CG-6020Nickel/AluminumFilled FIP PasteProperty Report

Mar. 2017

Rev. A

NysteinR&D Center

Contents

Section 1: Overview

Purpose

Product Description

Section 2: Volume Resistivity

Section 3: Mechanical Properties

Density

Hardness

Tensile Strength and Elongation

Tear Strength

Adhesive Force

Static Compression Force Measurements

Section 5: Flammability

Section 6: Reliability Performance

Compression Set

Section 7: Environmental & Safety

RoHS & Halogen Compliance

MSDS

Section 8: Summary

1 / 18

Section1: Overview

Purpose

These measurements reportsare those most needed to describe the product properties, provide information needed by customers and the sales force, and to make decisions regarding the suitability of the product for a given application.

The properties described in this report are relevant to the product’s importantcharacteristic including normal functionreliability test.

Product Description

CG-6020, a Nystein star product, is a hightemperature curing Nickel/Aluminum conductiveFIP gasket. With guaranteed resistance tocorrosion, adhesion strength and conductivity, itfeatures low cost and has competitive advantage.

Featured good elasticity and low compression set,CG-6020 has good adhesion strength on metal andplastic surfaces. The product can be applied tooptical transceivers, telecommunication basestations, radar equipment, and hand held device andconsumer electronics.

1 / 18

Section2: Volume Resistivity

Volume resistivity was measured with pressure probe method according to MIL-DTL-83528C.

1)Having a contacting surface area equal to 0.25 square inch (0.564 inchdiameter) with suitable provisions for attaching ohmmeter leads. Here we use Φ14.3mm*2mm disc.

2)Device is shown in Figure 1. Positiondice between electrodesand test it.

3)Test procedure:

a)Measure and record thickness of material at contact areas to be tested using thickness gauge.Material being tested must have sufficient area to contact entire electrode area.

b)Position material between electrodes and apply pressure of 100±5psi across contact surface area (25pounds load or force).

c)Maintain constant pressure until electrical requirement is met with a maximum time of two minutes.

4)Calculate volume resistivity in ohms-cm using the following formula:

Where:ρ= Volume resistivity (ohm-cm)

R = Observed resistance (ohms)

A = Area of specimen (cm2)

L = Thickness of specimen (cm)

5)Test results

Table 1: Volume Resistivity test with Pressure probe method of CG-6020

Test Item / Result / Spec. / Conclusion
Sample 1 / Sample 2 / Sample 3 / Average
Volume Resistivity / Ohm-cm / 0.039 / 0.056 / 0.041 / 0.045 / ≤0.1 / Pass

1 / 18

Section3: Mechanical Properties

Density

1)Density of CG-6020was measured with a QL-300AW ElectronicDensimeteraccording to standard ASTM D792, method A.

2)Sample preparation: small patch for about 5g.

3)Test procedure:

a)Measure and record the water temperature.Weigh the specimen in air. Record the mass of the specimen, a.

b)Put samples on the support which is immerged in the water.Weight it. Record thisapparent mass as b.

c)The Densimeter would calculate it automatically.

4)Test results are shown in table 2:

Table 2:Density of CG-6020

Test Item / Result / Spec. / Conclusion
Sample 1 / Sample 2 / Sample 3 / Average
Density / (g/cm3) / 1.82 / 1.87 / 1.86 / 1.85 / 1.9±0.2 / Pass

Hardness

1)Shore A hardness was measured in accordance with ASTM D2240 with a manual hardness tester, verified by a traceable calibration block, on 3pcsstack-upcured Φ14mm*2mm slabs.

2)Test procedure

a)Place the specimen on the specimen support table, Actuate the release lever of the operatingstand or activate the electromechanical device, allowing thedurometer to descend at a controlled rate and apply the presserfoot to the specimen. Operate the leveror other mechanism to apply the specimen to the indentor in amanner that assures parallel contact of the specimen to thedurometer presser foot without shock.

b)Oncethe presser footis in contact with the specimen, for example, when the initialindentor travel has ceased, the maximum indicated readingshall be recorded. The time interval of 1 s, between initialindentor travel cessation and the recording of the indicatedreading, shall be considered standard.

