CE 321 INTRODUCTION TO FLUID MECHANICS 200SPRINGFALLSPRING 200964
LABORATORY 12: FLOW RATE MEASUREMENT
OBJECTIVESS
- :TtTo measure flow rates,
Tto assess the uncertainty of the measured
- flow rates, and
- Tto to determine how to make accurate flow rate measurements.
EQUIPMENT:
HhHydraulic bench, wrist watch, and stopwatch.
APPROACH
: In this lab you will learn to “measure” flow and assess the uncertainty
of the measured values. We have intentionally selected a method that produces
significant uncertainty at times. Once you understand the sources of this
uncertauncertaiinty, you can refine the measurement procedure to obtain accurate flow
rate measurements in later labs.
PROCEDURE:
A. FAMILIARIZE YOURSELF WITH THE EQUIPMENT.
- 1. Open the dump valve and .s Switch on the power. Open the flow control valve
about ½ turn. (Whenever the power is on, the valve should be open so that
some water flows through the system. Otherwise, the pump will overheat.)
2. Adjust the valve so that maximum flow occurs. O (open valve slowly to
- prevent water from shooting around the lab).
- 3. Trace out the path of water starting at the sump tank.
4. Before making any measurements, fill the measuring tank to the top
several times to remove air trapped in the site gage tube that can cause
measurement errors.Mand make sure that the dump valve seals when it is
closed. This should be done at the beginning of each lab period in which
you measure flow. Before filling the measuring tank, ensure that the sump
- contains adequate water, so that the pump does not run dry.
5. Close the dump valve and watch the water level in the site gage rise as
water accumulates in the measuring tank. Open the dump valve once the
water level in upper site gage reaches 5 liters. Drain the tank completely
and fill it to 5 liters several more times whilesimply observing the system’s
- behavior.
6. Now repeat the drain-and-fill operation and try measuringmeasure the time
required to accumulate 5 liters of water.;Sstart timing when the bottom of
the meniscus1 1in the site gage reaches the 0 liter mark, and stop when it
- reaches the 5-liter mark. Do not bother recording this first set ofese data, as it is simply a trial. Now you are ready to begin collecting data.
1Meniscus1 -the U-shaped upper surface of the water in the site gage. Start timing when the bottom of the meniscus is aligned with the zero mark.
B. COLLECT FLOW-RATE DATA.
1. Open the flow control valve completely, and do not change its position
during the experiment.
2. With the water level well below the 0 level on the upper scaleof the site
gage, measure the time required for the water level to increase from 0 to 1
on the upper scale. You cannot obtain accurate measurements under
these circumstances. That is intentional. Try to obtain measurements that
are as accurate as possible within the limits imposed by the equipment
and procedure.
3. Repeat this measurement 10 times. Record the measured times in
column 2 of Table 1.
4. Without adjusting the flow valve, repeat the procedure of steps 2-3, using
an interval on the site gage from 0 to 5 liters. Record these data in column
4 of Table 1.
C. ESTIMATE THE LEAST COUNT.
1. The smallest change in volume that you can reliably identify using the site
gage is its least count (). Estimate the value of for the measurement
conditions you encountered. Remember that you had to decide when the
water level was exactly zero as the water level was rising. The moving
water level makes it more difficult to identify small changes in volume, so it
should increase your estimate of the least count.
D. TURN OFF THE MOTOR AND OPEN THE DUMP VALVE.
E. COLLECT UNCERTAINTY DATA.
1. Use a quartz wristwatch to define a 5-minute time interval.
2. Measure the length of this interval using one of the lab stopwatches.
3. If the two watches show less than 1% difference, keep the stopwatch.
Otherwise, get a different one and try again.
4. Use the stopwatch to time a 15-second interval on the lab clock. Make
sure you time the lab clock. Be sure to stand so that you minimize
parallax while watching the sweep second hand on the lab clock. WAnd
watch the clock, not the stopwatch. Make 10 measurements and record
them in Table 2. Don’t throw out any of them out!
ANALYSIS
A. CALCULATE AND PLOT THE FLOW RATES.
Put the calculated values in Table 1. Plot the flow rates as a function of the
measurement volume as shown in the attached example.
Assuming steady flow, describe any uncertainty that your flow measurements
suggest. Describe how the measurement volume and time appear to
influence uncertainty.
B. ESTIMATE THE UNCERTAINTIES IN YOUR BASIC MEASUREMENTS.
Assume the uncertainty (V)V in each measured volume is 2(twice the least
count). Estimate the uncertainty t to be twice the standard deviation
observed in the values of t. That is t = 2SD t. Determine the standard
deviation (SDSDt) of the 10 measurements of t using the STDEV function in
EXCEL.
C. ESTIMATE THE UNCERTAINTY IN Q DUE TO UNCERTAINTIES IN V AND T.
Use the propagation of uncertainty equation derived in Lab and your values of
V and t to calculate dQ/Q.
