GUIDE TO CYMER ELS 5400

Outline

1Introduction

2Operation

2.1Turn on laser

2.2Setup and Fire

2.3Measure intensity

2.4Turn off laser

3Maintenance

3.1Gas refill

3.2Replace nitrogen cylinder

4Trouble Shooting

5Contact

6Update

Ver 1.1

Call Xinyu Gu at (512)744-8453 if you have any questions.

Introduction

CYMER ELS 5400 is one of the earliest models of KrF laser among CYMER’s products. This laser is a single chamber and outputs 248 nm light. The laser in our lab was generously donated by CYMER Inc.

  1. Getting Start

1.1Turn on laser

There are three modes in CYMER ESL 5400, i.e. long off, standby and on. In the long off mode, laser is off. In the standby mode, the memory and logcould be read by an external PC. In the on mode, the laser is running.

a)Open the main valve on the nitrogen cylinder. (at the corner of the room between New yellow and Old yellow) Do NOT touch the regulator unless the pressure is not between 60 and 90 psi.

b)Open the main valve on the water line.

c)Open the bottom red valve.

d)Switch on the alarm light for New yellow lab.

e)Push “on” button on the remote control paddle. The laser goes into standby mode.

f)Push “on” button again. The laser goes into on mode.

1.2Setup and Fire

a)Push “Burst Count” button, enter the number of pulses you need and press “enter”.

b)Push “Rep Rate” button, enter the number of pulse frequency you need and press “enter”. To keep the laser running safely and consistently, set the REP RATE below 100 Hz.

c)Place your sample above the smelling face and close the door on the exposure box.

d)Push “Shutter” button to open the electric shutter in the laser. “OO” will be shown on the right top corner of the display screen on the remote control paddle instead of “##”.

e)Push the rod switch at the outlet of the laser to open the mechanical shutter. Both the mechanical shutter and the electrical shutter need to be opened for firing laser.

f)Look around and make sure that everybody has protection goggle on.

g)Push “Fire” to shoot laser on your sample.

1.3Measure intensity

a)Plug the smaller photosensor (J-10MB-LE) onto the photometer (Coherent Field MAX II).

b)Plug the power cord onto the photometer.

c)Long press the illumination button to turn on the meter.

d)Open the cap on the photo sensor.

e)Place the sensor in the center of the laser beam. (Do NOT fire the laser right now)

f)Close the door on the exposure box and leave a crack for the wire to the sensor.

g)Set the Burst Count to ZERO and the Rep Rate to FIVE. (ZERO means infinite number of pulses.)

h)Fire the laser.

i)Push “↑” and “↓” to adjust the maximal scale of the meter. (Usually 300 μJ and keep in mind that trigger threshold is 8% of the maximal scale)

j)Push “Stat” button to switch between statistics mode and real time mode. (Statistics mode collects the energies of 30 pulses and calculates the MAX, MIN, MEAN and STDV)

k)Push “Zero” to redo the statistics again.

l)Write down the average energy and make sure that the standard deviation is less than 5%.

m)Calculate the real intensity on your sample. ( usually thin films on wafers)

1.4Measure intensity

a)Close both the mechanical and electric shutters.

b)Press “Off” twice to switch to long off mode.

c)Close the main valve on the nitrogen cylinder.

d)Close the valve on the cooling water line.

  1. Maintenance

2.1Gas refill

As the laser runs for long time, the composition of the gases in the chamber deviates from the ideal values, which lowers the laser performance. Therefore, we have to refill the gas in the chamber in the laser. During the refilling process, the machine first pumps down to exhaust all the old gas to the fume hood and then receives the gases from the cylinders.

In gas room:

a)Open nitrogen.

b)Open the main valve on the Helium cylinder in the blue cabinet.

c)Open the main valves on the right two cylinders which contain Krypton.

In New Yellow:

d)Make sure that the Helium is connected to CYMER laser. Check the valves above the ArF laser machine.

e)Make sure the laser is off.

f)Press “Gas Refill” button on the right bottom corner of the control paddle until the screen shows GAS REFILL.

g)Press “Enter”.

h)Wait until the screen returns back to the initial state and the gas request light comes off on the panel.

In gas room:

i)Close all the mail valves that you just opened in the blue cabinet.

2.2Replace nitrogen cylinder

The nitrogen runs out very fast when the laser is being used. Remember to CLOSE the main valve on the nitrogen cylinder after you are done.

To replace the empty nitrogen cylinder:

a)Disassemble the regulator and put the safety cap back on.

b)Take a cart from the basement.

c)Ask Xinyu for the password to buy the nitrogen gas.

d)Carry the cylinder to RLM 3.104 and get new nitrogen.

e)Put everything back on and return the cart.

Trouble Shooting

Refer to CYMER Manual.

Contact

Israel Gracia

Technical support
Cymer, Inc.
Austin, TX
Cell: (512) 705-4359
Pager: (512) 205-1433
Alpha:

Email:

Update

Xinyu Gu, Version 1.0, June 29th 2009

Xinyu Gu, Version 1.1, July 15th 2009