Schedule of Accreditation
Issued To
Calibration Laboratory
Integrated Calibration Services (ICS)
Flat 101, 7 Mahmoud Abu El Auyoun St., Helipolis, Cairo, Egypt
Valid to : May 31, 2019 / Revision No. (3): May21, 2014 / Issue No. (4): August 13, 2015 / Schedule No.: 20640 BBrief Description of
Measurement and Equipment Used
/Calibrationand Measurement Capability * (±)
/Range
/Measured Quantity
Pressure CalibratorSN: 61051975, 6032042205
with indicatorSN: 2930013 / DKD-R6-1-2003 / 1.3 mbar / 0 – 2 bar / Pressure
Calibration of Pneumatic Gauge
(Permenant)
Pressure Calibrator
SN: 61051975, 6032042205 / DKD-R6-1-2003 / 6.7 mbar / 0 – 20 bar
Pressure Calibrator
SN: PC1556,
with indicatorSN: 3586208 / DKD-R6-1-2003 / 13 mbar / 0 – 60 bar
Pressure Calibrator
SN: PC1556,
with indicatorSN: 2296748 / DKD-R6-1-2003 / 7.2 psi / 0 – 5000 psi
Pressure Calibrator
SN: PC1556,
with indicatorSN: 2296748 / DKD-R6-1-2003 / 11 psi / 0 – 10000 psi / Pressure
Calibration of Fluid Gauge
(Permenant)
Brief Description of
Measurement and Equipment Used
/Calibrationand Measurement Capability * (±)
/Range
/Measured Quantity
Pressure CalibratorSN: 61051975, 6032042205
with indicatorSN: 2930013
Digital Multimeter
SN: 3146A20942 / DKD-R6-1-2003 / 1.8 mbar / 0 – 2 bar / Pressure
Calibration of Pneumatic Transmitter
(Permenant)
Pressure Calibrator
SN: 61051975, 6032042205
Digital Multimeter
SN: 3146A20942 / DKD-R6-1-2003 / 7.0 mbar / 0 – 20 bar
Pressure Calibrator
SN: PC1556 , with indicator SN: 3586208 / DKD-R6-1-2003 / 16 mbar / 0 – 60 bar
Brief Description of
Measurement and Equipment Used
/Calibrationand Measurement Capability * (±)
/Range
/Measured Quantity
Pressure CalibratorSN: 61051975, 6032042205,
with indicatorSN: 2930013 / DKD-R6-1-2003 / 1.9 mbar / 0 – 2 bar / Pressure
Calibration of Pneumatic Gauge
(On Site)
Pressure Calibrator
SN: 61051975, 6032042205 / DKD-R6-1-2003 / 9.3 mbar / 0 – 20 bar
Pressure Calibrator
SN: PC1556,
with indicatorSN: 3586208 / DKD-R6-1-2003 / 19.6 mbar / 0 – 60 bar
Pressure Calibrator
SN: PC1556,
with indicatorSN: 2296748 / DKD-R6-1-2003 / 7.9 psi / 0 – 5000 psi
Pressure Calibrator
SN: PC1556
with indicatorSN: 2296748 / DKD-R6-1-2003 / 11 psi / 0 – 10000 psi / Pressure
Calibration of Fluid Gauge
(On Site)
Brief Description of
Measurement and Equipment Used
/Calibrationand Measurement Capability * (±)
/Range
/Measured Quantity
Pressure CalibratorSN: 61051975, 6032042205
with indicatorSN: 2930013
Digital Multimeter
SN: 3146A20942. / DKD-R6-1-2003 / 2.0 mbar / 0 – 2 bar / Pressure
Calibration of Pneumatic Transmitter
(On Site)
Pressure Calibrator
SN: 61051975, 6032042205
Digital Multimeter
SN: 3146A20942 / DKD-R6-1-2003 / 8.2 mbar / 0 – 20 bar
Pressure Calibrator
SN: PC1556,
with indicator SN: 3586208. / DKD-R6-1-2003 / 25 mbar / 0 – 60 bar
Brief Description of
Measurement and Equipment Used
/Calibrationand Measurement Capability * (±)
/Range
/Measured Quantity
Dry Block CalibratorSN: 095.03.07
Dry Block Calibrator
SN: 292.07.02
Standard Platinum Resistance Thermometer
SN: 294.07.02
Digital Multimeter
SN: 3146A20942 / In-House Method
ICS-CP-T02 / 0.18 ºC / From -20 to 400 ºC / Temperature
Calibration of RTDs
(PermenantOn Site)
Dry Block Calibrator
SN: 095.03.07
Dry Block Calibrator
SN: 292.07.02 / In-House Method
ICS-CP-T03 / 0.11 ºC / From -20 to 400 ºC / Temperature
Calibration of Temperature System
(indicator with Sensors either thermocouples or RTDs / PT100)
(PermenantOn Site)
Dry Block Calibrator
SN: 292.07.02 , SN: 095.03.07
Standard Platinum Resistance Thermometer
SN: 294.07.02
Digital Multimeter
SN: 3146A20942 / In-House Method
ICS-CP-T03-1 / 0.17 ºC / From -20 to 400 ºC
Temperature Calibrator
SN: 526011
Digital Multimeter
SN: 3146A20942 / In-House Method
ICS-CP-T04 / 0.22 ºC / From -20 to 1350 ºC / Temperature
Calibration of Thermocouple Temperature Indicators by Simulation, without Sensors (mV input)
(PermenantOn Site)
Brief Description of
Measurement and Equipment Used
/Calibrationand Measurement Capability * (±)
/Range
/Measured Quantity
Standard ResistanceSN: 29416
Digital Multimeter
SN: 3146A20942 / In-House Method
ICS-CP-T04-1 / 0.18 ºC / From -20 to 800 ºC / Temperature
Calibration of RTD Temperature Indicators by Simulation, without Sensors (ohm input)
(Permenant)
Standard Platinum Resistance Thermometer
SN: 294.07.02
Digital Multimeter
SN: 3146A20942 / In-House Method
ICS-CP-T05 / 0.17 ºC / From -20 to 400 ºC / Temperature
Calibration of Dry Block
(Permenant)
Standard Resistance
SN: 29416
Digital Multimeter
SN: 3146A20942 / In-House Method
ICS-CP-T04-1 / 0.18 ºC / From -20 to 800 ºC / Temperature
Calibration of RTD Temperature Indicators by Simulation, without Sensors (ohm input)
(On Site)
Dry Block Calibrator
SN: 095.03.07
SN: 292.07.02, with
Flexible RTD Class A
SN: ICS-0003-T, ICS-0004-T / In-House Method
ICS-CP-T06 / 0.18 ºC / From -20 to 250ºC / Temperature
Calibration of Ovens, Incubators, Autoclaves, Refrigerators and similars.
(On Site)
TC+RTD Data Logger Thermometer
SN: 11030942 with type K thermocouple. / In-House Method
ICS-CP-T06-1 / 1.6 ºC / From 0 to 800 ºC
(*)Calibration & Measurement Capabilitiesare to be expressedas expanded uncertainty (k=2)i.e. providing a level of confidence of approximately 95%.
كورنيش المعادى –برج رياض المعادى1- القاهرة –مصر Kornish El-Maadi, Riad El-Maadi Tower 1 - Cairo - Egypt تليفون :25275227/ 25275225 /5275224 2 /25275220 (202) Tel.: (202) 25275220/25275224/25275225/25275227 فاكس :25275226 (202)Fax: (202) 25275226