Helpstarts on page10.

Region:

EDAX Inc.
Ametek, Materials Analysis Division
91 McKee Drive
Mahwah, NJ 07430
USA
Tel: 201 529 4880
Fax: 201 529 4760

Acceptance Test Report

Date:
Customer:
EDAX Order Number (S#):
Analyzer Serial Number:
Ardexus Product Number:

June 6, 2016

Rev 1.97

CustomerInformation / System Information
Customer: / Analyzer Model: / S/N:
User: / Microscope Vendor:
Address: / Microscope Model:
Address: / EDAX Computer: / S/N:
Address: / Operating System: / Svc Pack:
City: / Printer: / S/N:
State: / Monitor: / S/N:
Zip code: / Order Number (S#):
Country:
Phone:

SystemConfiguration

EDS (Yes/No)? / YesNo / WDS (Yes/No)? / YesNo / EBSD (Yes/No)? / YesNo
User Interface / Electrical
Check monitor function: / YesNoOKN/A / Check Mains supply: / YesNoOKN/A
Check Keyboard function: / YesNoOKN/A / Measured Input Voltage (in VAC):
Check Mouse function: / YesNoOKN/A / Line Frequency (in Hz):
Check / Set regional settings: / YesNoOKN/A / Check Ground/Earthing: / YesNoOKN/A
N/A / YesNoOKN/A
Analyzer / Check Hyperthreading is disabled: / YesNoOKN/A
Check cooling fans in analyzer: / YesNoOKN/A / Check condition of all cables and connectors: / YesNoOKN/A
Check HDWrite-cache setting: / YesNoOKN/A / Check PCBs inside EDAX Analyzer (seated properly): / YesNoOKN/A
Imaging / Stage andColumn Control
Check all software and hardware functions: / YesNoOKN/A / Stage Control: / YesNo
Check and adjust imaging option: / YesNoOKN/A / Column Control: / YesNoOKN/A
Imaging (+ imaging option): / YesNoOKN/A / DCOM Control (Yes/No): / YesNo
Mapping: / YesNoOKN/A / N/A:
Line Scan: / YesNoOKN/A / Microscope Software Version:
Line Measurements, Scan Extent: / Microscope Control notes:
X = / Port or IP address
Y =
Use SG2 Shell program, print results (PDF + Printer): / YesNoOKN/A

Customer Instruction

Advise customer to store system restoration software: / YesNoOKN/A
Advise customer to store critical hardware: fuses, tools, boxes: / YesNoOKN/A
Install TEAM Remote software on customer’s computer: / YesNoOKN/A

Refer to EDAX web site for additional training:

Login Setup

Check Administrator and User logins with appropriate privileges: / YesNoOKN/A
Check Windows Login: / YesNoOKN/A
Check EDAX Software Login: / YesNoOKN/A
Show the user where they can add their company’s logo in TEAM: / YesNoOKN/A

Installation Documentation Complete

Installation problems reported to TFS? Data entered in Ardexus?

Complete these pages: Pages 1-3

Comments / Notes

Warranty Start Date:
Customer:
User:
Signature:
Date:
EDAX Representative:
Date:

Contact information

Sales: / Applications:
Phone: / Phone:
E-Mail: / Email:
Service:
Phone:
E-Mail:

Printthese pages: 1-3

File, Save As…. “ Acceptance.doc” or “ Acceptance.pdf

Email a copy of this document to

Skip this page. This page does not apply!

EDS Information

Detector Model: / Apollo 10Apollo PrimeApollo XFApollo XLTApollo XLTWApollo XLT2Apollo XLT2 WOctane Elite 25Octane PlusOctane PrimeOctane ProOctane SuperOctane UltraOctane T Ultra WOctane T Optima-30Octane T Optima-60Octane T Plus / S/N:
Detector Part Number: / Motorized Slide (Yes/No)? / YesNo
EDAX Software (select): / TEAMGenesisGenesis + TEAMTEAM PrimeGenesis Prime / eDPP: Yes
EDAX Sales Order Number (S#): / Ardexus Product Number:

