ABB Instrumentation Division Inc.

Warminster, Pa.

Magnetic Flowmeter Calibration Test Report

Sales Order No.:RA78696FW Date :19 MAY 2000

Serial No.:00W021664 Line Item:1

Reference#:96W046070

Meter Size :2 inch Model No.:10D1475SN1

Sec. Serial No.:00W021664A Sec Model No.:50XM13NX

Excitation :7.5 Line frequency :60 Hz

Meter Factor 1.8127

Std Deviation .083

Std best fit: .056 Zeroshift: -.005 Ft/sec

Test Max Freq : 10000.0 HZ Test Velocity: 12.440 Ft/Sec

TEST DATA

Run # Weight Density Rate cm3/Sec Time Counts

01 564.4138 0.99818 6295.41422 40.751 406526

02 564.8142 0.99816 6279.48275 40.884 407170

03 454.9045 0.99813 5037.62000 41.048 327844

04 452.6522 0.99809 4988.39900 41.248 326222

05 342.7924 0.99807 3795.10863 41.059 246733

06 345.6453 0.99803 3786.61190 41.497 248729

07 232.0318 0.99799 2534.89792 41.612 166971

08 227.7775 0.99797 2522.19173 41.058 163950

09 211.3111 0.99803 1276.42816 75.249 152004

10 210.8106 0.99805 1267.17370 75.618 151651

Run # Hertz Actual GPM Indicated GPM % Max % Rate Tol

01 9975.85 99.78440 99.82808 +.044 +.044 OK

02 9959.15 99.53188 99.66096 +.129 +.130 OK

03 7986.84 79.84794 79.92413 +.076 +.095 OK

04 7908.80 79.06777 79.14309 +.075 +.095 OK

05 6009.23 60.15373 60.13420 -.020 -.032 OK

06 5993.90 60.01905 59.98082 -.038 -.064 OK

07 4012.57 40.17897 40.15366 -.025 -.063 OK

08 3993.13 39.97757 39.95915 -.018 -.046 OK

09 2020.01 20.23181 20.21422 -.018 -.087 OK

10 2005.49 20.08512 20.06886 -.016 -.081 OK

Required Accuraccy :1/4% OF RATE

Secondary used in calibration 50XM

Hydraulic test performed by :R. Spencer

TEST EQUIPMENT USED :

E1028 3852A COUNTER

E1054 3852A COUNTER

E3560 MULTIMETER

M**98 METTLER 3000 LBS SCALE

T**90 THERMISTOR

SWS18.BAS, RM 120998 15:05:26 21 42

ABB Instrumentation Division Inc.

Warminster, Pa.

Magnetic Flowmeter Calibration Test Report

Sales Order No.:RA78696FW Date :19 MAY 2000

Serial No.:00W021664 Line Item:1

Reference#:96W046070

Meter Size :2 inch Model No.:10D1475SN1

Sec. Serial No.:00W021664A Sec Model No.:50XM13NX

Excitation :7.5 Line frequency :60 Hz

Max Flow: 100.000 GPM Sp. Gr. : 1.000

Meter Factor 1.8127

Cal Factor 268.14 GPM

Run # Actual GPM Indicated GPM Diff % Rate

01 99.784 99.828 +.044

02 99.532 99.661 +.130

03 79.848 79.924 +.095

04 79.068 79.143 +.095

05 60.154 60.134 -.032

06 60.019 59.981 -.064

07 40.179 40.154 -.063

08 39.978 39.959 -.046

09 20.232 20.214 -.087

10 20.085 20.069 -.081

All Flowmeters are hydraulically calibrated in accordance with

ANSI/Z540 and are traceable to the NATIONAL INSTITUTE OF STANDARDS

AND TECHNOLOGY. The ABB Instrumentation Div. Warminster facility

is certified to ISO 9001.

This Calibration report may not be reproduced, except in full,

without written permission.

Secondary used in calibration 50XM

Hydraulic test performed by :R. Spencer

TEST EQUIPMENT USED :

E1028 3852A COUNTER

E1054 3852A COUNTER

E3560 MULTIMETER

M**98 METTLER 3000 LBS SCALE

T**90 THERMISTOR

SWS18.BAS, RM 120998 15:05:26 21 42

ABB Instrumentation Division Inc.

Warminster, Pa.

Magnetic Flowmeter Calibration Test Report

AS RECEIVED

Sales Order No.:RA78696FW Date :19 MAY 2000

Serial No.:00W021664 Line Item:1

Reference#:96W046070

Meter Size :2 inch Model No.:10D1475SN1

Sec. Serial No.:00W021664A Sec Model No.:50XM13NX

Excitation :7.5 Line frequency :60 Hz

Max Flow: 100.000 GPM Sp. Gr. : 1.000

Meter Factor 1.81344

Cal Factor 268.4 GPM

Run # Actual GPM Indicated GPM Diff % Rate

01 166.353 166.738 +.231

02 82.391 82.630 +.289

03 32.933 32.974 +.126

All Flowmeters are hydraulically calibrated in accordance with

ANSI/Z540 and are traceable to the NATIONAL INSTITUTE OF STANDARDS

AND TECHNOLOGY. The ABB Instrumentation Div. Warminster facility

is certified to ISO 9001.

This Calibration report may not be reproduced, except in full,

without written permission.

Secondary used in calibration 50XM

Hydraulic test performed by :R. Spencer

TEST EQUIPMENT USED :

E1028 3852A COUNTER

E1054 3852A COUNTER

E3560 MULTIMETER

M**98 METTLER 3000 LBS SCALE

T**90 THERMISTOR

SWS18.BAS, RM 120998 15:05:26 21 42