CT CHEST LOW DOSE - PULMONARY NODULES - BEACON HILL
INDICATIONS: We must have a prior CT on PACS documenting known pulmonary nodules in order to use the Low Dose
Protocol. EXCEPTION: “If there is anything else: i.e. nodules + adenopathy: may need routine dose.”
PATIENT PREP: No Prep
IV CONTRAST: None
ORAL CONTRAST: None
POSITIONING: Feet First Supine, Arms above head
TOPOGRAMS: AP. Range: Above lung apices through lung bases/costophrenic angles or adrenals if history of lung cancer. 35 mA, 120 kV.
Lateral. Range: Above lung apices through lung bases/costophrenic angles or adrenals if history of lung cancer. 35 mA, 120 kV.
SCAN TYPE: Spiral/Helical
SCAN RANGE: Above Lung Apices through Lung Bases/Costophrenic Angles (CPA)
*If history of Lung Cancer scan through Adrenals
NOTES: Packs Per Day X Number of Years Smoking = Pack-Years
ALARA – Keep radiation dose As Low As Reasonably Achievable.
*XCARE ON*
CHEST LOW DOSE – PULMONARY NODULES
ScanRange / Scan Direction
Scan Type / Respiration / Scan
Delay
(Seconds) / CARE
DOSE
4D / Quality
Reference
mAs / CARE
kV / Quality
Reference
kV / XCARE / Detector
Configuration
Slices Per Tube Rotation / Pitch
Table Increment/Speed: (mm/rotation) / Rotation
Time
(Seconds) / SFOV
(cm)
-Above Lung Apices
Thru Bases/CPA
-If H/O Lung CA
Scan thru Adrenals / Craniocaudal
Spiral/Helical / Suspended
Inspiration / 5
Seconds / ON / 50 / Semi / 100 / ON / Detectors: 64 x 0.6 mm
Slices Per Tube Rotation: 32 / Pitch: 0.6
Table Increment/Speed:
11.52 mm/rotation / 0.33
Seconds / 50
cm
Plane / SAFIRE
Strength / Slice
Thickness / Interval / Kernal / Window
Width/Level / DFOV
(cm)
RECON – AXIAL SOFT TISSUE / Axial / ON
3 / 3.0 mm / 3.0 mm / I31f
Medium Smooth / Mediastinum
350/50 / FOV just beyond
patient’s body
RECON – AXIAL LUNG / Axial / ON
2 / 3.0 mm / 3.0 mm / I70f
Very Sharp ASA / Lung
1200/-600 / FOV just beyond
patient’s body
RECON – AXIAL MIPS / Axial MIPS
Recon Card 3D Axial MIP Thin / ON
2 / 10.0 mm / 2.0 mm / I70f
Very Sharp ASA / Lung
1200/-600 / FOV just beyond
patient’s body
RECON – CORONAL SOFT TISSUE MPR
Coronals Angled in Plane to Body Part / Coronals Angled in Plane to Body Part
Recon Card 3D Coronal MPR / ON
3 / 3.0 mm / 3.0 mm / I31f
Medium Smooth / Mediastinum
350/50 / FOV just beyond
patient’s body
RECON – SAGITTAL SOFT TISSUE MPR
Sagittals Angled in Plane to Body Part / Sagittals Angled in Plane to Body Part
Recon Card 3D Sagittal MPR / ON
3 / 3.0 mm / 3.0 mm / I31f
Medium Smooth / Mediastinum
350/50 / FOV just beyond
patient’s body
*The operator must check the CTDIvol before and after the scan to ensure it is within the allowed dose range. Scans performed outside of the allowed range must be
documented and reviewed by the designated radiologist and/or physicist.
Allowed CTDIvol Dose Ranges: 0.5 mGy – 50 mGy
XR29 Dose Notification Value (CTDIvol): 50 mGy
Page 1 of 2
Approximate Values for CTDIvolPatient Size / Weight (kg) / Weight (lbs) / CTDIvol (mGy)
Small / 50-70 / 110-155 / 4-10
Average / 70-90 / 155-200 / 8-16
Large / 90-120 / 200-265 / 14-22
Reference: AAPM
*The AAPM recommended NEMA XR29 Dose Notification Value for an adult torso is 50 mGy. Dose notification levels less than the AAPM recommended can be set. The maximum CTDIvol should match the dose notification value. Exams with CTDIvol values less than the minimum allowed range should not be performed unless approved by a radiologist.
NETWORK: Exam to PACS