EU Declaration of Conformity
Name of Apparatus: Masterflex® Brushless PWM L/S® Pump Drives
Model Number: 7523-80,-90; 7551-00,-10; 7575-10
Description of Apparatus: Brushless PWM, variable speed, peristaltic pump motor drive. Used with pump head and tubing to pump fluids.
Thermo Fisher Scientific declares that the above models are in conformity to the following harmonized standards and directives:
Applicable Directives / Applicable Standards / Manufacturer’s Report Number
2006/95/EC /
EN61010-1:2001
/ TR0105
2004/108/EC / EN61326-1:2006 / TR0105
The last two digits of the year in which the current configuration of the above models was assessed per the Low Voltage Directive is :09.
Manufacturer:
Thermo Fisher Scientific
28W092 Commercial Avenue
Barrington, IL 60010-2392
USA
Tel.: (847) 381-7050
Manufacturer’s Signature:
______/ 17 November 2009
Price Hodson Date
Director of Engineering

Edition 01 Sht 1 of 2 A-1299-5164

Performance Certification to EMC Directive
Normative Standard: EN 61326-1: 2006 Test Report: TR0105
EMISSONS

TEST METHOD

/

LIMITS

CISPR 11/A2:2006

Radiated Emissions /

B

(res / com / lt. indus)

CISPR 11/A2:2006

Conducted Emissions / B
(res / com / lt. indus)

IEC 61000-3-2:2005

Harmonic Currents / A
(motor drive)

IEC 61000-3-3/A2:2005

Flicker / Pass
(soft-start)

IMMUNITY

TEST METHOD

/

STRESS LEVEL

/ PERFORMANCE

IEC61000-4-2/A2: 2000

Electrostatic Discharge / 4kV contact
4kV air / A
IEC61000-4-3/A1: 2002
Radiated RFI / 3 V/m / A
IEC61000-4-4/A2: 2001
Fast transient Burst / 1.0 kV Power Supply
0.5 kV I/O / A
IEC61000-4-5: 2005
Surge Immunity / ±0.5 kV line-line, 10x
±1.0 kV line-earth,10x / A
IEC61000-4-6/A1: 2004
Conducted RFI / 3 Vrms / A
IEC61000-4-11: 2004
Line Voltage Dropout / 0.5 cycle @ 100%
1 cycle @ 100%
0.5 sec @ 30%
5.0 sec @ 100% / A
A
A
C
Performance: A - During testing normal performance within the specification limits.
B - During testing, temporary degradation, or loss of function or performance which is self-
recovering.
C - During testing, temporary degradation, or loss of function or performance which requires
operator interventionor system reset occurs.
D - Degradation or loss of function which is not recoverable due to damage to equipment,
components, software, or to loss of data. Degradation or loss of function resulting in a safety hazard.
Manufacturer:
Thermo Fisher Scientific
28W092 Commercial Avenue
Barrington, IL 60010-2392
USA
Tel.: (847) 381-7050
Manufacturer’s Signature:
______/ 17 November 2009
Price Hodson Date
Director of Engineering

Edition 01 Sht 2 of 2 A-1299-5164