/RESET to stop alarm.
Rinsing
- Don all the PPE before removing holder/basket from acid tank.
- Lift off the white tank lid and set aside.
- Go to QUICK DUMPRINSER procedures.
Quick Dump Rinser (QDR)
Start
- Gently lift clear lid of QDR and slide lid down into slot.
- While wearing PPE slowing lift wafer holder/basket from Etch Bath. Hold above tank to allow dripping of acid solution to stop before moving wafer holder/basket to QDR.
- Lower holder/basket into compartment of quick dump rinser (QDR).
- Gently lift clear lid of QDR and cover tank.
- Replace the white tank lid to the acid bath.
- Carefully rinse bench top surface to remove any acid drips.
Rinse
- Rinse and remove chemical gloves before operating controllers.
- Do not PWR offETCH BATH
- With chemical gloves off, press START on RINSER controller.
Fill/dump cycles continue until 10 cycles are completed or the water measures10M-Ohm.
Ends with a final fill and5-second pre-alarm
- Press STOP/RESET to stop alarm.
End
- Go to the drying procedures.
Technique #1: Manually holding wafer or piece with tweezers and blow dry with nitrogen gun.
Technique #2: For full wafers only. Automatically dry wafers by transferring rinsed wafers into holder for Verteq spin dryer. Follow operating procedures.
- Stored DAE wafer holders/basket in QDR
Do not remove holders and use at other process benches.
DAEbench.doc- 1 -