ABB Instrumentation Division Inc.
Warminster, Pa.
Magnetic Flowmeter Calibration Test Report
Sales Order No.:RA78696FW Date :19 MAY 2000
Serial No.:00W021664 Line Item:1
Reference#:96W046070
Meter Size :2 inch Model No.:10D1475SN1
Sec. Serial No.:00W021664A Sec Model No.:50XM13NX
Excitation :7.5 Line frequency :60 Hz
Meter Factor 1.8127
Std Deviation .083
Std best fit: .056 Zeroshift: -.005 Ft/sec
Test Max Freq : 10000.0 HZ Test Velocity: 12.440 Ft/Sec
TEST DATA
Run # Weight Density Rate cm3/Sec Time Counts
01 564.4138 0.99818 6295.41422 40.751 406526
02 564.8142 0.99816 6279.48275 40.884 407170
03 454.9045 0.99813 5037.62000 41.048 327844
04 452.6522 0.99809 4988.39900 41.248 326222
05 342.7924 0.99807 3795.10863 41.059 246733
06 345.6453 0.99803 3786.61190 41.497 248729
07 232.0318 0.99799 2534.89792 41.612 166971
08 227.7775 0.99797 2522.19173 41.058 163950
09 211.3111 0.99803 1276.42816 75.249 152004
10 210.8106 0.99805 1267.17370 75.618 151651
Run # Hertz Actual GPM Indicated GPM % Max % Rate Tol
01 9975.85 99.78440 99.82808 +.044 +.044 OK
02 9959.15 99.53188 99.66096 +.129 +.130 OK
03 7986.84 79.84794 79.92413 +.076 +.095 OK
04 7908.80 79.06777 79.14309 +.075 +.095 OK
05 6009.23 60.15373 60.13420 -.020 -.032 OK
06 5993.90 60.01905 59.98082 -.038 -.064 OK
07 4012.57 40.17897 40.15366 -.025 -.063 OK
08 3993.13 39.97757 39.95915 -.018 -.046 OK
09 2020.01 20.23181 20.21422 -.018 -.087 OK
10 2005.49 20.08512 20.06886 -.016 -.081 OK
Required Accuraccy :1/4% OF RATE
Secondary used in calibration 50XM
Hydraulic test performed by :R. Spencer
TEST EQUIPMENT USED :
E1028 3852A COUNTER
E1054 3852A COUNTER
E3560 MULTIMETER
M**98 METTLER 3000 LBS SCALE
T**90 THERMISTOR
SWS18.BAS, RM 120998 15:05:26 21 42
ABB Instrumentation Division Inc.
Warminster, Pa.
Magnetic Flowmeter Calibration Test Report
Sales Order No.:RA78696FW Date :19 MAY 2000
Serial No.:00W021664 Line Item:1
Reference#:96W046070
Meter Size :2 inch Model No.:10D1475SN1
Sec. Serial No.:00W021664A Sec Model No.:50XM13NX
Excitation :7.5 Line frequency :60 Hz
Max Flow: 100.000 GPM Sp. Gr. : 1.000
Meter Factor 1.8127
Cal Factor 268.14 GPM
Run # Actual GPM Indicated GPM Diff % Rate
01 99.784 99.828 +.044
02 99.532 99.661 +.130
03 79.848 79.924 +.095
04 79.068 79.143 +.095
05 60.154 60.134 -.032
06 60.019 59.981 -.064
07 40.179 40.154 -.063
08 39.978 39.959 -.046
09 20.232 20.214 -.087
10 20.085 20.069 -.081
All Flowmeters are hydraulically calibrated in accordance with
ANSI/Z540 and are traceable to the NATIONAL INSTITUTE OF STANDARDS
AND TECHNOLOGY. The ABB Instrumentation Div. Warminster facility
is certified to ISO 9001.
This Calibration report may not be reproduced, except in full,
without written permission.
Secondary used in calibration 50XM
Hydraulic test performed by :R. Spencer
TEST EQUIPMENT USED :
E1028 3852A COUNTER
E1054 3852A COUNTER
E3560 MULTIMETER
M**98 METTLER 3000 LBS SCALE
T**90 THERMISTOR
SWS18.BAS, RM 120998 15:05:26 21 42
ABB Instrumentation Division Inc.
Warminster, Pa.
Magnetic Flowmeter Calibration Test Report
AS RECEIVED
Sales Order No.:RA78696FW Date :19 MAY 2000
Serial No.:00W021664 Line Item:1
Reference#:96W046070
Meter Size :2 inch Model No.:10D1475SN1
Sec. Serial No.:00W021664A Sec Model No.:50XM13NX
Excitation :7.5 Line frequency :60 Hz
Max Flow: 100.000 GPM Sp. Gr. : 1.000
Meter Factor 1.81344
Cal Factor 268.4 GPM
Run # Actual GPM Indicated GPM Diff % Rate
01 166.353 166.738 +.231
02 82.391 82.630 +.289
03 32.933 32.974 +.126
All Flowmeters are hydraulically calibrated in accordance with
ANSI/Z540 and are traceable to the NATIONAL INSTITUTE OF STANDARDS
AND TECHNOLOGY. The ABB Instrumentation Div. Warminster facility
is certified to ISO 9001.
This Calibration report may not be reproduced, except in full,
without written permission.
Secondary used in calibration 50XM
Hydraulic test performed by :R. Spencer
TEST EQUIPMENT USED :
E1028 3852A COUNTER
E1054 3852A COUNTER
E3560 MULTIMETER
M**98 METTLER 3000 LBS SCALE
T**90 THERMISTOR
SWS18.BAS, RM 120998 15:05:26 21 42