Heather Garen

Digital Image Qualities

A radiographer views an image to determine if it is acceptable photographically and

geometrically. To compensate for a quality they would have to determine what needs to be

changed to provide an adequate picture. The following factors may have to be altered to make

an image sufficient.

Contrast is the difference between adjacent densities on a radiograph. It is the

distinctness of black and white. As photons pass through an object they are absorbed, pass

through, or slightly pass through reaching the IR. In tissues where absorption differs

visualization of densities varies. An image with few densities with strong differences is one with

high contrast. This can also be defined as a long scale contrast. In conclusion it can be

described as long and low, or short and high. The anatomic tissue radiographed and kVp used

determines contrast. Low kVp used results in higher contrast. Thinner the patient the higher

the contrast, kVp would have to be adjusted to get that same contrast level on a larger patient.

Scatter radiation decreases contrast adding to unwanted density. When OID is

increased it creates an air gap causing low contrast. Collimation increased results in less scatter

reaching IR so high contrast. Tube filtration increases contrast by decreasing low energy

photons. Contrast can be altered by window width on the monitor after processing. The more

pixel depth and number the higher contrast. A wide dynamic range also helps have a high

contrast short scale of gray.

Spatial resolution con simply is defined as the blurring of an image, an edge’s degree of

sharpness on an image. It is determined by multiple factors. Including increasing the number

of pixels in the image matrix increases SP. Increasing OID decreases sp, object is close to the IR

making no magnification. Increase in SID would increase sharpness. Larger the FSS the worse

the SP. Add tube, patient, or IR angulations and you will decrease SP. Motion from the patient

ofxray equipment will cause poor SP. In the reader the smaller and faster the laser spot size

the better. The better the pixel depth the better the SP. The greater crystal size in the cassette

the less spatial resolution. CR angle increase decreases detail because it increases distortion.

Recorded detail is the accuracy of structural lines recorded.Optimal geometric quality is

achieved by maximizing the amount of recorded detail and decreasing distortion. As FSS

increases, unsharpness increases and recorded detail increases. So small FSS is chosen for

better recorded detail. Increasing the SID decreases the amount of unsharpness and increases

the amount of recorded detail in the image. Source of photons is closer so less scatter results in

less added density which inturn makes better recorded detail high contrast. Increasing the OID

is inversely proportional to recorded detail (degree of sharpness). The more OID causes an air

gap that scatter is made and diverging beams hit IR.

Density is the amount of brightness the image has. Picture must have sufficient density

to demonstrate all essential anatomy. Controlling factors would include mAs, influencing

factors are the following. The 15% rule states that changing the kVp by 15% will have the same

effect on density as doubling the mAs, or reducing the mAs by half. SID affects the quanity of

radiation reaching the IR. As SID increase density decrease. As SID increases the xray beam

intensity is spread over a larger area and more diverging beam hits IR. As SID increases the

density decreases as a result of the square of the distance. OID is increased less density results.

With this air gap less intensity of the beam and diverging beam will strike the receptor.

Increasing collimation makes field size smaller which decreases density, less scatter to add

density. Anode heel effect decrease the primary beam on the anode end of the tube making

density appears not uniformed on an image. A patient’s part thickness changes density. The

thicker the part the less radiation reaching IR so decrease in density. Generator must be

calibrated for consistent output or it will affect density. Less tube filtration adds density.Adding

a compensating filter to the primary beam produces more uniform density, but requires an

increase in mAs.