NAME:______
DEPARTMENT: ______
DATE: ______
1. The Bowie-Dick test should be run:
a.Weekly
b.Monthly
c.Daily before the first processed load
d.During routine maintenance
2. The Bowie-Dick test is designed to evaluate the effectiveness of the air removal system of which kind of steam sterilizer?
- Gravity-displacement
- Prevacuum
- Both a and b
3. The Bowie-Dick test is designed to detect residual air in the sterilizer chamber due to:
- Inadequate air removal
- Air leaks
- Non-condensable gases in the steam
- All of the above
4. The appropriate placement of the Bowie-Dick Test Pack in the sterilizer is:
- On the floor over the drain
- On a basket over the drain
- Horizontally, label side up, on the bottom shelf of the sterilizer rack, over the drain in the first full load of the day
- Horizontally, label side up, on the bottom shelf of the sterilizer rack, over the drain in an otherwise empty chamber
5. It is recommended to run a warm-up cycle before running the Bowie-Dick test to ensure accurate results and avoid false failures.
True
False
6. To save time, it is acceptable to run the Bowie-Dick test pack in the same cycle as the biological indicator test pack.
True
False
7. When running the Bowie-Dick test, the sterilizer should be programmed to run which of the following cycles?
- 270-273ºF prevacuum for 4 minutes or longer with 20 minute dry time
- 270-273ºF prevacuum for 3 ½ - 4 minutes with 1 minute dry time
- 270-273ºF gravity-displacement for 3 ½ - 4 minutes with 1 minute dry time
8. The 3M Comply 00135LF Bowie-Dick Plus Test Pack contains:
- A biological indicator
- A Bowie-Dick Test Sheetonly
- A Bowie-Dick Test Sheet and an Early Warning Test Sheet
9. A uniform color-change of the primary test sheetfrom yellow to dark brown/black is considered a pass.
True
False
10. If the indicator dot on theEarly Warning Test Sheet shows a non-uniform color change but the primary test sheet shows a pass, what should you do?
- Use the sterilizer and schedule routine maintenance
- Take the sterilizer out of use and call a repair person
- Ignore the result
11. If a major repair of the sterilizer is required, qualification testing should be conducted before the sterilizer is put back into routine use.
True
False
12. During sterilizer qualification testing, the Bowie-Dick test is:
- Run in three consecutive cycles after running three consecutive biological indicator test pack cycles
- Run in three consecutive cycles before running three consecutive biological indicator test pack cycles
- Not needed
- Run once
13. For flash sterilizers, the Bowie-Dick test:
- Is not needed
- Should be done daily if the sterilizer is used for prevacuum cycles
- None of the above
14. After the Bowie-Dick test results are interpreted and recorded, the test sheets:
- Can be thrown away
- Should be maintained in a paper record keeping system or scanned into an electronic record keeping systemso they are available for future comparison
- None of the above
NAME:______
DEPARTMENT: ______
DATE: ______
Answers:
- c
- b
- d
- d
- True
- False
- b
- c
- True
- a
- True
- a
- b
- b