عطاء رقم 1/ل/2016

توريد وتركيب وتشغيل أجهزة مخبرية/ كلية العلوم

المواصفات الخاصة:

Item / Description / quantity
1 / Real Time PCR System with Notebook, Open System

Features:

5 channels multiplexing for discrimination of up to five targets in a single reaction well*

Strong flexibility thanks to the 96-well format suited for standard PCR microplates, test tubes and strips, Optimal signal/noise ratio, Light emitting diodes (LED)

detection using multiple exposure method

Technical Characteristics :
Thermal block format 96 test tubes of 0.2ml (12x8)
Range of thermal block temperature control 0C…100C
Resolution of temperature setting 0.1C
Absolute accuracy of temperature maintenance ±0.2C
Uniformity of thermal block temperature ±0.15C
Hot cover Temperature 105C ±1C
Excitation source Light –emitting diode (LED)
Detector CCD ( Charge coupled device )
Number of the fluorescence measurement 5
Excitation/detection wave length 470/252, 532/570, 585/633, 633/670, 690/750
Threshold sensitivity of each of the channels for solution of standard fluorophores / 1
2 / Analytical Balances
Capacity g : 210
Readability mg : 0,1 ( 0.0001g)
Linearity mg : ± 0,2
Repeatability mg : ± 0,05
functions and characteristics
  • LCD display with small decimal digits
  • Membrane keyboard, water proof and solvent resistant, easy to use with TARE, ON/OFF, PRINT and MODE keys
  • Indication of the reached stable weight
  • Bar-graph indicator of dosage and remaining capacity of the balance
  • Parameters configurable by menu: reading in g (grams), lb (pound), oz (ounce), ct (carats), pcs (pieces), % (percentage)
  • Full scale automatic calibration with external mass
  • Weighing underneath the balance
  • Selectable response time “fast/slow” (see table)
/ 1
3 / Autoclaves
Specification :
Control / Microprocessor PID Control
Chamber Volume (L) / 50 / 1.8
Temperature 1) / Sterilization Temp. (℃) / 110 to 123
Melting Temp. (℃/℉) / 60 to 100
Safety / Over Temperature Protector / Pressure Safety Valve / Over Current Pr
tector
Warning Alert (Over Temp. / Low Temp. / Low Heat / Sterilization Fail / Low Water Level)
Timer / 0 to 999 min
Programs / Standard modes (Decontamination, Solids Sterilization, Liquids Sterilization, Melting)
Program modes (Sterilization 1, Sterilization 2, Liquid Sterilization, Melting)
/ 1
4 / pH/Temperature Bench Meter
This pH/temperature bench meter provides intuitive software and a well-organized LCD display, making operation quick and simple. It also offers a functional design and reliable calibration results, which are tracked on the display, ensuring precise measurements time after time. This meter measures the pH and the temperature of a sample and includes a built-in 2-in-1 plastic electrode, temperature sensor, and a pH buffer
Specification
Calibration / 2 Pointsat least
1 Predefined Buffer Group
(4.01, 7.00, 10.01)
Error Limits / ±0.01 pH
±0.5° C
Power Supply / 110 to 240 V/50 Hz, DC 12 V
Range / 0.00 to 14.00 pH
0 to 100° C
Resolution / 0.01 pH
0.1° C
/ 1
5 / REFRIGERATED BENCH TOP CENTRIFUGE
Maximum speed : 4.100 rpm Swing-out / 14.000 rpm Angle
Maximum RCF : 3.045 xg Swing-out / 18.405 xg Angle
Maximum capacity : 4 x 200 ml Swing-out / 24 x 1,5/2 ml Angle
Temperature range : -9°C / +40°C (+4°C at max. speed )
Programmable microprocessor control system
Digital displays for prog. no, speed, rcf, time, temperature, acceleration and braking rates
10 program memories
Timer : 1 - 99 minutes and hold position
10 acceleration / 10 braking rates
Safety features : Imbalance detection, lid interlock, chamber guard ring, motor
over temperature , rotor over speed, diagnostic system
ACCESSORIES
Puller for rotor removal
Swing-out rotor 4 x 200 ml, 4.100 rpm, 3.045 xg
Set of 4 buckets
Set of 4 buckets with sealing cap
Set of 4 200 ml PP bottles with cap
Set of 4 inserts 1 x 100 ml
Set of 4 inserts 1 x 50 ml conical
Set of 4 inserts 7 x 15 ml
Set of 4 inserts 3 x 15 ml conical
Set of 4 inserts 7 x 5/7 ml
Angle Rotor 30 x 15 ml 4.100 rpm 3.007 xg
Adaptor for 1,5/2 ml (Ø11mm) tubes, (Should be ordered according to the capacity of the rotor)
Adaptor for 13 x 100 mm vacuumed / non-vacuumed tubes(Should be ordered according to the capacity of the rotor)
Adaptor for 13 x 75 mm vacuumed / non-vacuumed tubes (Should be ordered according to the capacity of the rotor)
Angle rotor 24 x 1,5/2 ml 14.000 rpm, 18.407 x g
Adaptor 500 / 800 μl
Adaptor 200 μl PCR
Adaptor 250 / 400 / 700 μl
Angle rotor 6x50 ml 9.000 rpm, 9.418 xg
Set of 6 adaptors 1x15 ml conical
Set of 6 rubber base adaptors for 50 ml conical tubes
Microtitre rotor 4.100 rpm, 2.349 xg / 1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
6
7 / Fluorometer Dual-channel fluorometer
Equipped with two fluorescence channels for nucleic acid and protein quantitation:
Blue fluorescence channel: Excitation 495nm shortpass (wavelengths up to 495nm), emission 510–580nm
Red fluorescence channel: Excitation 640nm shortpass (wavelengths up to 640nm), emission 660–720nm