EDS Detector

Adjust mechanical setup of the detector: / YesNoOKN/A
Adjust the hard stops of the detector: / YesNoOKN/A
Verify the vacuum integrity of the microscope: / YesNoOKN/A
Use Shell program, calibrate all Amp Times using Al and Cu: / YesNoOKN/A
N/A / YesNoOKN/A
Use Shell program, print results (PDF + Printer): / YesNoOKN/A
Record MnKα resolution (eV):
Confirm the proper Temp Offset: / YesNoOKN/A
Confirm the Target Temperature: / YesNoOKN/A
Verify detector cool down: / YesNoOKN/A
N/A / YesNoOKN/A
N/A / YesNoOKN/A
Check detector status LEDs: / YesNoOKN/A

Customer Instruction for EDS

Check
EDS Detector handling including startup, window precautions: / YesNoOKN/A
System calibration, geometry parameters: / YesNoOKN/A
TEAM Overview of the user interface:Project Tree,Diagnostic Panel, etc.: / YesNoOKN/A
Genesis Overview of the user interface: / YesNoOKN/A
Recording and storing spectra, identifying elements, HPD function, background operations, SEC factors (in Spectrum Only and Point Analysis modes): / YesNoOKN/A
Mapping: Image acquisition, spectral mapping, phase mapping, drift correction, setting up, use of the pie chart: / YesNoOKN/A
Qualitative and quantitative Line Scan: / YesNoOKN/A
Processing of stored data sets (Quantitative mapping, RGB): / YesNoOKN/A
Evaluation / printout of spectra, storing of results for and analysis report, data storage, report generator: / YesNoOKN/A
Show location and access of software movies: / YesNoOKN/A
Advise customer on installing remote software: / YesNoOKN/A
Number of hours spent on user training:
Basic Operation Training by Service complete (Yes/No)? / YesNo

Refer to EDAX web site for additional training:

Comments / Notes:

Skip this page. This page does not apply!

Detector Performance

Resolution / Check
MnKα resolution / eV / YesNoOKN/A
BKα resolution / eV / YesNoOKN/A
CKα resolution / eV / YesNoOKN/A
NKα resolution / eV / YesNoOKN/A
OKα resolution / eV / YesNoOKN/A
FKα resolution / eV / YesNoOKN/A
Check Peak shift between 20% and 40% Dead Time / eV / YesNoOKN/A
Check low energy efficiency at 20kV using Cu (Net Int. CuLα/CuKα) / YesNoOKN/A

SEC Factor Correction

Element / Sample / HV / Wt% / Wt% / Peak pos. / Exp. SEC / Meas.SEC / Check
BKα / CaB6 / 10kV / B=61.81 / Ca=38.19 / 0.2-4 / YesNoOKN/A
CKα / SiC / 5kV / C=29.95 / Si=70.05 / 0.4-2 / YesNoOKN/A
NKα / CrN / 15kV / N=21.22 / Cr=78.78 / 0.1-1 / YesNoOKN/A
OKα / Fe203 / 15kV / O=30.06 / Fe=69.94 / 0.2-1 / YesNoOKN/A
FKα / CaF2 / 10kV / F=48.66 / Ca=51.33 / 0.2-1 / YesNoOKN/A

Skip this page. This page does not apply!

EBSD Information

EBSD Model Number: / S/N:
EBSD Part Number:
EDAX Software (select): / TEAMOIM + TEAMOIM
TEAM Software version:
N/A
EDAX Sales Order Number (S#): / Ardexus Product Number:

EBSD Detector

Check
Check Phosphor screen for damage / YesNoOKN/A
Check Hard Stop adjustment / YesNoOKN/A
Check Touch Alarm sensor / YesNoOKN/A
Check Motion control / YesNoOKN/A
Check / Adjust Analyze Scale settings / mm / YesNoOKN/A
Check SEM vacuum during interface I/O movement / YesNoOKN/A
Check Camera Focus and Centering. Adjust video if required / YesNoOKN/A
Verify Top/Bottom orientation of Camera / YesNoOKN/A

System Calibration

Calibrate on Ni Sample at center of image at specified optimal WD

Check
Calibrate on Ni sample at center of image / Fit < 1.0 / CI > 0.8 / YesNoOKN/A
SEM calibration (in Beam Setting Panel) / Width: / Height: / YesNoOKN/A
Check Beam Control
Pixels/µm / X= / Y= / Z= / YesNoOKN/A
Accessory: Forward Scatter Detector (FSD) / YesNoOKN/A

Skip this page. This page does not apply!

