لن ينظر باي استفسار يرد بعد تاريخ 14 /4/2013

item 1.ELECTRICAL BED, ICU, with OVERBED TABLE and BEDSIDE CABINET

Note: all three components must be from the same manufacturer

A: ICU Electric Bed. Qty(22)

  1. It must be stated clearly in the original catalogue that the offered item is intended for critical care.
  2. Heavy-duty construction suitable for rugged use.
  3. The dimensions of the sleeping surface platform to be a minimum ofapproximately (88 x 198) cm.
  4. The dimensions of the mattress to be a minimum ofapproximately (90 x 210) cm.
  5. Configuration: Four sections.
  6. Bed extension.
  7. Fully electrical
  8. Battery back-up.
  9. Built-in electronic weighing scale must either be standard or offered as an option.
  10. Siderails must incorporate embedded bed controls.
  11. Regular AC operation must be independent of battery status.
  12. Electrically controlled manoeuvres must include:
  1. Up/down (range must be clearly specified)
  2. Trendelenburg / Reverse trendelenburg
  3. Backrest
  4. Full Fowler position
  1. Easy release CPR control.
  2. Safe operational capacity minimum of 225 kg.
  3. Four heavy-duty medical grade electro-conductive, non-marking castors of not less than 15 cm diameter.
  4. Central brake system with steering function.
  5. Non-metallic side rails.
  6. Patient scale must be offered either as built-in or as optional.
  7. Radiolucent backrest.
  8. The bed should have separate controls for the patient and the caregiver.
  9. Removable non-metallic Headboards and Footboards.
  10. All available colours for headboards and footboards must be included in the offer.
  11. Four IV-pole receptacles to be located each side of the bed which can accommodate detachable IV-poles.
  12. Compatible IV poles with double hooks must be quoted in the offer and priced separately.
  13. Height adjustment from approximately 46 to approximately 85 cm or a wider range.
  14. Four protective revolving rubber bumpers located at each corner of the bed.
  15. Drainage bag hooks located at each side of the bed.
  16. Medical grade foam mattress with the following features:
  1. Fire retardant
  2. Anti microbial
  3. Solution resistant
  4. Complete with non-latex cover.
  1. Patient chart holder with heavy duty metallic paper clamp.
  2. Original Traction support to be listed and priced separately.
  3. Original Lifting-pole to be listed and priced separately
  4. All available options and accessories must be offered and priced separately.
  5. A list of standard accessories for the offered model (optional).

B: Bedside Cabinet. Qty(22)

  1. Standalone.
  2. Heavy-duty and robust.
  3. Complete with a minimum of one drawer.
  4. Lower cabinet.
  5. Mounted on four heavy-duty castors.
  6. Material, colour and design must be compatible with the beds and all available colours to be submitted with the offered.

C: Over-bed table. Qty(22)

  1. Standalone.
  2. Heavy-duty and robust.
  3. Mounted on four heavy-duty castors.
  4. Material, colour and design must be compatible with the beds and all available colours to be submitted with the offered.

item 2.Wall Suction Machine/low pressure Qty(30)

  1. Wall mounted original bottles holder to hold two autoclavable jars (≥ 1 litre each).
  2. Jars to be supplied with the unit.
  3. BOC straight connection for central suction lines with hose at last 2 meters long.

item 2.

item 3.Adult/ Paediatric Conventional ICU Ventilator with Consumables Qty(11)

A: ICU Ventilator:

General:

1.Adult and paediatric application.

2.Latest production that fulfils the required specs.

3.Unit must come without any type of a compressor.

4.Microprocessor controlled.

5.The unit must be mobile on a stand.

6.Ergonomically designed.

7.Castors of the mobile stand must be high quality medical grade.

8.Construction material must be robust and corrosion resistant.

9.Maintenance-free inspiratory/expiratory valves.

10.Inspiratory flow, L/min to cover the range 6-150

11.Surfaces must be flat, and easy to clean.

12.Tidal Volume: to cover the range 20-2000 ml

Ventilation: The following modes and functions are required:

13.Volume cycled.

14.CMV.

15.PEEP.

16.Inverse I:E capability.

17.Apnea back-up ventilation.

Monitored Parameters:

18.Expiratory Time

19.Expiratory Tidal volume.

20.Expiratory minute volume.

21.I:E ratio.

22.Total breath rate.

23.FiO2.

24.Pressure data: Plateau / Mean / Baseline.

Other features:

25.Audible and visual adjustable alarms for monitored parameters.

26.Complete with oxygen and medical air hoses with BS type probes.

27.Battery backup for both settings and operation.

Required Accessories:

28.Reusable adult and paediatric breathing circuit.

29.Articulated flexible arm for supporting the patient breathing circuit.

