Schedule of Accreditation
Laboratories of Egyptian Tank Plant (ETP 200)
Industrial Zone, Abo Zaabal, Kalyobia, Egypt
Valid to : March 31, 2016 / Revision No. (2): January 15, 2015 / Issue No. (2): February 08, 2013 / Schedule No.: 297004BBrief Description of
Measurement and Equipment Used
/Calibration and Measurement Capability*(±)
/Range
/Measured Quantity
Direct measurement / ElectricityFluke calibrator
Model: 5720A, SN. 5530023 / إجراء معايرة رقم (م 8) / 2.7 µV / up to 220 mV / Calibration of DC voltage measuring equipment
16 mV /
220 mV - 2.2 V
46 mV /2.2 - 11 V
0.09 mV /11 - 22 V
1.3 mV /22 - 220 V
13.9 mV /220 - 1100 V
Fluke multimeterModel: 8508A, SN. 46738 / إجراء معايرة رقم (ق د 2) / 2 µV / up to 220 mV / Calibration of DC voltage sources
30 µV / 220 mV - 2.2 V
0.083 mV / 2.2 - 11 V
0.12 mV / 11 - 22 V
1.3 mV / 22 - 220 V
8.8 mV / 220 - 1100 V
Fluke calibrator
Model: 5720A, SN. 5530023 / إجراء معايرة رقم (م 8) / 220 mV - 2.2 V /
Calibration of AC voltage measuring equipment
0.27 mV / 20 - 40 Hz0.13 mV / 40 Hz - 20 kHz
0.21 mV / 20 - 50 kHz
0.32 mV / 50 - 100 kHz
2.8 mV / 100 - 500 kHz
2.2 - 22 V
2.6 mV / 20 - 40 Hz
1.3 mV / 40 Hz - 20 kHz
2.1 mV / 20 - 50 kHz
2.8 mV / 50 - 100 kHz
28 mV / 100 - 500 kHz
22 - 220 V
24 mV / 20 - 40 Hz
14 mV / 40 Hz - 20 kHz
21 mV / 20 - 50 kHz
40 mV / 50 - 100 kHz
Fluke calibrator
Model: 5720A, SN. 5530023 / إجراء معايرة رقم (م 8) / 220 - 1100 V /
Calibration of AC voltage measuring equipment
107 mV / 20 - 60 Hz107 mV / 60 Hz - 400 kHz
107 mV / 400 - 1000 kHz
Fluke multimeter
Model: 8508A, SN. 46738 / إجراء معايرة رقم (ق د 2) / up to 220 mV /
Calibration of AC voltage sources
0.073 mV / 10 - 40 Hz0.08 mV / 40 Hz - 20 kHz
0.16 mV / 20 - 50 kHz
0.36 mV / 50 - 100 kHz
220 mV to 2.2 V
0.28 mV / 20 - 40 Hz
0.21 mV / 40 Hz - 20 kHz
0.51 mV / 20 - 50 kHz
1.4 mV / 50 - 100 kHz
2.2- 22 V
2.8 mV / 20 - 40 Hz
2.2 mV / 40 Hz - 20 kHz
5.2 mV / 20 - 50 kHz
14 mV / 50 - 100 kHz
22 - 220 V
28 mV / 20 - 40 Hz
16 mV / 40 Hz - 20 kHz
22 mV / 20 - 50 kHz
40mV / 50 - 100 kHz
220 - 1100 V
0.15 V / 20 - 60 Hz
0.14 V / 60 Hz - 400 kHz
0.11V / 400 - 1000 kHz
Fluke calibrator
Model: 5720A, SN. 5530023 / إجراء معايرة رقم (م 8) / 0.06 µA / up to 220 µA / Calibration of DC current measuring equipment
0.89 mA /
2.2 - 22 mA
8.12 mA /22 - 220 mA
0.23 mA /220 mA - 2.2 A
Fluke multimeterModel: 8508A, SN. 46738 / إجراء معايرة رقم (ق د 2) /
10.2 µA
/up to 220 µA
/ Calibration of DC current sources108 µA
/220 µA - 2.2 mA
1.2mA
/2.2 - 22 mA
12 mA
/22 - 220 mA
0.2 A
/220 mA - 2.2 A
Fluke calibratorModel: 5720A, SN. 5530023 / إجراء معايرة رقم (م 8) /
uP to 22 mA
/ Calibration of AC current measuring equipment0.24 mA
/40 Hz - 1 kHz
0.022 mA
/1 - 5 kHz
0.03 mA
/5 - 10 kHz
22 - 220 mA
0.03 mA
/40 Hz - 1 kHz
0.05 mA
/1 - 5 kHz
0.27 mA
/5 - 10 kHz
220 mA - 2.2 A
0.5 mA
/40 Hz - 1 kHz
1.14 mA
/1 - 5 kHz
17.78 mA
/5 - 10 kHz
Fluke multimeterModel: 8508A, SN. 46738 / إجراء معايرة رقم (ق د 2) /
up to 22 mA
/ Calibration of AC current sources0.026 mA / 40 Hz - 1kHz
0.024 mA / 1 - 5kHz
0.036 mA / 5 – 10 kH
22 mA – 220 mA
0.03 mA / 40 Hz - 1kHz0.05 mA / 1 - 5kHz
Fluke multimeter