3)Test results are shown below:

Table 3: Sample dimension test of CG-6020

Sample dimension / Results / Spec. / Conclusion
Sample 1 / Sample 2 / Sample 3 / Average
H/mm / 5.91 / 5.93 / 5.93 / 5.92 / 6.0+/-0.2 / Pass
Φ/mm / 13.08 / 13.02 / 13.05 / 13.05 / 13.0+/-0.2 / Pass

Table 4: Shore A Hardness of CG-6020

Item / Results / Spec. / Conclusion
Sample 1 / Sample 2 / Sample 3 / Average
Hardness
Shore A / 61 / 62 / 61 / 61 / 60±5 / Pass

Tensile Strength and Elongation

1)The samples for tensile strength and % elongation measurements were done in accordance with ASTM D412, using the universal tensile testing machine. The tests were run at a constant speed of 500mm/min.

2)Sample preparation:Cut 2mm thickness slab into dumbbell specimens with Die C dimension. Dumbbell specimensshall be marked with the bench marker which distance shall be as follows: 25.00+/-0.25 mm.

3)Test procedure:

a)Place the dumbbell in the grips of the testing machine, using care to adjust the specimen symmetrically to distribute tension uniformly over the cross section. The rate of grip separation shall be 500 +/- 50 mm/min.

b)Start the machine and note the distance between the bench marks, taking care to avoid parallax. Record the force at the elongation(s) specified for the test.

c)The elongation measurement is made preferably through the use of an extensometer, an autographic mechanism or a spark mechanism.

d)Calculation:

  1. Tensile strength (PSI):

where:

TS = tensile strength,PSI

Fmax = the maximum force, N

A = cross-sectional area of unstrained specimen, mm2

  1. Elongation:

where:

E = the elongation in percent (of original bench markdistance), %

L = observed distance between bench marks on theextended specimen, mm

L0 = original distance between bench marks,25mm

4)Test results are shown below:

Table 5: Tensile Strength and Elongation of CG-6020

Test Item / Result / Spec. / Conclusion
Sample 1 / Sample 2 / Sample 3 / Average
Tensile Strength / PSI / 171.2 / 170.9 / 173.5 / 171.9 / ≥150 / Pass
Elongation / % / 123 / 120 / 122 / 121.7 / ≥100 / Pass

Tear Strength

1)CG-6020 samples for tear strength testing were prepared according to ASTM-D624. Tests were conducted at a constant speed of 500mm/min using the universal tensile testing machine.

2)Sample preparation:Cut 2mm thickness slab intoType B (crescent-shaped) test piece.

3)Test procedure:

a)Test pieces shall be conditioned & tested at a standard laboratory temperature of 23+/- 2°C

b)Place the test piece in the grips of the testing machine using care to adjust the test piece so that it will be strained uniformly along its length, and that sufficient material is clamped in the grips to minimize slippage.

c)Start the machine at a steady rate of grip separation with 500+/-50 mm/min. Record the maximum force.

d)Calculation the tear strength(PPI):

Where:

Ts = the tear strength, in PPI

F = the maximum force, in N,

d = the median thickness of each test piece, in mm.

4)Test results list below:

Table 6: Tear Strength of CG-6020

Test Item / Result / Spec. / Conclusion
Sample 1 / Sample 2 / Sample 3 / Average
Tear Strength / PPI / 49.3 / 50.4 / 48.9 / 49.5 / ≥45 / Pass

Adhesive Force

1)Standard:Nystein standard,QA-WI-FIP-002

2)Equipment:Push Force Scale

3)Sample Preparation:Dispense H=0.9mm / W=1.1mm D shape beads and cured.

4)Method:

a)Put 10mm width head tap near beads and parallel to it. Then push till break the beads.

b)Record the vales. Make sure choose test point randomly to ensure average result.

5)Test Results:

Table 7: Adhesive Force of CG-6020

Item / Results / Spec. / Conclusion
Sample 1 / Sample 2 / Sample 3 / Average
Adhesive Force
N/cm / 14.18 / 14.21 / 14.46 / 14.28 / ≥10 / Pass

Static Compression Force Measurements

1)Compression force measurements were conducted on H0.9mm*W1.1mm*L10mm D shapestrip of CG-6020, representing static conditions.