B Consumables for A:

  1. All the running requirements for system as recommended by the manufacturer the including flow-sensors, oxygen sensors, filters, any other consumables and maintenance kits sufficient for 360 patients at a rate of a different patient for every 24 hours for a period of 360 consecutive days.
  2. Requirements must be listed in a clear and itemized manner and prices fixed for 5 years from the date of installation.

item 1.

item 2.

item 3.

item 4.Infusion Pump Qty(30)

  1. Compact design.
  2. A flow rate of at least 1 to 999 ml/hr.
  3. The pump should be an open system (major set suppliers) while maintaining accuracy.
  4. Accuracy within 4% of displayed flow rate and volume.
  5. LCD display of flow and volume.
  6. Low battery warning.
  7. Infusion should be interrupted if battery is depleted.
  8. The pump should revert to keep vein open (KVO) when necessary.
  9. Adjustable alarm.
  10. The following alarms should be included:
  11. Up stream and down stream occlusion.
  12. Malfunction alarms.
  13. Air bubble detector.
  14. Low battery alarm.
  15. Easy to transport and to clamp to a pole.
  16. Visible and clearly identified controls.
  17. Spill proof design.

A list of standard accessories

item 5.Syringe pumpQty. (30 )

  1. Robust, compact design.
  2. Easy to transport and to clamp to a standard IV pole.
  3. Protected against fluid spillage.
  4. Microprocessor controlled unit.
  5. Start up self-test capability with an error code system.
  6. Hygienic, flat touch control panel.
  7. Adjustable flow rate (increments of 0.1 ml/hr.).
  8. To accept syringe sizes of up to 60 ml from several well-known manufacturers (open system).
  9. Settings and dose delivered should be retained to memory after the unit is turned off.
  10. Adjustable alarm volume control that cannot be completely turned off.
  11. Prime/purge control.
  12. Accuracy to within 5% of set flow rate.
  13. Clear and easy to read LCD display.
  14. Built-in high-capacity rechargeable battery.
  15. The following alarms should at least be included:
  16. High-pressure / occlusion.
  17. System malfunction.
  18. Empty reservoir (including alarms for infusion near end and infusion end).
  19. Syringe unlocked.
  20. Low battery alarm.
  1. A list of standard accessories for the offered model (optional).

item 6.Air mattressQty. ( 22 )

1.A complete Air Mattress for ICU patients use with air supply unit to prevent bed sores.

2.Full-length approx.200cm.

3.Silent running compact air supply unit that can be attached to the bed.

4.Quick and smooth inflation and deflation with intermittent action.

5.Patient weight capacity approximately 140 kg

6.Easy to clean smooth surface.

7.Transport mode, mattress remains inflated during patient transfer.

8.Inflation valve to be included.

9.A list of standard accessories for the offered model (optional).

item 7.Portable Feeding pump Qty. ( 2 )

A heavy-duty, portable, microprocessor controlled enteral feeding pump is required to accurately control the flow of liquid feeding solutions to adult patients through gastrostomy tubes. The unit must meet or surpass the following specifications:

1.Compact, light-weight design.

2.Durable, splash-proof exterior with easy to clean surfaces.

3.Visible and clearly identified controls.

4.Easy to transport and to clamp to standard IV poles.

5.Built-in rechargeable battery with operating time approximately ≥ 24hr on full charge.

6.Maximum flow rate approximately ≥ 400 ml/hr.

7.Accuracy: 10% or better.

8.Digital LCD display of rate, duration, delivered volume etc.

9.Parameters retained in memory: rate, volume, interval. Parameters should be retained for at least approximately 24 hours.

10.Audio-visual alarms:

  1. Occlusion.
  2. Malfunction alarms.
  3. Empty line.
  4. Low battery alarm.

11.Compatible sets, enteral feeding tubes, and feeding formulas must be available from several manufacturers.

12.A list of standard accessories for the offered model (optional).

B:

  1. All the running requirements for (A) including consumables and maintenance kits sufficient for 360 patients for each system over a period of one year at a rate of use per day.
  2. Requirements must be listed in a clear and itemized manner and prices fixed for 5 years from the date of installation.

item 8.AEDDefibrillatorQty. (13)

  1. Microprocessor controlled, biphasic technologyAEDDefibrillator based system.
  2. LCD screen ≥ 4.5 inches.
  3. Synchronizer capability.
  4. Selected and delivered energy display (Digital readout).
  5. Charge/ discharge controls on main unit and paddles.
  6. Charging time to maximum energy level ≤ 8 seconds.
  7. Standard adult paddles and built-in paediatric size.
  8. Full 3-lead ECG monitoring through leads or paddles.
  9. Variable heart rate Hi/ Low.
  10. Clear heart rate indicator.
  11. Variable beeper volume.
  12. Battery charging indicator and low battery indicator.
  13. Paddle contact integrity indicator.
  14. Large battery capacity (min. of 50 shocks at maximum setting).
  15. Self-test capability.
  16. Built-in annotative recorder with 25 mm-sec nominal speed.
  17. (3-5) Lead patient cable x2.
  18. Original or local trolley. If a local trolley is offered, a sample must be submitted at DRMS for evaluation purposes. The purchasing committee shall have the right to reject unsatisfactory samples, and subsequently, the whole offer.
  19. A list of standard accessories for the offered model (optional).