Model: 8508A, SN. 46738 / إجراء معايرة رقم (ق د 2) / 0.3 mA / 5 – 10 kHz / Calibration of AC current sources
220 mA - 2.2 A
0.5 mA / 40 Hz - 1kHz1.1 mA / 1 - 5kHz
Fluke calibrator
Model: 5720A, SN. 5530023 / إجراء معايرة رقم (م 8) / 0.12 mΩ / 2 Ω / Calibration of DC resistance measuring equipment
0.2 mΩ / 20 Ω
1.3 mΩ / 200 Ω
9 mΩ / 2 kΩ
90 mΩ / 20 kΩ
1.6 Ω / 200 kΩ
28 Ω / 2 MΩ
48 Ω / 20 MΩ
1 kΩ / 100 MΩ
Fluke multimeter
Model: 8508A, SN. 46738 / إجراء معايرة رقم (ق د 2) / 0.12mΩ / 2 Ω /
Calibration of DC resistance sources
19 mΩ / 20 Ω19 mΩ / 200 Ω
9 Ω / 2 kΩ
17 Ω / 20 kΩ
18 Ω / 200 kΩ
28kΩ / 2 MΩ
480 kΩ / 20 MΩ
26 kΩ / 200 MΩ
1.18MΩ / 2000 MΩ
Digital multimeter
Model: 3114A, SN. 13436
Fluke high voltage probe
Model: 80K-40, SN. 5605067 / إجراء معايرة رقم (م 9) / 1.3 % of Reading / 0 - 40 kV / Calibration of DC high voltage sources
5 % of Reading / 0 - 40 kV / Calibration of AC high voltage sources (50/60 Hz)
Digital multimeter
Model: 3114A , SN. 13436
Shunt resistance
Model: WT1000-100, SN. 0001 / إجراء معايرة رقم (م 9) / 2.5 % of Reading / 0 - 1000 A / Calibration of DC high current sources
3.5 % of Reading / 0 - 1000 A / Calibration of AC high current sources (50/60 Hz)
Frequency
Programmable calibration generatorModel: TM5003/CG5011, SN. B011630 / إجراء معايرة رقم (م 10) / 4x10 -7 Hz / 1 kHz - 1 GHz /
Calibration of frequency
Electronic stroboscopeModel: 1538-A, SN. 76669 / إجراء معايرة رقم (م 12) / 0.38 % of reading / 1000 - 150000 rpm / Calibration of tachometers
Pressure
Pressure transducers
Model: 525A-PV4, SN. 591203
Model: 525A-P05, SN. 591291
Model: 525A-P07, SN. 590706
Model: 525A-P29, SN. 591293
Model: 700P31, SN. 89403106
Dead weight tester
Model: RPM4-E-DWT, SN. 1511 / DKD-R 6-1:2003 / 0.03 psi
0.5 psi
2 psi / -15 to 0 psi
0 to 500 psi
500 to 1000 psi
/ Calibration of pneumatic pressure gauge using temperature/pressure calibrator(Permanent)
4.4 psi10 psi
/ 0 to 3000 psi3000 to 10000 psi
/ Calibration of hydraulic pressure gauge using temperature/pressure calibrator(Permanent)
BalanceClass I weight sets
SN. 96-092542, CN. 21-900-0004-001
CN. 21-900-0005-001, SN. 69089
SN. 69090, SN. 69091, SN. 69092
SN. 69093, CN. 21-900-0006-001
CN. 21-900-0007-001 / OIML R 76-1:2006 / 0.52 mg / 10 mg - 220 g / Calibration of electronic balances(Permanent & on-site)
0.84 mg / 220 - 500 g
2.9 mg / 500 g - 2 kg
6.6 mg / 2 kg - 5 kg
15 mg / 5 kg - 10 kg
0.11 g / 10 kg - 75 kg
5.8 g / 75 kg - 250 kg
Brief Description of
Measurement and Equipment Used
/Calibration and Measurement Capability* (±)
/Range
/Measured Quantity
DimensionGauge blocks comparator
Model: 130B-24, SN. 185B92
Gauge blocks set
Model: 516-937-71, SN. 0103074
Model: 350002, SN. 060817 / ISO 3650:1998 / 0.2 µm / up to 100 mm / Calibration of gauge blocks
Gauge blocks set
Model: 516-942, SN. 751260
Individual gauge blocks
CN. 21-400-1007-001
Electronic indicator
Model: M406, SN. 100045 / ES 313:2008 / 7 µm / 0 - 150 mm / Calibration of calipers
8 µm / 0 - 300 mm
10 µm / 0 - 450 mm
12 µm / 0 - 600 mm
Gauge blocks set
Model: 516-942, SN. 751260
Measuring machine
Model: 305M, SN. 2625 / ES 617:2008 / 5 µm / up to 10 mm / Calibration of dial indicator
Measuring machine