Imager System

Imager software with USB connection to PC
Built-in, 21 x 26cm pullout UV-transilluminator
With Imager Analysis software
Viewing window for gel observation with standard 560 nm amber filter.
Specification :
UV light / 312(302) nm
Camera / Minimum 5.0 MP
ImageStorage / PC
Connection / USBto PC
Display / LED
Viewing Window / For gel observation
Light / White light (standard)
Optional Accessory / White light table(power built-in)
470 nm Blue light(power built-in)
Safety / CE, FCC
USB port / for PC connection
Accessories
Optical EtBr filter (620 nm) for SmartView Pro Imager
Optical SYBR Green filter (520 nm) for SmartView Pro / 1
1
8 / Hotplate Mignatic Stirrers
Stirrers Bar set
/ 1
1
9 / Incubated Shakers
Digital incubating benchtop shakers
Operating Features
General Control
 Intuitive touch screen LCD controller with easy icons and logical menus.
 Automatic run is available with non-volatile set point memory after power interruption.
 Built-in RS-232 port and USB port for external control and data collection.
Shaking Control
 Speed range: 10 to 500 rpm.
 Orbit diameter: 19mm.
 Timer range: 1 min. to 999 min.
 Audible/Visible alarm activated by shaking speed deviation.
 Worry-free operation.
- Automatic shaking speed adjustment in case of excessive workload such as unbalanced load placement, unusual vibrations caused from unstable floor or external shock.
 Shaker protection system against over-current, stalled platform check
Specification:
Temperature / Range (℃) / Amb. +5 to 80
Fluctuation 1) at 37℃
in flask(±℃) / 0.1 / 0.18
Variation 1) at 37℃
in flask (±℃) / 0.5 / 0.90
Shaking System / Motion Type / Orbital
Amplitude Size
(mm / inch, dia.) / 19.1 / 0.75
SpeedRange (RPM) / 10 to 500
Accuracy / ±1% of set speed (>100rpm) / ±1 (<100rpm)
Timer / 1 min. to 999 min.
Max. Load
(kg ) / 10 / 22.0 at 500 rpm
18 / 39.7 at 400 rpm
/ 1
10
11
/ TOPLOADING BALANCES
Capacity g : 210
Readability g : 0,001
Linearity g : ± 0.002
Repeatability mg : ± 0.001
Response time : ~ 3 sec.
Pan mm: Ø 110
Calibration mass : External1
WATER DISTILLER
-Capacity: 3.5 lt/hr. at least.
-Stainless steel boiler and condenser
-Stainless steel heater
-Distillation led
-Manometer for water pressure
-Warning ledfor shortage of water or overheating / 1
1