System Performance for EBSD

On Standard TSL Ni Sample / Check
PC Values saved in software (Z = ~7 % nominal)
Pattern center (PC) calibration performed / YesNoOKN/A
WD=(-4mm) / X= / Y= / Z= / YesNoOKN/A
WD=(-2mm) / X= / Y= / Z= / YesNoOKN/A
WD=() / X= / Y= / Z= / YesNoOKN/A
WD=(+2mm) / X= / Y= / Z= / YesNoOKN/A
WD=(+4mm) / X= / Y= / Z= / YesNoOKN/A
WD=(+6mm) / X= / Y= / Z= / YesNoOKN/A
WD=(+8mm) / X= / Y= / Z= / YesNoOKN/A
Check values in .ini file / YesNoOKN/A
Check image of Nickel sample / YesNoOKN/A
Software indexed patterns (CI > 0.100) on standard TSL Ni sample / YesNoOKN/A
Check software scans and indexes 500 x 500 µm grid at 10 µm step / YesNoOKN/A

Customer Instruction

EBSD detector handling / YesNoOKN/A
Calibration of the camera for the EBSD analysis / YesNoOKN/A
Calibrating indexing / YesNoOKN/A
Phase list / YesNoOKN/A
Hough space / YesNoOKN/A
Interactive work / YesNoOKN/A
Recording a EBSD scan / YesNoOKN/A
Loading the EBSD scans in the OIM Analysis software / YesNoOKN/A
First steps with the Analysis software / YesNoOKN/A
Advise customer on installing remote software / YesNoOKN/A
Number of hours spent on User training
Basic Operation Training by Service complete (Yes/No)? / YesNo

Refer to EDAX web site for additional training:

Comments / Notes:

Skip this page. This page does not apply!

WDS Information

WDS Model Number: / S/N:
WDES Part Number:
EDAX Software (select): / TEAMGenesisGenesis + TEAM / TEXS or LEXS ? LEXSTEXS
TEAM Software version:
N/A
EDAX Sales Order Number (S#): / Ardexus Product Number:

P10 Gas

Check Regulator pressure (0.14-0.28 Bar / 2.03-4.06 PSI) / YesNoOKN/A
Check P10 Gas Flow (Max 2 liter / hour) / YesNoOKN/A

Specimen Current Check

Check Specimen Current Meter / YesNoOKN/A
Check RS232 Communication / YesNoOKN/A
Compare read out between Meter and Current Monitor / YesNoOKN/A

This section does not apply to LEXS

Power / YesNoOKN/A
USB / YesNoOKN/A
X Motor / YesNoOKN/A
Y Motor / YesNoOKN/A
Z Motor / YesNoOKN/A
X Encoder / YesNoOKN/A
Y Encoder / YesNoOKN/A
Z Encoder / YesNoOKN/A
FPGA / YesNoOKN/A
Check optimal settings for Horizontal (H) / YesNoOKN/A
Check optimal settings for Vertical (V) / YesNoOKN/A
Check optimal settings for Length (L) / YesNoOKN/A

WDS Performance Test

Use WDS Shell program + Test Certificate (LEXS 10 kV / TEXS 20 kV)
Calibrate Diffractor 1 using B Kα (LEXS) / YesNoOKN/A
Calibrate Diffractor 1 using Cu Kα (TEXS) / YesNoOKN/A
Calibrate Diffractor 2 using C Kα (LEXS + TEXS) / YesNoOKN/A
Calibrate Diffractor 2 using N Kα (LEXS + TEXS) / YesNoOKN/A
Calibrate Diffractor 3 using O Kα (LEXS + TEXS) / YesNoOKN/A
Calibrate Diffractor 3 using Cu Kα (LEXS + TEXS) / YesNoOKN/A
Calibrate Diffractor 4 using AL Kα (LEXS + TEXS) / YesNoOKN/A
Calibrate Diffractor 5 using Si Kα (LEXS + TEXS) / YesNoOKN/A

Skip this page. This page does not apply!