Model: 305M, SN. 2625
with Standard setting ring
SN. 41179, 41071 / DIN 13 part 14:1982
DIN 13 part 15:1982 / 3 µm / up to 300 mm / Calibration of plug & ring gauge
3 µm / up to 200 mm / Calibration of thread plug gauge
3 µm / up to 125 mm / Calibration of thread ring gauge
Gauge blocks set
Model: 516-942, SN. 751260
Precision spirit level
Model: 679, CN. 21-120-0007-001 / ES 1567:2003 / 5 µm / 150 × 250 mm / Calibration of general measurement microscope
Gauge blocks set
Model: 516-942, SN. 751260
Individual gauge blocks
CN. 21-400-1007-001
Surface plate
Model: Grade AA-lab, SN. 900617
Electronic indicator
Model: M406, SN. 100045
Steel square
CD. 21-820-0005-004 / ES 911:2003 / 10 µm / up to 600 mm / Calibration of height gauge
Dimension
Laser measurement system
Model: 55292A, SN. US47040327
Model: 5518A, SN. 2532A02180
with Hamar laser
Model: L-110, SN. 94112438
Edge Straight
CD. 21-880-0043-0005 / DIN 876 Part 1:1984
DIN 876 Part 2:1984 / 1.1 µm / from 250 × 400 to 2000 mm / Calibration of surface plate
Gauge blocks set
Model: 516-942, SN. 751260
Individual gauge blocks
CN. 21-400-1007-001
Surface plate
Model: Grade AA-lab, SN. 900617
Measuring machine
Model: 305M, SN. 2625
with Standard setting ring
SN. 41179, 41071
Electronic indicator
Model: M406, SN. 100045
Opticalflats
CD.21-640-0009-0001 / ES 163:2008
ES 528:2008 / 4 µm / 0 - 25 mm / Calibration of micrometer outside
6 µm / 25 - 50 mm
7 µm / 50 - 75 mm
8 µm / 75 - 100 mm
9 µm / 100 - 125 mm
10 µm / 125 - 150 mm
12 µm / 150 - 600 mm
12 µm / up to 600 mm / Calibration of inside micrometer
10 µm / up to 200 mm / Calibration of depth micrometer
Standard steel square
CD. 21-820-0005-004
Gauge blocks set
SN. 751260 / DIN 875:1981 / 5 µm / up to 700 mm / Calibration of steel square angle
Laser measurement system
SN. 2532A02180
with Hamar laser
SN. 94112438 / ASME B 89.4.1:1997 / 5 µm / up to 6000 mm / Calibration of coordinate measuring machine
Temperature
Platinum resistance thermometerModel: 5628, SN. 0461
(SPRT) with resistance bridge
Model: 1575, SN. 4B054
Water bath
Model: Neslab, SN. 1314135-1
Oil bath
Model: Fluke6331, SN. 2001811
Calibration bath
Model: 7380, SN. A56258
Sand bath
Model: FSB-3, CN. 21-181-0020-001 / ES 400:2005
إجراء معايرة رقم (م 1) / 0.14oC / - 40 to 20 oC / Calibration of glass thermometer
(permanent)
0.13 oC / 20 to 100 oC
0.13 oC / 100 to 220 oC
0.4 oC / 220 to 400 oC
Standard thermocouple type S
Model: 5650, SN. 9187
Platinum resistance thermometer
Model: 5628, SN. 0461
Digital thermometer
Model: 820, SN. T-233272
Model: 54II, SN. 14690038
Water bath
Model: Neslab, SN. 1314135-1
Oil bath
Model: Fluke6331, SN. 2001811 / ES 962:1993
إجراء معايرة رقم (م 5) / 0.71 oC / 30 to 1200 oC / Calibration of thermocouples (K,T,J,S,R,E) (permanent)
Temperature calibrator
Model: 525, SN. 9019004
Model: 724, SN. 2007023
Digital thermometer
Model: 820, SN. T-234784, T-233272, T-235237 / إجراء معايرة رقم (م 3) / 0.07oC / - 200 to 1100 oC / Calibration of digital thermocouple indicator
(permanent &on-site)
Temperature calibrator
Model: 525, SN. 9019004
Model: 724, SN. 2007023
Digital thermometer
Model: 820, SN. T-234784, T-233272, T-235237
