BERKSHIRE HEALTH SYSTEM LABORATORIES
BerkshireMedicalCenter
FairviewHospital
POLICY:Pre-Surgical Staphylococcus for Methicillin-ResistantStaphylococcus Aureus (MRSA) and Methicillin-SensitiveStaphylococcus Aureus (MSSA) Screening by PCR
Purpose:
Staphylococcus bacteria are extremely common in the environment, and are usually not a problem for healthy individuals. Historically, staph infections begin only after the individual has suffered a skin break or open wound. Recently, strains of Staphylococcus aureus have developed a resistance toMethicillin, common antibiotics used to treat infections. These infections have been reported worldwide and outbreaks have occurred in many community settings and among different populations.
In an effort to control further spread of this strain Methicillin-ResistantStaphylococcus Aureus (MRSA), BerkshireMedicalCenterHospital has put into place a series of precautionary steps to check for the possible existence of MRSA and MSSA in all orthopedic surgical patients who will be admitted to the hospital after surgery.
Procedure:
Copan swab (red lettering) must be used for PCR testing.
Nasal Specimen Collection and Transport:
To obtain adequate specimen, follow the instructions in this section closely.
- Open the Collection Device by peeling back the outer packaging.
- Ask the patient to tilt his/her head back. Insert dry swabs approximately 1 – 2 cm into each nostril.
- Rotate the swabs against the inside of the nostril for 3 seconds. Apply slight pressure with a finger on the outside of the nose to help assure good contact between the swab and the inside of the nose.
- Using the same swabs, repeat for the second nostril, trying not to touch anything but the inside of the nose.
- Remove the plastic transport tube. Twist off the tube cap and discard it. Place the swabs into the plastic transport tube. The swabs should go all the way into the tube until they rest on top of the sponge at the bottom of the tube. Make sure the redcap is on tightly. Note: the swabs should stay attached to the redcap at all times.
- Label the plastic transport tube with the patient ID and place them in a biohazard bag. Each patient’s swabs should be placed in an individual biohazard bag for transport to the laboratory.
- Store swab specimen at room temperature (15- 30° C) if it will be processed within 24 hours, otherwise store swab at 2 – 8° C. The swab specimen is stable up to 5 days when stored at 2 - 8° C.
Reference: Cepheid – molecular diagnostics
Sunnyvale, CA
Ordering Instructions for Meditech:
- Book testing for Mondays and Tuesdays – this allows for turnaround time for testing – results available by end of week.
- Lab order code: “SAS”
Miscellaneous Instruction:
- Supplies: Call the storeroom at BMC for additional supplies at 447-2487
- Technical consultation: Call Dr. Rick Intres at 447-2599
Prepared by: C. Jenks new 07-01-09
Revised / 1-16-12
Revised: / 4-24-12 C. Jenks
Reviewed
Reviewed
Policy ID: Mol/Micro PreSurg Staph Screen