Describe what these calculations show.: Do they suggest that all of the
uncertainty you observed in Q comes from your uncertainty in V and t? Do
they suggest that uncertainty in Q varies as you change the size of the
measurement volume and/or the measuring time? If V and t are the only
sources of error in the experiment, does this equation suggest that you could
obtain measurements that have only 1% uncertainty at this flow rate? (This
final question is for discussion, but not necessary for the lab report.)
D. EXPLORE THE UNCERTAINTY IN Q THAT IS POSSIBLE FOR DIFFERENT MEASUREMENT
SITUATIONS.
Use the propagation of uncertainty equation derived in Lab to develop a table
that shows how dQ/Q varies as a function of Q and V. In this equation, the
filling time (t) is equal to the volume (V) divided by the flow (Q). Use values
of 0.1 l/s Q 0.7 l/s and 2 l/s V 10 l/s.
Later during thisin the semester you will run experiments where the flow is set
using the valve and determine the flow rate by measuring volume and time where the flow is set using the valve..
In order to assess the uncertainty in calculated values you will need to obtain
values of Q that have less uncertainty than a specified amount. Use the
approach suggested by your lab instructor and your data to develop a method
for making precise measurements of Q. Remember that you will not know the
value of Q before you measure V and t.
E. DETERMINE A METHOD FOR MEASURING FLOW ACCURATELY IN FUTURE
EXPERIMENTS.
Use the information developed in part D and assume that it will be sufficient to
obtain values of Q that have an uncertainty less than 5%. Remember, you will
not know the flow rate until after you have measured it., and y Your method must
take this into account. Describe your method and provide an example of how
you will use it to measure an unknown flow accurately.
CHECK LIST FOR THE FLOW MEASUREMENT LAB
DATA APPEAR TO BE REASONABLY ACCURATE
Your calculated values of Q should show a trend similar to what can be observed
in Sample Fig. 1.
Check the values of your uncertainty estimates. Are they reasonable given the
measurement device and the circumstances?
RESULTS SECTION
Your facts are correct, clearly stated, and supported by the data you present.
Did you describe any uncertainty that your flow measurements suggest? Did you
describe how the measurement volume and time appear to influence uncertainty
in flow?
Did you state the amount of uncertainty that you estimated your measured values
are subject to? Did you describe the method of estimating you used to make the
estimates?
Did you describe what your uncertainty calculations show about the values of Q
that you measured? Do they suggest that all of the uncertainty you observed in
Q came from your uncertainty in V and t? Do they suggest that uncertainty in Q
varies as you change the size of the measurement volume and/or the measuring
time? If V and t are the only sources of uncertainty in the experiment, does this
equation suggest that you could obtain measurements that have only 1%
uncertainty at this flow rate? (This last question is for discussion and is not
necessary for the lab report.)
DISCUSSION
The logical arguments are correct and clearly stated.
Did you describe the method you have developed for accurately measuring flow
in future experiments? Will your method meet the accuracy criteria?
Did you present a logical argument that justifies the method? Did you describe
how you would account for the fact that you will not know the flow rate until after
you have measured it?
Did you provide an example that shows how you can use your method to
measure an unknown flow accurately?
Make sure the important results and conclusions are restated in the Conclusions
Section and summarized in the Abstract
.
CHECK LIST FOR THE FLOW MEASUREMENT LAB
Data appear to be reasonably accurate.
Your calculated values of Q should show a trend similar to what can be observed
in Sample Fig. 1.
Check the values of your uncertainty estimates. Are they reasonable given the
measurement device and the circumstances?
Your facts are correct, clearly stated, and supported by the data you present.
Focus on the Results Section.
Your facts are correct, clearly stated, and supported by the data you present.
Did you describe any uncertainty that your flow measurements suggest? Did you
describe how the measurement volume and time appear to influence uncertainty
in flow?
Did you state the amount of uncertainty that you estimated your measured values
are subject to? Did you describe the method of estimating you used to make the
estimates?
Did you describe what your uncertainty calculations show about the values of Q
that you measured? Do they suggest that all of the uncertainty you observed in
Q came from your uncertainty in V and t? Do they suggest that uncertainty in Q
varies as you change the size of the measurement volume and/or the measuring
time? If V and t are the only sources of uncertainty in the experiment, does this
equation suggest that you could obtain measurements that have only 1%
uncertainty at this flow rate? (This last question is for discussion and is not
necessary for the lab report.)
The logical arguments are correct and clearly stated. Focus on the Ddiscussion.
The logical arguments are correct and clearly stated.
Did you describe the method you have developed for accurately measuring flow
in future experiments? Will your method meet the accuracy criteria?
Did you present a logical argument that justifies the method? Did you describe
how you would account for the fact that you will not know the flow rate until after
you have measured it?
Did you provide an example that shows how you can use your method to
measure an unknown flow accurately?
Make sure the important results and conclusions are restated in the Conclusions
Section and summarized in the Abstract.
1