Digital themometer
Model: 54II, SN. 14690038 / إجراء معايرة رقم (م 2) / 0.5 oC / 50 to 1100 oC / Calibration of oven indicator (permanent &on-site)
إجراء معايرة رقم (م 4) / 0.3 oC / 20 to 90 oC / Calibration of water bath indicator (permanent &on-site)
0.3 oC / 85 to 205 oC / Calibration of oil bath indicator (permanent &on-site)
0.4 oC
/200 to 580 oC
/ Calibration of sand bath indicator (permanent &on-site)Fluke calibrator
Model: 5720A, SN. 5530023
Reference multimeter
Model: 8508A, SN. 46738 / إجراء معايرة رقم (م 6) / 0.02 oC
0.03 oC / Source
Measure / Calibration of temperature calibrator (permanent)
Standard platinum resistance thermometer (SPRT)
Model. 5628, SN. 0461
Resistance bridge
Model. 1575, SN. 4B054
Reference multimeter
Model.8508A, SN. 46738
Water bath
Model: Neslab, SN. 1314135-1
Oil bath temperature
Model: Fluke6331, SN. 2001811
Sand bath temperature
Model: FSB-3, CN. 21-181-0020-001
Low bath temperature
Model. 7380, SN. A56258 / إجراء معايرة رقم (م 30) / 0.2 oC / 50 to 400 oC / Calibration of platinum resistance thermometer with digital indicator
Brief Description of
Measurement and Equipment Used
/Calibration and Measurement Capability* (±)
/Range
/Measured Quantity
Force
Load cellModel: 41/9267-01-02, SN. 998919
Model: KTN-D, SN. 54993 / ASTM E4:2007 / 1.4 % / 7 - 60 ton / Calibration of universal testing machine (tension & compression)
Reference reckwell hardness blocks
Model: 62.32HRC, SN. IN602N
Model: 27.60HRC, SN. IN25629M
Model: 47.68HRC, SN. IN45801M / ISO 6508-2:2005 / 1 HRC / 20 - 60 HRC / Calibration of rockwell hardness tester
Reference brinell hardness blocks
Model: 401.6 HBW, SN. EP08108266
Model: 404.5 HBW, SN. EP08105104
Model: 190.7 HBW, SN. EP08108280
Model: 137.2 HBW, SN. EP08108286
Model: 289.9 HBW, SN. EP08108292
Model: 281.8 HBW, SN. EP08108298 / ISO 6506-2:2005 / 2 % / 140 - 400 HB / Calibration of brinell hardness tester
Load cell
Model: KTN-D, SN. 54993
Calibers
Model: SCM-40, CN. 21-200-0005-006
Model: S-225, SN. 8279
Protractor
Model: C493B, CN. 21-120-0038-9009
Spirit level
Model: 679, CN. 21-120-0007-001
Stop watch
Model: 220-172100, CN. 307SW / EN 10045-pII:1991 / 3 % / 40 – 400 J / Calibration of impact tester
Torque
Torque transeducers
Model: TTC10, S.N.: 06098001 M-8125
Model: TTC11, S.N.: 06098001 M-8250
Model: TTC14, S.N.: 06098001 M-8200 / ISO 6789:2003 / 1.25 % / 20 - 2500 Nm / Calibration of torque wrench CW
Standard Specifications / Techniques Used / Main Devices / Types of Tests / PropertiesMeasured / Range of Measurements / Materials / Products Tested
Force testing
Reckwell hardness tester
Model: B526R, SN. 97214304 / ASTM E 18:2007 / 20 - 70 HRC / Rockwell hardness test / Metallic materials
Brinell hardness tester
Model: K-10, SN. 912262 / ASTM E 10:2007a / 200 - 400 HB / Brinellhardness test
Universal testing machine
Model: 4210, SN. H0101 / ASTM E 8M:2004 / 10 – 45 T / Tension test
كورنيش المعادى –برج رياض المعادى1- القاهرة – مصر Kornish El-Maadi, Riad El-Maadi Tower 1 - Cairo - Egypt تليفون : 5275227/ 5275225 /5275224 /5275220 (202) Tel.: (202) 5275220/5275224/5275225/5275227 فاكس : 5275226 (202) Fax: (202) 5275226