item 9.DRESSING CARTS Qty (20 )

This section describes the requirements for a Dressing Cart. In general, this Dressing Cart should be designed to transport dressing supplies and instruments. It shall have the following features:

  1. It shall not have side panels.
  2. Approximate overall dimensions approximately (W x D x H): 900 x 450 x 900 mm with 4-leg frame construction
  3. Removable waste container
  4. Two utility drawers
  5. It shall be made completely of chrome-nickel alloy steel / stainless steel or an equivalent non-metallic material with high quality finish to assure durability.
  6. It shall be corrosion resistant, disinfectant proof and easy to clean.
  7. The cart should have a minimum of two removable shelves
  8. It shall be mounted on 4 castors with the following features:
  9. Anti-static.
  10. Wear resistant.
  11. Approximate Diameter: approximately 75 mm.
  12. Swivel.

item 10.Medication CartsQty ( 16)

This section describes the requirements for medication carts. In general, medication carts should be designed with a locking mechanism for secure drug storage and should be expressly designed to aid in distributing medication on patient floors. Medication carts should provide an efficient, mobile means of organizing medication and supplies, such as cups and syringes, and transporting them to the patient.

  1. The medication cart shall be a unit-dose cart with individual drawers for each patient’s dosage.
  2. The medication cart should include a work surface.
  3. The medication cart should include storage wells and fenced sides.
  4. Approximately 30 drawers
  5. Large lower utility drawer(s)
  6. Dividers should be included.
  7. Master lock or locking mechanism that secures the entire cart and that must be unlocked before any individual compartment can be opened.

The following features and accessories should be included

  1. A push handle.
  2. Non-marking and durable bumpers
  3. Waste container
  4. Medicine cup dispenser
  5. The cart shall be mobile on swivel caster with locks

item 11.Portable patient monitor, preconfigured type Qty. ( 3 )

1.User friendly.

2.Portable type monitor

3.Heavy duty with robust design.

4.Two, Three lead ECG cable for each monitor.

5.Adult, paediatric and neonatal cuffs must be quoted for

6.Reusable adult and neonate SPO2 sensors for each monitor.

7.Easy to mount on anaesthesia trolley upper shelf.

8.All mounting options to be offers including wall mount and bench top.

9.All leads, cables, sensors and accessories needed for function must be stated clearly and priced separately.

10.Standard accessory sets must be itemized.

11.Adjustable ECG gain.

12.Baseline adjustment capability.

13.High resolution display with sweeping motion.

14.All data shall be in real time (waveform and numeric).

15.Large LCD screen (not less than 10.4 inches).

16.Built-in rechargeable battery.

17.Where applicable, any software shall be of the latest version, and any service / settings password must be disclosed and perpetual.

18.Vendors to list all prices of optional software and hardware for offered system

19.A list of standard accessories for the offered model (optional).

item 4.

item 5.

item 6.

item 7.

item 8.

item 9.

item 10.

item 11.

item 12.Heating Mattress Qty.(6)

  1. Microprocessor controlled unit.
  2. Mobile unit with high quality castors.
  3. Robust design with heavy-duty construction.
  4. Air type with smooth quiet operation.
  5. Temperature range to be indicated clearly. Temperature range will be an important evaluation criterion in decision making.
  6. Digital temperature read-out for both set and measured values.
  7. Highest level of safety with alarm indicators.
  8. Mattresses to be made of high quality material that is easy to clean and disinfect.

item 13.TABLES, OPERATING, GENERAL, ELECTRICAL Qty ( 6 )

  1. This section describes the requirements for a general operating table. It shall meet or exceed the following components/features:
  2. Electric, mobile operating table column.
  3. It shall have integrated, maintenance-free, rechargeable batteries with integrated batterycharger.
  4. 4-section (minimum) operating tabletop for general surgeries.
  5. Full function, corded hand pendant control (wireless [IR] remote control to be offered as anoption).
  6. Approximate body weight load capacity: 250 Kg or better.
  7. The table shall meet or exceed the following specifications:
  8. Approximate total length of the operating tabletop with all of its section: 2000 mm.
  9. Approximate tabletop width: 500 mm (550 mm with the accessory side rails).
  10. Height: 700 – 1000 mm.
  11. Trendelenburg / reverse Trendelenburg: 25°.
  12. Lateral tilt: left / right: 15°.
  13. Back plate: up / down: 40° / 40°.
  14. Leg plate downward: 90°.
  15. Side rails for attachment of accessories.
  16. Anesthesia screen with attachment clamp.
  17. Left and right arm rests with attachment clamps
  18. X-ray film arm.
  19. Complete orthopedic and traction attachments and accessories.
  20. Holder/stand for storing accessories.
  21. All options and accessories shall be listed and quoted separately.
  22. Accessories MUST be priced separately and MUST include at least the following specialities: General, orthopaedics, gynaecology, neurology, and urology.

