Manual of RAIRS Mattson Instrument
(Cygnus 100 Rev.7)
Guangming Liu
Postdoctoral researcher in Zaera's lab
Zaera's Laboratory
Department of Chemistry
University of California Riverside
Jan 26, 2004
1. Optical Route Alignment:
- Press a sample as bare clean Au film onto sample holder—home made reflectance unit. This position of sample can be adjusted in 3 directions (X, Y, Z),and it has already been optimized with the maximum signal intensity, try to keep the same position for every measurement. If possible, suggest to change another reflectance unit which can be adjusted with the incidence angle (the price may be around thousands of bucks).
- Fill sufficient liquid nitrogen into the MCT detector. Turn the MCT preamplifier on. Remember fill this tank at least 1 hour earlier before each measurement. For better performance, pump this detector each month if possible.
- Set the parameters of FTIR for RAIRS measurement in Winfirst program. Detailed Instructions for Winfirst is written in its manual. Typically I set the parameters as follows: gain = 4, resolution = 4, Iris = 40%, scan = 4096.
- Display interferogram on the oscilloscope connected preamplifier of MCT detector.
- Move mirrors for getting maximum intensity of interferogram as possible as you can. Thermal sensitive film (IR card) can be used to track the IR beam and also focus the beam on sample surface.
- Check interferogram for both p and s polarizations.
2. Optimization of RAIRS Setting up
I. The gain of the instrument can be set to 1, 2, 4. The limit of the centerburst in the interferogram is 9.999 V, max= 4.999 V, min= -4.999 V. Check the detector was saturated or not when the gain set to high values. Using a stable sample as M192/Au as standard sample, checked the different gain values influence. The following parameters were set to see how the spectra changing with the different gain values, scan= 1024, Iris= 40 %, resolution= 2.0 cm-1.
Gain = 1 noise: RMS = 0.0225 (2150-2200 cm-1), p2p = 4.869 V, Signal = 15.582 (970-1040 cm-1), S/N = 692.53;
Gain = 2 noise: RMS = 0.0424 (2150-2200 cm-1), p2p = 6.301 V, Signal = 18.393 (970-1040 cm-1), S/N = 433.80;
Gain = 4 noise: RMS = 0.0410 (2150-2200 cm-1), p2p = 9.999 V, Signal = 22.358 (970-1040 cm-1), S/N = 545.32.
From these data, the best S/N can be achieved at gain= 1, but sometime if you want to get strong signal intensity, set Gain to 4 is also necessary.
II. Using bare Au film as sample and keep the beam completely open, took back to back spectra of the same sample and ratioing them, the measured RMS region was from 2100 to 2150 cm-1, the gain = 2, resolution = 4, Iris = 40 %.
Scan= 1 RMS= 0.0861
Scan= 4 RMS= 0.0399
Scan= 16 RMS= 0.0226
Scan= 64 RMS= 0.0111
Scan= 256 RMS= 0.0045
Scan= 1024 RMS= 0.0038
The noise level approximately goes down by a factor of 2 with increasing the scan number from 1 to 256 except from 256 to 1024.
III. Choose a fixed scan number of 256 (Iris= 40%, Gain= 2, Resolution= 4), and run back to back spectra at the different mirror speeds. The velocity of mirror fwd and rev can be selected as the following values: 80, 40, 20, 10, 5.7, 3.6, 2.6. A series of experiments were done by choosing the value of 40, 20, 10 and 5.7, repectively. However, the speed of 80 cannot be properly run by the Winfirst program, there is an error signal of “laser fault” appeared in the display panel.
Speed set to 40 time= 2.13 min, RMS= 0.0059
Speed set to 20 time= 4.05 min, RMS= 0.0067
Speed set to 10 time= 7.50 min, RMS= 0.0044
Speed set to 5.7 time= 12.58 min, RMS= 0.0041
So the mirror speed of 40 is the best speed which can get with the shortest acquisition time and the appropriate noise.
IV. To maximize the signal intensity in the interferogram by optimizing the optical route, we can slightly change the micrometer settings in the fixed comer cube mount. The horizontal micrometer was kept at 3.525 without change, vertical micrometer was decreased from 4.059 to 4.025. At this position, the centerburst in the interferogram reached the maximum intensity although the bottom part was a little clipping.
V. Took spectra versus opening of the iris by using M192/Au polymer as standard sample. Fixed the following parameters: Resolution = 4.0, Gain = 4.0, Scan = 1024, Mirror speed: fwd = rev = 40.0.