DQX Adjustments

Calibrate Diffractor 1 using B Kα (LEXS) / YesNoOKN/A
Calibrate Diffractor 1 using Cr Kα + Cu Kα (TEXS) / YesNoOKN/A
Calibrate Diffractor 2 using C Kα + N Kα / YesNoOKN/A
Calibrate Diffractor 3 using O Kα + Cu Lα / YesNoOKN/A
Calibrate Diffractor 4 using Al Kα + Cu Lα / YesNoOKN/A
Calibrate Diffractor 5 using Si Kα + Mo Lα / YesNoOKN/A
Print and attach Acceptance Test Certificate (2x) to report / YesNoOKN/A

Customer Instruction

Check
WDS Detector handling (P10 gas, flow regulator/meter, WDS optics): / YesNoOKN/A
Basic settings for a WDS measurement (WD, auto focus, element selection): / YesNoOKN/A
TEAM Overview of the user interface Project Tree, Diagnostic panel, etc.: / YesNoOKN/A
Recording, storing spectra (element & Energy scan, Intelligent scan), identifying elements / YesNoOKN/A
Simultaneous acquisition of EDS spectra, overlay WDS-EDS spectra, multi-element scans / YesNoOKN/A
Quantitative WDS analysis of SiO2 (also in combination with EDS) / YesNoOKN/A
WDS elemental mapping images (combined WDS/EDS mapping, sequential WDS mapping) / YesNoOKN/A
Line Scan / YesNoOKN/A
Storing of results / YesNoOKN/A
WDS Shell software-brief explanation, setting the proportional counter for a new element / YesNoOKN/A
Number of hours spent on User training
Basic Operation Training by Service complete (Yes/No)? / YesNo

Refer to EDAX web site for additional training:

Comments / Notes:

Help

Entering Data

The Acceptance report is made with fields that should be filled out electronically. Because of different regional requirements and configurations, some fields will change based on region and configurations. Always start at the top and work down.

If the form does not open and allow editing, select View and thenEdit Document first.

Filling out the form automatically populates some fields later in the form, such as Customer and Sales numbers.

Fill in the shaded fields with information. You can use the mouse to move from field-to-field. Do NOT use [Enter].

In many cases, a pull down selection can be made to make data entry easy.

The [DEL] key in a field clears the entry.

Security

This document uses macros to work properly. Be sure to Enable Macros in Trust Center settings.

Select View, then Edit in Word 2013 to allow the document to be edited.

Help

Some help on some of the entries can be found by clicking on the underlined heading. Hover the mouse over a subject to see if it links to Help. For example, ctrl-click on Customer Information on Page 2.Otherwise, scroll through the Help notes, or use the Search function. Use the Service Manuals for installation, test and adjustment procedures. [Alt] [Left arrow] returns to spot in the report.

Using pull-down choices

Select OK when a check of a section is completed successfully to acknowledge it is done.

Select Yes to confirm a configuration.

Select No if the check is not done, or not successful.

Select N/A if the check does not apply to the configuration, or can not be done for some reason.

Page 1

Selecting a Region

Selecting the region will affect the report. Be sure to start the report with the correct region.

If the region you need is not one of the available selections, you can add a region on page18.Then select Other region. The Region information will update after a region is selected and the next field is selected. Or click the Update button on page 2.

Date of Report

The date field on page 1 should be the date the report is filled out.

Customer, EDAX Order #, Analyzer serial number

These are needed to identify the system.

The customer name should be the name of the Company where the system is installed.

The EDAX Order # (Sales Number or S#), can be found on the Shipping paper work, or in the About window from the TEAM or Genesis software.

