RFI Attachment B – In Situ Dissolved-Gas Mass Spectrometer (MASSP):
Instrument SpecificHardware & Performance
This table contains Instrument Specific Hardware & Performance Requirements for a mass spectrometer (MASSP) attached to the Regional Scale Nodes (RSN) cabled observatory network. The University of Washington Purchasing Department requests that respondents to this RFI complete and return this form. For those requirements that cannot be met, you may offer an alternative (other*). Respondents may add additional pages, noting the line item, in order to expand upon their responses and comment on the appropriateness of a requirement and/or whether it may be a significant cost-driver.
RESPONDENT INFORMATIONNAME: / PHONE:
ADDRESS 1: / FAX:
ADDRESS 2: / EMAIL:
CITY: / STATE / ZIP:
B-1
RFI Attachment B – In Situ Dissolved-Gas Mass Spectrometer (MASSP):
Instrument SpecificHardware & Performance
SECTION 1 - MASSP HARDWARE REQUIREMENTSITEM / REQUIREMENT / MEET? / OTHER* /COMMENT/ EXPLANATION
MASS-001 / The MASSPshould be capable of monitoring masses in the range of 2 to 200 daltons. This is an objective. / --Select--YESNOOTHER*
MASS-002 / The MASSPshall be capable of measuring dissolved gases with a mass resolution of 0.50 daltons. / --Select--YESNOOTHER*
MASS-003 / The MASSP shall be capable of operating and transmitting data in selective ion monitoring and full scan modes. / --Select--YESNOOTHER*
MASS-004 / The MASSP shall be capable of mass calibration while deployed. / --Select--YESNOOTHER*
MASS-005 / The MASSP should be capable of collecting and transmitting profile data at full mass resolution of the mass spectrometer. This is an objective. / --Select--YESNOOTHER*
MASS-006 / The MASSP should have two electron generating filaments and have the ability to power either upon command. This is an objective. / --Select--YESNOOTHER*
MASS-007 / The MASSP shall be capable of making the measurements with sample fluid temperatures of 2 °C to 40 °C. / --Select--YESNOOTHER*
MASS-008 / The MASSP should be capable of making the measurements with sample fluid temperatures of 2 °C to 75 °C. This is an objective. / --Select--YESNOOTHER*
MASS-009 / The MASSP shall measure the dissolved gases to a water depth of 2000 m. / --Select--YESNOOTHER*
MASS-010 / The MASSP shall be capable of quantitativelymeasuring a single dissolved gas with a response time of 3 mins. / --Select--YESNOOTHER*
MASS-011 / The MASSP shall be capable of quantitativelymeasuring a suite of dissolved gases with a response time of 10 mins.
(See PERFORMANCE section for prioritization of specific gases.) / --Select--YESNOOTHER*
MASS-012 / The MASSP shall be capable of quantitativelymeasuring thesuite of dissolved gasesonce every 6 hours over the deployment period of 13 months. (Provide details regarding limiting factors in following section.) / --Select--YESNOOTHER*
MASS-013 / The MASSPshould be capable of quantitatively measuring thesuite of dissolved gasesonce every hour over the deployment period of 13 months. This is an objective. (Provide details regarding limiting factors in following section) / --Select--YESNOOTHER*
B-1
RFI Attachment B – In Situ Dissolved-Gas Mass Spectrometer (MASSP):
Instrument SpecificHardware & Performance
GENERAL COMMENTS ON MASSP HARDWARE REQUIREMENTS (e.g. reasonableness, significant cost drivers). See RFI for additional information requested:Specifically, please include the most significant factors (e.g. gas waste) that may limit the combination of frequency of measurements with the length of deployment and what technique(s) you have employed to mitigate those factors.
B-1
RFI Attachment B – In Situ Dissolved-Gas Mass Spectrometer (MASSP):
Instrument SpecificHardware & Performance
B-1
RFI Attachment B – In Situ Dissolved-Gas Mass Spectrometer (MASSP):
Instrument SpecificHardware & Performance
Please detail the performance capabilities of your instrumentation by populating the following table:SECTION 2 – MASSPMEASURMENT PERFORMANCE REQUIREMENTS
ITEM / PARAMETER / EXPECTED
IN SITU RANGE
(M) / METHOD
DETECTION
LIMIT (M) / PRACTICAL
QUANTITATION
LIMIT (M) / ACCURACY
(M) / PRECISION
(M) / APPLICABLE
RANGE
(M) / SAMPLE SIZE OR
FLOW RATE
(Specify units)
DGAS-001 / The following gases have the highest priority and are considered the minimal suite of dissolved gases for quantitative analysis using the in-situ MASSP:
-001a / Hydrogen / 50 to 1000
-001b / Methane / 0.002 to 100,000
-001c / Ethane / Unknown
(Methane x 10-5)
-001d / Carbon Dioxide / 2,000 to 100,000
-001e / Argon / ~10 to 20
-001f / Hydrogen Sulfide / 0 to 30,000
DGAS-002 / The following suite of gases also are of interest and quantitative analysis of their concentrations by in-situ MASSP is desired:
-002a / Carbon monoxide / Unknown
-002b / Nitrous Oxide / <0.002 to 0.60
-002c / Oxygen / 0 to 50
-002d / Nitrogen / ~ 500 to 625
-002e / C3-C5 alkanes / Unknown
-002f / C2-C5 alkenes/alkynes / Unknown
-002g / Organosulfur Compounds / Unknown
Enter NM (Not Measured) for those analytes that are either not detectable or not quantifiable using your instrumentation. If your instrumentation is capable of measuring an analyte but a specific parameter is currently unknown, please flag using NA (Not Available) and explain in the comments below. The Applicable Range refers to the metrics of Accuracy and Precision. Sample Size/Sample Flow Rate refer to that required to attain the measurements.
B-1
RFI Attachment B – In Situ Dissolved-Gas Mass Spectrometer (MASSP):
Instrument SpecificHardware & Performance
COMMENTS ON MASSP MEASUREMENT PERFORMANCE REQUIREMENTS (e.g. reasonableness, significant cost drivers, additional descriptions, technology options). See RFI for additional information requested:B-1
RFI Attachment B – In Situ Dissolved-Gas Mass Spectrometer (MASSP):
Instrument SpecificHardware & Performance
B-1
RFI Attachment B – In Situ Dissolved-Gas Mass Spectrometer (MASSP):
Instrument SpecificHardware & Performance
SECTION 3 – MASSPCALIBRATION METHODOLOGIESPlease provide a description of the Calibration Methods that you recommend and/or utilize to enable quantification of the target analytes listed in Section 2 above. If you are able to provide calibration infrastructure as an option, please include a detailed summary of the instrumentation and protocols.
B-1
RFI Attachment B – In Situ Dissolved-Gas Mass Spectrometer (MASSP):
Instrument SpecificHardware & Performance
B-1