Iris = 100% RMS = 0.0574 (2150-2200 cm-1), signal = 18.77 (970-1040 cm-1), centerburst: p2p = 9.842 V (Min= -4.842 V, Max= 4.999 V); S/N= 327.00
Iris = 80% RMS = 0.0574 (2150-2200 cm-1), signal = 18.76 (970-1040 cm-1), centerburst: p2p = 9.771 V (Min= -4.772 V, Max= 4.999 V); S/N= 326.83
Iris = 60% RMS = 0.0560 (2150-2200 cm-1), signal = 18.66 (970-1040 cm-1), centerburst: p2p = 9.652 V (Min= -4.660 V, Max= 4.999 V); S/N= 333.21
Iris = 40% RMS = 0.0366 (2150-2200 cm-1), signal = 17.52 (970-1040 cm-1), centerburst: p2p = 8.408 V (Min= -3.408 V, Max= 4.999 V); S/N= 478.69
Iris = 30% RMS = 0.0377 (2150-2200 cm-1), signal = 16.804 (970-1040 cm-1), centerburst: p2p = 8.072 V (Min= -3.072 V, Max= 4.999 V); S/N= 445.73
Iris = 20% RMS = 0.0351 (2150-2200 cm-1), signal = 15.1465 (970-1040 cm-1), centerburst: p2p = 7.658 V (Min= -2.658 V, Max= 4.999 V); S/N= 431.52
Iris = 10% RMS = 0.0248 (2150-2200 cm-1), signal = 11.63 (970-1040 cm-1), centerburst: p2p = 5.097 V (Min= -1.826 V, Max= 3.276 V); S/N= 468.95
Both of the signal intensity and the noise level increased with the opening of the iris size, the best S/N was found at the iris size of 40% opening.
VI. Using mask to define beam at different size of 1.2, 1.0, 0.5 (diameter), checked the size of the centerburst, the noise level, and the actual size of the strongest peaks, for each beam size spectra were measured by using two iris sizes of 40% and 100%, respectively.
A. Beam size= 1.2 ,
Iris= 40% RMS= 0.0322 (2150-2200 cm-1), signal= 6.70 (970-1040 cm-1), centerburst: p2p=6.656 V (Min= -2.355 V, Max= 4.340 V)
Iris= 100% RMS= 0.0347 (2150-2200 cm-1), signal= 6.75 (970-1040 cm-1), centerburst: p2p=7.187 V (Min= -2.522 V, Max= 4.750 V)
B. Beam size= 1.0 ,
Iris= 40% RMS= 0.0256 (2150-2200 cm-1), signal= 4.15 (970-1040 cm-1), centerburst: p2p=5.797 V (Min= -2.149 V, Max= 3.685 V)
Iris= 100% RMS= 0.0260 (2150-2200 cm-1), signal= 4.18 (970-1040 cm-1), centerburst: p2p=6.333 V (Min= -2.002 V, Max= 4.335 V)
C. Beam size= 0.5 ,
Iris= 40% RMS= 0.0082 (2150-2200 cm-1), signal= 1.0512 (970-1040 cm-1), centerburst: p2p=2.569 V (Min= -0.925 V, Max= 1.660 V)
Iris= 100% RMS= 0.0090 (2150-2200 cm-1), signal= 1.0521 (970-1040 cm-1), centerburst: p2p=2.568 V (Min= -0.936 V, Max= 1.645 V)
D. Beam size= 3.0 (without mask),
Iris= 40% RMS= 0.0496 (2150-2200 cm-1), signal= 18.61 (970-1040 cm-1), centerburst: p2p=8.880 V (Min= -3.880 V, Max= 4.999 V)
Iris= 100% RMS= 0.0570 (2150-2200 cm-1), signal= 18.60 (970-1040 cm-1), centerburst: p2p=4.176 V (Min= -4.187 V, Max= 4.999 V)
VII. Test the quality of the spectra taken at different resolutions. There are several options for the resolution of this instrument, such as 0.5, 1.0, 2.0, 4.0, 8.0, 16.0, 32.0, 64.0, 128.0, cm-1, higher resolution does increase the sensitivity of the spectra, make it easier to find the peak we want, however, it also means more acquisition time and much more noise. Using the stable M192/Au as standard sample, checked the spectra quality with the different resolution (1,2, 4, 8 cm-1), as shown in the figure below.
Resolution = 8 cm-1 t = 5.12 min, RMS = 0.0252 (2150-2200 cm-1), signal = 18.865 (970-1040 cm-1);
Resolution = 4 cm-1 t = 8.51 min, RMS = 0.0332 (2150-2200 cm-1), signal = 21.93 (970-1040 cm-1);
Resolution = 2 cm-1 t = 15.36 min, RMS = 0.0529 (2150-2200 cm-1), signal = 22.857 (970-1040 cm-1);
Resolution = 1 cm-1 t = 29.01 min, RMS = 0.0536 (2150-2200 cm-1), signal = 22.77 (970-1040 cm-1);
So the resolution 2 or 4 cm-1 is the desirable resolution we can select in the experiments.
