CT CHEST LOW DOSE – LUNG CANCER SCREENING - CANYON PARK
CRITERIA: 1. Age 55-79
2. >29 pack-years of smoking
3. Currently smoking or quit within 15 years
4. Stop screening when quit 15 years
5. No co-morbidities that would significantly limit life expectancy
6. Willing and able to have curative surgery
PATIENT PREP: No Prep
IV CONTRAST: None
ORAL CONTRAST: None
POSITIONING: Feet First Supine, Arms above head.
TOPOGRAMS: AP. Range: Above lung apices through lung bases/costophrenic angles. 25 mA, 110 kV
Lateral: Range: Above lung apices through lung bases/costophrenic angles. 25 mA, 110 kV
SCAN TYPE: Spiral/Helical
SCAN RANGE: Above Lung Apices through Lung Bases/Costophrenic Angles (CPA)
DISPLAY FIELD OF VIEW: DFOV: 1cm beyond the rib cage. Does not need to include entire chest wall thickness.
NOTES: ALARA – Keep radiation dose As Low As Reasonably Achievable.
Packs Per Day X Number of Years Smoking = Pack-Years
CARE DOSE OFF
kV: 110
Adjust mAs According to BMI
BMI <30: 25 mAs
BMI 30-34.9: 30 mAs
BMI >35: 40 mAs
BMI Calculator Website: www.nhlbisupport.com/bmi/
CHEST LOW DOSE – LUNG CANCER SCREENING
ScanRange / Scan Direction
Scan Type / Respiration / Scan
Delay
(Seconds) / CARE
DOSE
4D / mAs / kV / Detector Configuration
Slices Per Tube Rotation / Pitch
Table Increment/Speed: (mm/rotation) / Rotation
Time
(Seconds) / SFOV
(cm)
-Above Lung Apices Thru Bases/CPA / Craniocaudal
Spiral/Helical / Suspended
Inspiration / 6
Seconds / OFF / BMI <30: 25 mAs
BMI 30-34.9: 30 mAs
BMI >35: 40 mAs / 110 / Detectors: 16 x 1.2 mm
Slices Per Tube Rotation: 16 / Pitch: 0.8
Table Increment/Speed: 15.36mm/rotation / 0.6
Seconds / 50
cm
Plane / Slice
Thickness / Interval / Kernal / Window
Width/Level / DFOV
(cm)
RECON – AXIAL SOFT TISSUE / Axial / 2.0 mm / 2.0 mm / B35s
HeartView Medium / Mediastinum
400/40 / 1cm beyond
rib cage
RECON – AXIAL LUNG / Axial / 2.0 mm / 2.0 mm / B70s
Sharp / Lung
1200/-600 / 1cm beyond
rib cage
RECON – AXIAL MIPS
Axial MIPS / Axial MIPS
Recon Card 3D Axial MIP Thin / 10.0 mm / 2.0 mm / B70s
Sharp / Lung
1200/-600 / 1cm beyond
rib cage
RECON – CORONAL SOFT TISSUE MPR
Coronals Angled in Plane to Body Part / Coronals Angled in Plane to Body Part
Recon Card 3D Coronal MPR / 2.0 mm / 2.0 mm / B35s
HeartView Medium / Mediastinum
400/40 / 1cm beyond
rib cage
RECON – SAGITTAL SOFT TISSUE MPR
Sagittals Angled in Plane to Body Part / Sagittals Angled in Plane to Body Part
Recon Card 3D Sagittal MPR / 2.0 mm / 2.0 mm / B35s
HeartView Medium / Mediastinum
400/40 / 1cm beyond
rib cage
Page 1 of 2
*The operator must check the CTDIvol before and after the scan to ensure it is within the allowed dose range. Scans performed outside of the allowed range must be
documented and reviewed by the designated radiologist and/or physicist.
Allowed CTDIvol Dose Ranges: 0.25 mGy – 8 mGy
XR29 Dose Notification Value (CTDIvol): 8 mGy
Approximate Values for CTDIvolPatient Size / Weight (kg) / Weight (lbs) / CTDIvol (mGy)
Small / 50-70 / 110-155 / 0.25-2.8
Average / 70-90 / 155-200 / 0.5-4.3
Large / 90-120 / 200-265 / 1.0-5.6
Reference: AAPM
Allowed CTDIvol Dose Ranges: 0.25 mGy – 8 mGy
XR29 Dose Notification Value (CTDIvol): 8 mGy
The allowed CTDIvol is for a *standard size patient. Individual patients may have higher or lower values to reflect adjustment for patient size.
*The ACR Reference Dose for a "standard size patient" (by definition, is approximately 5' 7" and 155 lbs or 170 cm and 70kg, with a BMI of about 24) is a CTDIvol of less than 3 mGy.**There is no AAPM recommended NEMA XR29 Dose Notification Value for lung screening scans. In general, lung screening exams should not have a CTDIvol greater than 7 mGy. Exams with CTDIvol values less than the minimum allowed range should not be performed unless approved by a radiologist.
NETWORK: Exam to PACS
12/2017 Page 2 of 2