Name of Apparatus: Masterflex® Brushless PWM L/S® Pump Drives
Model Number: 7523-80,-90; 7551-00,-10; 7575-10
Description of Apparatus: Brushless PWM, variable speed, peristaltic pump motor drive. Used with pump head and tubing to pump fluids.
Thermo Fisher Scientific declares that the above models are in conformity to the following harmonized standards and directives:
Applicable Directives / Applicable Standards / Manufacturer’s Report Number
2006/95/EC /
EN61010-1:2001
/ TR01052004/108/EC / EN61326-1:2006 / TR0105
The last two digits of the year in which the current configuration of the above models was assessed per the Low Voltage Directive is :09.
Manufacturer:
Thermo Fisher Scientific
28W092 Commercial AvenueBarrington, IL 60010-2392
USA
Tel.: (847) 381-7050
Manufacturer’s Signature:
______/ 17 November 2009
Price Hodson Date
Director of EngineeringEdition 01 Sht 1 of 2 A-1299-5164
Performance Certification to EMC DirectiveNormative Standard: EN 61326-1: 2006 Test Report: TR0105
EMISSONS
TEST METHOD
/LIMITS
CISPR 11/A2:2006
Radiated Emissions /B
(res / com / lt. indus)CISPR 11/A2:2006
Conducted Emissions / B(res / com / lt. indus)
IEC 61000-3-2:2005
Harmonic Currents / A(motor drive)
IEC 61000-3-3/A2:2005
Flicker / Pass(soft-start)
IMMUNITY
TEST METHOD
/STRESS LEVEL
/ PERFORMANCEIEC61000-4-2/A2: 2000
Electrostatic Discharge / 4kV contact4kV air / A
IEC61000-4-3/A1: 2002
Radiated RFI / 3 V/m / A
IEC61000-4-4/A2: 2001
Fast transient Burst / 1.0 kV Power Supply
0.5 kV I/O / A
IEC61000-4-5: 2005
Surge Immunity / ±0.5 kV line-line, 10x
±1.0 kV line-earth,10x / A
IEC61000-4-6/A1: 2004
Conducted RFI / 3 Vrms / A
IEC61000-4-11: 2004
Line Voltage Dropout / 0.5 cycle @ 100%
1 cycle @ 100%
0.5 sec @ 30%
5.0 sec @ 100% / A
A
A
C
Performance: A - During testing normal performance within the specification limits.
B - During testing, temporary degradation, or loss of function or performance which is self-
recovering.
C - During testing, temporary degradation, or loss of function or performance which requires
operator interventionor system reset occurs.
D - Degradation or loss of function which is not recoverable due to damage to equipment,
components, software, or to loss of data. Degradation or loss of function resulting in a safety hazard.
Manufacturer:
Thermo Fisher Scientific
28W092 Commercial Avenue
Barrington, IL 60010-2392
USA
Tel.: (847) 381-7050
Manufacturer’s Signature:
______/ 17 November 2009
Price Hodson Date
Director of Engineering
Edition 01 Sht 2 of 2 A-1299-5164