***The Ardexus Product number can be obtained from the local Sales/Service office.

Page 2

Customer Information – Fill out as required. If all of the address fields are not needed, enter a [-] in the empty field. The Customer field will be automatically carried over from Page 1.

System information – Enter the information as available.

The Analyzer Model in many cases will be the EDAX Computer, such as an “X1” or “X2”. Serial number can be found on the system label on the back or top of the unit.

Enter the rest of the information as configured. If a printer is not supplied, enter “Not supplied”.

The EDAX Order Number will be carried over from Page 1.

System Configuration – Fill out as configured

If the system has an EDS detector, select Yes, otherwise select No.

If the system has an WDS detector, select Yes, otherwise select No.

If the system has an EBSD detector, select Yes, otherwise select No.

The Update buttons on the report, run a macro that updates the fields throughout the form to populate fields, and configure the form to the specific configuration.

Hyperthreading

To check if hyperthreading is enabled

Type wmic in run box

Type cpu Get NumberofCores,NumberofLogicalProcessors /Format:List in the window

If the Number of Logical Processors is greater than Cores, then hyperthreading is enabled

Type exit to quit

Hard Disk Write-cache buffer flushing:It is recommended that you turn off the Windows write-cache buffer flushing option, leaving it on can slow down the system response times. This setting is in the Windows Device Managers – Disk Drives – Policies tab.

Electrical

Check the Mains Supply. Measure the actual input voltage where the EDAX system will plug into.

Record the Line Frequency in Hz where the EDAX system will plug into.

Check the Ground/Earthing where the EDAX system will plug into. Request information from the customer or building operations as needed. Select the appropriate response from the pull down selections.

Imaging – Set up the imaging for the microscope model per the specific microscope interface documentation. Use the service Shell to configure the imaging interface parameters.

Print the parameters from the Shell to a PDF and printer, if available. Select the response from the pull down selections.

Stage and Column Control – Select whether the system is configured with stage and/or column control software per the order. If the system uses DCOM control, select Yes. Enter the Microscope software version. Add any specific notes as desired, such as IP address, or Com port used to the microscope control notes field.

Customer Instruction – There are some checks to confirm that the customer has been informed about storing software and hardware that should be safely stored.

Login Setup – This section confirms that the User has been shown how to log in to Windows and the EDAX software.

Page 3

Installation documents – If any quality problems occurred during the installation, such as missing or damaged parts, they should be reported via the TFS system. See the local service manager for access.

Comments / Notes – A free form area to add notes about the system, that should be noted.

Enter the Warranty start date.

The Customer and User names will be carried over from previous entries.

There is a line for a customer’s physical signature, if the form is printed.

EDAX Representative is the installing service engineer’s name and the date.

Contact information

Enter the information for the customer with the regional EDAX contacts for service, sales and applications.

The Update button will ensure that all of the fields enter so far, populate the complete report.

Printing – The report is meant to be filled out electronically. Then once the report is complete, the pages to be printed are listed on Page 3.

Example:

Under the Print menu option, set the pages to print, select the Page per sheet option, and select the Paper Size appropriate for your region. Select Scale to Paper Size for correct formatting.

Save the file as a doc or pdf file, using a name format “S###### Acceptance Report.doc” where S###### is the Sales Order number. Save the file to C:\EDAX Service\.

Save it as a PDF using CutePDF. Select CutePDF as the Printer. Set the Custom Print Range as shown on Page 4

Email a copy of the report after the installation has been completed.

Page 4

EDS Information – List the actual information from the label on the detector.

Select the software installed on the system and version information.

EDAX Sales and Ardexus numbers are automatically filled in from previous entries.

EDS Detector – The detector should be installed and calibrated according to installation procedures in the service manual (on disk in c:\EDAX Service\Service Manuals) and the calibration Help files in the Shell program. Check each of the requested functions during installation and setup.

Note the data requested in each of the fields.

Mechanical stops should be set so the analyze position matches geometry on included data sheets, without touching any part in microscope.

The vacuum level of the microscope should be checked against the vacuum level before the detector was installed, per the installation procedure in the service manual.