VIII. Optimize the incidence angle:
The incidence angle was optimized to set around 83. If the incident angle was set near 85, the signal intensity was very small at this position; most of the beam was directly transmitted while not reflected to the mirror. This means much signal intensity was lost, so it is very difficult to set the angle over 85, 83 is probably the most incident angle we can reach with high intensity.
3. Measurements
I. Preparation of the Pr SAMS: SAMs of porphyrins were prepared by immersing the freshly deposited Au coated substrates into 1 mM deoxygenated solutions of the desired porphyrin in CH2Cl2 at room temperature for 1-7 days (the deposition time was varied to reach the desired SAM surface concentration and to ensure the formation of well organized SAMs). Upon removal from the solution, the samples were then rinsed thoroughly with CH2Cl2 and ethanol, to remove physisorbed species and blown dry with a steam of argon. They were used immediately for spectral measurements.
II. Before testing, flow the purging gas liquid nitrogen with appropriate time, the longer time maybe reduce the water noise, but the integrity of the sample would be less, so make sure the stability of your sample, then decide how much time you may require (typically 1-2 hours for porphyrin molecules).
III. Press the sample to the optimized positionon the sample holder surface, try to keep the sample at the same position at each measurement.
IV. After a certain time, start background spectral measurement at s-polarizer position. After that, immediately start-acquiring p polarized spectra.
V. After the measurement, the sample should be pulled up from the sample holder and make ready for the next sample.
4. Maintenance
Checking Laser Signals on Rev. 7 Polaris Benches
- Disconnect the voice coil drive cable from P 7 (the black an white wire pair) located just below the heat sinks on the center right side of the main control board.
- Record the micrometer settings for the fixed corner cube mount. These numbers will be necessary if the micrometers are changed inadvertently during this procedure.
Horizontal Micrometer 3.525 Vertical Micrometer 4.059
3. Connect the two oscilloscope input probes to TP 1 and TP 21 on the main control board. The test pints are located 2 inches to the left of the beamsplitter under the front lip of the touch panel display. Connect the grounds to the ground loop near P 4. Set the oscilloscope as follows:
Channel A input Connect to TP 1
Channel B input Connect to TP 21
Ground Lead Connect to ground pin
Channels A & B input coupling: Set both to ac
Vertical gain Set to 1.0 volt/cm
Horizontal gain Set to 20 Sec/cm
Trigger source Channel A
Trigger mode Auto
Adjust the position, brightness, etc., controls until the signals are displayed. They should be DC signals with varying voltages (close to noise until the mirror is moved by hand).
Adjust the vertical position controls on the oscilloscope so that the grounded signal from TP 1 and TP 21 are overlapping on the center of the screen. Set the scope inputs back to AC. Move the mirror and observe the phase difference of the signals. The phase of the signal from TP 21 should change; (The signal from TP 1 should not change because it is the trigger source).
Move the linear bearing slowly back and forth at about 1 to 2 cm/sec. The two signals being displayed should appear as sine waves with a frequency, which varies with the velocity of the mirror. Set the trigger level control on the scope to stabilize the display.
The peak-to-peak voltage of the signals should be between 2 and 4 volts. Carefully adjust the two micrometer handles while continuing to move the bearing. Set the micrometers to the point, which produces the maximum laser signal voltage. If the sine waves do not appear when you move the mirror, check the micrometer setting by referring to the numbers you recorded above.
Adjusting the amplitude of the laser signals:
If the signal intensity is low, check where the laser is striking the diode. The brightest dot should strike the center of the diode window. If not, loosen the diode arm with a 3/32” hex wrench, adjust and re-tighten. You can remove the diode from the holder and place a piece of tape over the hole to see the laser’s path. At the same time check the optical quality of the beamsplitter, is it starting to fog?
If you cannot increase the laser-signal by optical alignment then you can adjust the amplification of the signals on the control board. The gain pots in the circuit are R 26 for TP 21 and R 25 for TP 1. If adjusting gain, makes sure that the signal is not clipping (flattening off the bottom of the signals). It is better to have a 1 to 2 volt signal, than a signal that is clipping.
The proper phase of the two signals is shown. The TP 21 signal should lead the TP1 signal by 90 degrees when the mirror is moving towards the Beamsplitter, and should lag by 90 degrees when the mirror is moving away from the Beamsplitter. The phase angle should be 90 degrees +/- 10 degrees. If the phase is inaccurate, the reliability of the scanning of the mirror may be affected, or the mirror may not scan at all.
5. Typical conditions of FTIR-RAIRS
Mattson Cygnus 100
Scan speed40
Number of scan1024 or 4096
Aperture25-40%
Amplifier Gain4
Othersfactory setting
Interferogram intensity 3-6 V peak-peak
1
1