2)All strips were extruded then post-cured and put for over 24 hours prior to testing. Data was collected from tests of a new sample strip, measuring the force the first time the strip was compressed. Subsequent compressions will yield a smaller value than the first compression of the strip.

3)Static compression force measurements were made according to NysteinmethodQA-WI-FIP-003. Compression of the strip to a fraction of the original free height was followed by a hold of thirty seconds to allow the strip to equilibrate before the force was measured. The strip was then further compressed, equilibrated, and another measurement performed. Typical results for CG-6020 are shown in Figure 2Table 9, below.

4)The data shown below can help customer to design structure & application.

Table 8: D shape sample dimension of CG-6020

D shape sample dimension / Results / Spec. / Conclusion
Sample 1 / Sample 2 / Sample 3 / Average
H/mm / 0.88 / 0.85 / 0.82 / 0.85 / 0.9+/-0.15 / Pass
W/mm / 1.01 / 1.03 / 1.01 / 1.02 / 1.1+/-0.15 / Pass

Table 9:Static Compression Force, CG-6020H0.9mm*W1.1mm*L10mm strip

Compression (%) / Force (N/cm)
Sample 1 / Sample 2 / Sample 3 / Average
0 / 0 / 0 / 0 / 0
5 / 3 / 4 / 5 / 4
10 / 11 / 10 / 9 / 10
15 / 16 / 15 / 17 / 16
20 / 28 / 25 / 27 / 27
25 / 36 / 37 / 38 / 37
30 / 45 / 43 / 48 / 45
35 / 55 / 54 / 57 / 55
40 / 64 / 62 / 67 / 64
45 / 71 / 75 / 78 / 75
50 / 85 / 84 / 88 / 86
55 / 101 / 102 / 109 / 104
60 / 115 / 119 / 128 / 121

1 / 18

Section 5: Flammability

1)Cured CG-6020 is UL-94 tested and is V-0 rated for sample size is 127mm*12.7mm*2mm.

2)Test procedure:

a)Clamp the specimen from the upper 6 mm of the specimen, with the longitudinal axis vertical, so that thelower end of the specimen is 300 ±10 mm above a horizontal layer of not more than 0.08 g of absorbent 100percent cotton thinned to approximately 50 x 50 mm and a maximum thickness of 6 mm.

b)Adjust the burner to produce a blue flame 20 ±1 mm high.

c)Apply the flame centrally to the middle point of the bottom edge of the specimen so that the top of theburner is 10 ±1 mm below that point of the lower end of the specimen, and maintain it at that distance for10 ±0.5 seconds.

d)After theapplication of the flame to the specimen for 10 ±0.5 seconds, immediately withdraw the burner at a rate ofapproximately 300 mm/sec, to a distance at least 150 mm away from the specimen and simultaneouslycommence measurement of the afterflame time t1 in seconds. Record t1.

e)As soon as afterflaming of the specimen ceases, even if the burner has not been withdrawn to the full150 mm distance from the specimen, immediately place the burner again under the specimen and maintainthe burner at a distance of 10 ±1 mm from the remaining portion of the specimen for an additional 10 ±0.5seconds, while moving the burner clear of dropping material as necessary. After this application of the flameto the specimen, immediately remove the burner at a rate of approximately 300 mm/sec to a distance of at least150 mm from the specimen and simultaneously commence measurement of the afterflame time, t2, and theafterglow time, t3. Record t2 and t3.

3)Judgement:

Criteria conditions / V-0 / V-1 / V-2
Afterflame time for each individual specimen t1 or t2 / 10s / 30s / 30s
Total afterflame time for any condition set (t1 plus t2 for the 5 specimens) / 50s / 250s / 250s
Afterflame plus afterglow time for each individual specimen after the second flame application (t2+t3) / 30s / 60s / 60s
Afterflame or afterglow of any specimen up to the holding clamp / No / No / No
Cotton indicator ignited by flaming particles or drops / No / No / Yes

4)In house test result:

Table 10: Flame retardant test of CG-6020

No. / Standard / Conclusion
1stafterflame time
t1/s≤10s / 2ndafterflame time
t2/s≤10s / 2nd afterglow time
t3/s / Total afterflame time for 5 samples (t1+t2)/s≤50s / Afterflame plus afterglow time for (t2+t3)/s≤30s / Afterflame or afterglow of any specimen up to the holding clamp / Cotton indicator ignited by flaming particles or drops
1 / 3 / 7 / 0 / 46 / 7 / No / No / V-0
2 / 2 / 6 / 0 / 6 / No / No
3 / 3 / 5 / 0 / 5 / No / No
4 / 5 / 6 / 0 / 6 / No / No
5 / 4 / 5 / 0 / 5 / No / No

1 / 18

Section6: ReliabilityPerformance

Compression Set

1)Compression set was measured in accordance with MIL-DTL-83528C / ASTM D395, method B.

2)Sample size: 6+/-0.2mm height, Φ13+/-0.2mm. The compression device is shown in figure 3.

3)Compression set is reported as the percent of compressed distance that was not recovered.

4)Test procedure:

a)Original Thickness Measurement—Measure the original thickness of the specimen.

b)Place the test specimen between the plates of the compression device with the spacers on each side, allowing sufficient clearance for the bulging of the rubber when compressed. Tighten the bolts so that the plates are drawn together uniformly until they are in contact with the spacers. The amount of compression employed shall be approximately 25 %. (spacer’s thickness is 4.50+/-0.01mm).

c)Sample would be test under the condition of 100℃ / 70 hour.

d)Final column height was measured after 30 minutes recovery at room temperature.

5)Calculate the compression set expressed as a percentageof the original deflection as follows:

where:

CB = compression set expressed aspercentage of the original deflection, %

t0 = original thickness of specimen, mm

ti = final thickness of specimen,mm

tn = thickness of the spacer bar used, 4.5mm

Table 11: Compression Set of CG-6020

Test Item / Result / Spec. / Conclusion
Sample 1 / Sample 2 / Sample 3 / Average
t0/ mm / 6.086 / 6.074 / 6.040 / 6.067 / —— / ——
ti/ mm / 5.749 / 5.735 / 5.722 / 5.735 / —— / ——
CB / % / 21.2 / 21.5 / 20.6 / 21.1 / ≤30 / Pass

1 / 18

Section7: Environmental & Safety

RoHS & HalogenCompliance

Testing at SGS-CSTC, Ltd., demonstrated compliance of CG-6020 with RoHS requirements. Documentation is available. Pleasecontact local engineer for it.

MSDS

An MSDS has been prepared forCG-6020 and should be reviewed prior to use and handling of this product.Pleasecontact local engineer for it.

1 / 18

Section8: Summary

CG-6020is a nickel/Aluminum filled silicone elastomer gasket material.It shows good compression set, regardless of conditions of high temperature.

Typical Values

Standard / Unit / Spec / CG-6020
Elastomer / - / - / - / Silicone
Filler / - / - / - / Nickel/Aluminum
Electrical Properties
Volume Resistivity / MIL-DTL-83528C / Ohm-cm / ≤0.1 / 0.1
Shield Effectiveness
500M / MIL-DTL-83528C
Para. 4.5.12 / dB / - / 105
2G / dB / - / 100
10G / dB / - / 95
Physical Properties
Hardness / ASTM D2240 / Shore A / 60±5 / 61
Density / ASTM D792 / g/cm3 / 1.9±0.2 / 1.85
Tensile Strength / ASTM D412 / PSI / ≥150 / 171.9
Elongation / ASTM D412 / % / ≥100 / 121.7
Tear Strength / ASTM D624 / PPI / ≥45 / 49.5
Adhesion Force / QA-WI-FIP-002 / N/cm / ≥10 / 14.28
Static Compression Force / at 20% compression / QA-WI-FIP-003 / N/cm / - / 27
at 40% compression / N/cm / - / 64
Reliability Properties
Compression Set / ASTM D395-B / % / ≤30 / 21.1
Temperature Range
Temperature Range / ASTM D1329 / ℃ / - / -55~+125
Flammability
UL rating / UL94 / — / - / V-0

______END______

1 / 18