SOLITAIRE PHARMACIA PVT. LTD.
- Objective
To define the procedure for Operating and Calibration procedure ofK.F. Autotitrator.
- Scope
This procedure is applicable to Quality Control departmentatSolitaire Pharmacia Pvt. Ltd., Baddi.
- Responsibility
Executive Quality Control
- Accountability
HeadQuality Control
- Procedure
- Instrument Detail:
Name / K.F.Autotitrator
Make / Esico
Model / 1780
Sr. No. / 1002030
Instrument ID. / Location
QCD/KFA/007 / Chemical Lab.
- Operation Procedure:
- Ensure that instrument is clean and free from dust.
- Switch “ON” the main.
- System then perform self test for neutralization.
- Press “CAL” it will display as CAL – 25.
- Press “Enter” for inject. After completion of Neutralization a beep sound will appear.
- Operation Procedure for Water Factor:
- After completion of Neutralization press “CAL”
- Press “Enter” twice display will show : H2O.
- Press “Enter” >enter weight of water drop >then press enter for injection.
- After completion of titration a beep sound will appear.4.3.5Press “%” for result .
- Operation Procedure for Water Content:
- After gating water content, press “SPL”.
- Enter the weight of sample.
- Press “Start”.
- After completion the titration a beep sound will appear.
- Press “%” for result, result will appear in % in Display.
- Take properly entry of water factor and water content on log book.
- Calibration Procedure:
- Neutralize the water in methanol by Karl Fischer Autotitrator (Use procedure 5.2).
- Weigh accurately 100 mg of sodium tartrate dihydrate as a standard. Add into the titration flask and titrate. Record the volume required for the sample.
- Take three reading and record obtained results by Karl fisher Autotitrator.
- Calculate the average water equivalent factor of Karl Fischer reagents.
- Standard value of titrer factor is 6.0 limits: ± 0. 5
- Fill the data.
- Records/Formats:
SPL/QC/043-Log book for Water factor.
SPL/QC/044-Log book for water content.
SPL/QC/045-Format for Calibration of Autotitrator.
- Annexure:
NA
- Reference:
Nil
- Abbreviations:
SPL-Solitaire Pharmacia Pvt. Ltd., Baddi
QC-Quality Control
SOP-Standard Operating Procedure
- Revision History:
Revision No / Effective date / Reason for Revision
00 / New Document
Format No.SPL/QA/001
SOLITAIRE PHARMACIA PVT. LTD.
Log book for Water factor
Format No : SPL/QC/043
Sr. No. / Date / Product Name / Batch No. / Operation Time / Operated By / Check By / RemarksOn Time / Off Time
Log book for Water Content
Format No : SPL/QC/044
Sr. No. / Date / Product Name / Batch No. / Operation Time / Water Factor / Water Content / Operated By / Check ByOn Time / Off Time
Format for Calibration of Autotitrator
Format No : SPL/QC/045
Date of Calibration:Next Due Date:
Name of Instrument:Autotitrator
Make:Esico
Model No.:1780
User Department:Quality Control
Sensitivity:0.001 mg/ml
Instrument Code:QCD/KFA/007
Calibration Schedule:Monthly
Environmental Parameter:Temperature 25oC
Sr. No. / Factor / Wt. of di-Sodium Tartrate( in mg) / Volume Consumed
( in ml) / Water Content
01
02
03
Average
% RSD
Remark: The instrument is working/ not working under satisfactory condition.
Analyzed By:Checked By:Approved By:
Date : Date : Date :
Format No.SPL/QA/001