item 14.Anaesthesia Machine with Ventilator. Qty. ( 6 )

The Unit

  1. High quality, high standard and heavy duty anaesthesia machine suitable for adults, paediatrics and neonates.
  2. Original cart to include:
  3. An upper shelf
  4. A working tray.
  5. Equipped with anti-static castors with brakes.
  6. The system must include a ventilator suitable for adults, paediatrics and neonates.
  7. Ventilator brand must be the same as the anaesthesia unit and manufactured by the same company.
  8. The unit must include all necessary regulators and flow-meters
  9. Flow meters to include O2, N2O and medical air.
  10. Flow meters type must be specified.
  11. The system must contain a suction unit.
  12. Circulation system must be compact, sterilizable and changeable from re-breathing (closed) to non re-breathing (open).
  13. Central gas supply (Pipeline gas inlets) a number of three inlets:
  1. O2
  2. N2O
  3. Medical Air (4 bar)

all with international colour code and pressure gauge according to NIST system and equipped with B.O.C probes for connection to central gas outlets.

  1. Gas cylinder yokes: (O2 and N2O) with clear gas name marking and pressure gauge equipped with international pin-index system.

Vaporizers

  1. Vaporizes to be made by the same manufacturer of the offered anaesthesia machine.
  2. Vaporizers to be selectatec type for fast and easy coupling.
  3. At least two vaporizers can be mounted on the machine at the same time.
  4. Machine must be equipped with modern and new inhalation agents’ vaporizers.
  5. Offered Vaporizes must include halothane, isoflurane and sevofluorane.
  6. Vaporizers must be maintenance free.
  7. Halothane and isoflurane vaporizers must be with key filling, Refilling adapters (keys) must be provided
  8. Sevofluorane vaporizers must be with quick fill (easy fill) system. Refilling adapters must be provided.
  9. O2/N2O safety control:
  10. Not less than 25% O2.
  11. Oxygen failure protection device (hypoxic guarding mechanism) with alarm.
  12. Oxygen-flush with a flow-rate not less than 30 litres/minute.
  13. Gas selector switch for 'no gases / full gases' (on/off gases switch).
  14. Unit to include a CO2 absorber with a capacity of approximately 0.5 kg or more.

Ventilator

  1. A modern type ventilator with an integrated a microprocessor.
  2. Fully coloured LCD display.
  3. For ventilators using bellows, bellows type must be not other than ascending or descending.
  4. Ventilation modes to be selectable and to include at least:
  5. Manual/Spontaneous
  6. Pressure control
  7. Volume control
  8. PSV
  9. SIM
  10. Parameter readings of ventilator monitor to include not less than:
  11. Airway pressure.
  12. Continuing pressure alarm.
  13. Expiratory volume/flow.
  14. Low pressure/apnea.
  15. Ventilator monitor graphic waveforms: to include volume and pressure.
  16. Type of sensor for Exp volume/ flow to be not other than hot wire, galvanic cell or diff pressure.
  17. Coloured display monitor of not less than 8.4 inches in size.

Ventilators Technical Data

  1. Respiration frequency: 5 - 60 Bpm or better.
  2. Inspiration Flow: 5-70 l/min or a wider range.
  3. Tidal Volume: 20-1400 ml or a wider range.
  4. Maximum minute volume to be not less than 25 l / minute.
  5. I: E Ratio: 2:1 to 1:4 or better.
  6. Inspiratory pause: to include not less than the range 0-50%.
  7. Pressure-limit: 15 – 70 cmH2O or better.
  8. Ventilator to include an internal rechargeable battery of time per charge not less than 30 minutes.

Alarms

  1. Audible and visual alarms to include :
  2. Power failure alarm.
  3. Low/high pressure.
  4. Low O2 concentration.
  5. Low /high volume.
  6. O2 inlet supply failure.
  7. Alarm interruption not more than one minute reset function.

Accessories

  1. Reusable Open-type patient circuits for adults and paediatrics (complete assembly).
  2. Reusable Closed-type patient circuit suitable for adults and paediatrics.
  3. Offer must include a full gasmonitor with BIS module priced separately as an option.
  4. Any other accessories must be listed and priced separately
  5. Complete with a full gas monitor either integrated or external.

A list of standard accessorie