This Computrac SMA-500 is revolutionary moisture analyzer for the plastics industry. Developed as an alternative to the Karl Fischer titration system with a drying oven.
Principles of Operation:
The SMA-500 consists of a cylindrical sample bottle heater, a dry carrier gas flow system and a moisture transducer. The instrument heats a sample of test material contained in a 20mL septum bottle. Evolved volatiles are passed through a cold trap filter to a analysis cell, where the moisture content of the flowing gas is measured. A microprocessor integrates the varying moisture signal and converts the integrated signal to micrograms of water for display. If the sample weight has been entered, the moisture concentration in the sample is calculated as part-per-million or % moisture. Phoemis instrument's revolutionary prediction algorithm automatically terminates the test when sufficient information is collected to allow accurate determination of the moisture content of the sample.
Accurate Results:
The heating range of the SMA-500 is 25℃to 275℃and is accurate within 1℃. Depending upon the sample and test temperature and test ending criteria can be altered to optimize speed and accuracy. Designed for either lab or production floor use, the SMA-500 Moisture Analyzer brings to the forefront state-of-art-technology specifically designed to provide accuracy and precision in a timely manner, so that over-drying is eliminated and product quality is maintained.
Feature:
No reagents
Results displayed in % moisture, ppm moisture, and micrograms water
Instrumentation set-up and calibration take less than 10 minutes
Automatic termination of test
Programmable features include sample name, lot or product ID number, and 7 memory positions for commonly used test parameters
Test time as short as 3 minutes depending upon sample and test temperature
Temperature accurate to +/-1℃
Factory calibration traceable to NIST
Continuous display (text and graphic) of test time, current temperature, programmed test temperature, current microgram water measurement and calculated moisture concentration
Separate three decimal place analytical balance with direct communication capabilities (RS 232)
Self-diagnostic detect hardware, software, sensor and flow system failures
Typical Application / Foods, plastics, chemicals, coatings, pharmaceuticals and paperSpecifications:
Sample Size / 10 milligram to 20 gramsMoistureRange / 10-10,000 micrograms water
10 ppm-1% moisture at 1 gram sample size
HeatingRange / 25- 275℃, Set 1℃increments and maintained to +/-1℃
Oven Capacity / 20 cubic centimeters
Display / 30 character, 9row LCD with black light display
Memory / Storage of last 50 test results
1 quality test program for water injection
Consumables / 20 mL glass sample bottles with septum top
Sample probe
Sample septa
High temperature bottle tops
Hamilton 1.0 microliter syringe
Nitrogen (industrial grade)
5 micron filter
Dimensions / 146mm H x 380mm W x 370mm D;
5.75" H x 15" W x 14.5" D
Weight / 8.6 kg; 19pounds
Power Requirements / 1000120V~50/60Hz, 1 amp(standby) or 8 amps (heat on), or 220-240V~50/60Hz, 0.5 amp(standby) or 4 amps(heat on)
Interfaces / RS-232 (9-pin standard); parallel printer (LPT- PC standard); 5 pin DIN connector for 101 key PC keyboard
PHOEMIS CORPORATION
171 Nan Ping Road, Taichung, Taiwan, R.O.C.
E-mail:
Website:
MSN:
SKPE:sma3000
Tel:886-4-22623476 Fax:886-4-22623471